Comparison of two strip test methods of whole blood glucose measurement in the neonatal period.

M Papp1, N Sharief.   

Abstract

UNLABELLED: The aim of this study was to compare the performance and accuracy of the BM Strip test used in conjunction with Reflectance photometry, and the new non-wipe strip test (Advantage) against a reference plasma glucose method. In total, 114 newborns consecutively admitted to the Neonatal Unit over a 6 mo period were enrolled into the study. Each newborn had their venous blood glucose measured by the BM Strip test and Advantage glucometer and the venous haematocrit was also determined. Plasma glucose was measured in the laboratory by the hexokinase method. The mean difference between the BM Strip test and plasma glucose was significantly less than the corresponding value for the Advantage glucometer (0.312, 95% confidence interval (CI) 0.11-0.51 vs 0.766, 95% CI 0.57-0.95], although the limits of agreement between both methods and plasma glucose were wide. Haematocrit did not influence significantly the accuracy of either test.
CONCLUSION: The new Advantage glucose meter does not offer any advantage over the BM Strip test. Owing to the wide limits of agreement of both methods compared with plasma glucose, their clinical value is limited in the neonatal period.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11683193     DOI: 10.1080/080352501316978138

Source DB:  PubMed          Journal:  Acta Paediatr        ISSN: 0803-5253            Impact factor:   2.299


  3 in total

1.  Evaluation of different POCT devices for glucose measurement in a clinical neonatal setting.

Authors:  Matthias Roth-Kleiner; Corinne Stadelmann Diaw; Jocelyne Urfer; Christiane Ruffieux; Dominique Werner
Journal:  Eur J Pediatr       Date:  2010-06-24       Impact factor: 3.183

2.  Evaluation of "point of care" devices in the measurement of low blood glucose in neonatal practice.

Authors:  H T Ho; W K Y Yeung; B W Y Young
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2004-07       Impact factor: 5.747

3.  Evaluation of three glucometers for whole blood glucose measurements at the point of care in preterm or low-birth-weight infants.

Authors:  Joon Ho Hwang; Yong-Hak Sohn; Seong-Sil Chang; Seung Yeon Kim
Journal:  Korean J Pediatr       Date:  2015-08-21
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.