Management of patient dose and image noise in routine pediatric CT abdominal examinations.

Francis R Verdun1, D Lepori, Pascal Monnin, Jean-François Valley, Pierre Schnyder, F Gudinchet.   

Abstract

The aim was to propose a strategy for finding reasonable compromises between image noise and dose as a function of patient weight. Weighted CT dose index (CTDI(w)) was measured on a multidetector-row CT unit using CTDI test objects of 16, 24 and 32 cm in diameter at 80, 100, 120 and 140 kV. These test objects were then scanned in helical mode using a wide range of tube currents and voltages with a reconstructed slice thickness of 5 mm. For each set of acquisition parameter image noise was measured and the Rose model observer was used to test two strategies for proposing a reasonable compromise between dose and low-contrast detection performance: (1) the use of a unique noise level for all test object diameters, and (2) the use of a unique dose efficacy level defined as the noise reduction per unit dose. Published data were used to define four weight classes and an acquisition protocol was proposed for each class. The protocols have been applied in clinical routine for more than one year. CTDI(vol) values of 6.7, 9.4, 15.9 and 24.5 mGy were proposed for the following weight classes: 2.5-5, 5-15, 15-30 and 30-50 kg with image noise levels in the range of 10-15 HU. The proposed method allows patient dose and image noise to be controlled in such a way that dose reduction does not impair the detection of low-contrast lesions. The proposed values correspond to high- quality images and can be reduced if only high-contrast organs are assessed.

Entities:  

Mesh:

Year:  2004        PMID: 14722730     DOI: 10.1007/s00330-003-2206-z

Source DB:  PubMed          Journal:  Eur Radiol        ISSN: 0938-7994            Impact factor:   5.315


  25 in total

1.  Technique factors and image quality as functions of patient weight at abdominal CT.

Authors:  W Huda; E M Scalzetti; G Levin
Journal:  Radiology       Date:  2000-11       Impact factor: 11.105

2.  Dose reduction in pediatric CT: a rational approach.

Authors:  John M Boone; Estella M Geraghty; J Anthony Seibert; Sandra L Wootton-Gorges
Journal:  Radiology       Date:  2003-08       Impact factor: 11.105

3.  Dose optimization in pediatric CT: current technology and future innovations.

Authors:  Christoph Suess; Xiaoyan Chen
Journal:  Pediatr Radiol       Date:  2002-08-29

4.  Helical CT of the body: are settings adjusted for pediatric patients?

Authors:  A Paterson; D P Frush; L F Donnelly
Journal:  AJR Am J Roentgenol       Date:  2001-02       Impact factor: 3.959

5.  Effective dose in paediatric computed tomography.

Authors:  C L Chapple; S Willis; J Frame
Journal:  Phys Med Biol       Date:  2002-01-07       Impact factor: 3.609

6.  Radiation exposure in multi-slice versus single-slice spiral CT: results of a nationwide survey.

Authors:  G Brix; H D Nagel; G Stamm; R Veit; U Lechel; J Griebel; M Galanski
Journal:  Eur Radiol       Date:  2003-04-10       Impact factor: 5.315

7.  Radiation exposure and image quality in chest CT examinations.

Authors:  J G Ravenel; E M Scalzetti; W Huda; W Garrisi
Journal:  AJR Am J Roentgenol       Date:  2001-08       Impact factor: 3.959

8.  [The smart-scan procedure of spiral computed tomography. A new method for dose reduction].

Authors:  S M Giacomuzzi; B Erckert; T Schöpf; M C Freund; P Springer; A Dessl; W Jaschke
Journal:  Rofo       Date:  1996-07

9.  Dose reduction in CT examination of children by an attenuation-based on-line modulation of tube current (CARE Dose).

Authors:  Holger Greess; Anton Nömayr; Heiko Wolf; Ulrich Baum; Michael Lell; Bernhard Böwing; Willi Kalender; Werner A Bautz
Journal:  Eur Radiol       Date:  2002-02-02       Impact factor: 5.315

10.  An approach for the estimation of effective radiation dose at CT in pediatric patients.

Authors:  W Huda; J V Atherton; D E Ware; W A Cumming
Journal:  Radiology       Date:  1997-05       Impact factor: 11.105

View more
  15 in total

1.  Multislice CT of the pelvis: dose reduction with regard to image quality using 16-row CT.

Authors:  Jessen Gurung; M Fawad Khan; Adel Maataoui; C Herzog; R Bux; H Bratzke; Hanns Ackermann; Thomas J Vogl
Journal:  Eur Radiol       Date:  2005-04-02       Impact factor: 5.315

2.  Evaluation of chronic infectious interstitial pulmonary disease in children by low-dose CT-guided transthoracic lung biopsy.

Authors:  Christoph M Heyer; Stefan P Lemburg; Thomas Kagel; Klaus-Michael Mueller; Thomas G Nuesslein; Christian H L Rieger; Volkmar Nicolas
Journal:  Eur Radiol       Date:  2005-01-26       Impact factor: 5.315

3.  Dosimetric changes with computed tomography automatic tube-current modulation techniques.

Authors:  Sofia Spampinato; Anna Maria Gueli; Pietro Milone; Luigi Angelo Raffaele
Journal:  Radiol Phys Technol       Date:  2018-04-06

Review 4.  CT dose reduction in children.

Authors:  Peter Vock
Journal:  Eur Radiol       Date:  2005-08-10       Impact factor: 5.315

5.  Detection of liver metastases under 2 cm: comparison of different acquisition protocols in four row multidetector-CT (MDCT).

Authors:  Amina Abdelmoumene; Patrick Chevallier; Marc Chalaron; Frédéric Schneider; Francis R Verdun; Phillipe Frascarolo; Reto Meuli; Pierre Schnyder; Alban Denys
Journal:  Eur Radiol       Date:  2005-05-03       Impact factor: 5.315

6.  Low-dose multidetector-row CT angiography of the infra-renal aorta and lower extremity vessels: image quality and diagnostic accuracy in comparison with standard DSA.

Authors:  Francesco Fraioli; Carlo Catalano; Alessandro Napoli; Marco Francone; Fiammetta Venditti; Massimiliano Danti; Federica Pediconi; Roberto Passariello
Journal:  Eur Radiol       Date:  2005-06-30       Impact factor: 5.315

7.  Low contrast medium and radiation dose for hepatic computed tomography perfusion of rabbit VX2 tumor.

Authors:  Cai-Yuan Zhang; Yan-Fen Cui; Chen Guo; Jing Cai; Ya-Fang Weng; Li-Jun Wang; Deng-Bin Wang
Journal:  World J Gastroenterol       Date:  2015-05-07       Impact factor: 5.742

8.  Assessment of image quality of a standard and three dose-reducing protocols in adult cranial CT.

Authors:  Sadi Gündoğdu; Kamran Mahmutyazicioğlu; Hüseyin Ozdemir; Ahmet Savranlar; Kiyasettin Asil
Journal:  Eur Radiol       Date:  2004-12-02       Impact factor: 5.315

9.  Model-based iterative reconstruction: effect on patient radiation dose and image quality in pediatric body CT.

Authors:  Ethan A Smith; Jonathan R Dillman; Mitchell M Goodsitt; Emmanuel G Christodoulou; Nahid Keshavarzi; Peter J Strouse
Journal:  Radiology       Date:  2013-10-29       Impact factor: 11.105

10.  CT radiation dose in children: a survey to establish age-based diagnostic reference levels in Switzerland.

Authors:  Francis R Verdun; Daniel Gutierrez; John Paul Vader; Abbas Aroua; Leonor Trinidad Alamo-Maestre; François Bochud; François Gudinchet
Journal:  Eur Radiol       Date:  2008-04-04       Impact factor: 5.315

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.