[The smart-scan procedure of spiral computed tomography. A new method for dose reduction].

S M Giacomuzzi1, B Erckert, T Schöpf, M C Freund, P Springer, A Dessl, W Jaschke.   

Abstract

PURPOSE: In this article a new method - the Smart-Scan-Technique - for dose reduction during spiral computed tomography is described. This technique allows a dynamical adaptation of the tube current according to the measured local density structure form and absorption values of the object of interest. Its clinical applicability for dose reduction is investigated. In addition, changes in image quality are evaluated.
METHODS: Specially designed, water filled plexiglas-phantoms, an Alderson-Rando phantom, and 20 patients were used to simulate possible measures of patients within the thoracic and abdominal region. For those two regions the dose reduction, the modulation transfer function, and the pixel noise were measured.
RESULTS: Depending on patient geometry dose reduction up to 20% could be achieved with the Smart-Scan-Technique. This technique also enabled a decrease of the modulation transfer function and an increase of the pixel noise by 8%.
CONCLUSION: Relevant dose reductions can be achieved by applying the Smart-Scan-Technique in spiral computed tomography.

Entities:  

Mesh:

Year:  1996        PMID: 8765357     DOI: 10.1055/s-2007-1015707

Source DB:  PubMed          Journal:  Rofo        ISSN: 1438-9010


  7 in total

1.  Management of patient dose and image noise in routine pediatric CT abdominal examinations.

Authors:  Francis R Verdun; D Lepori; Pascal Monnin; Jean-François Valley; Pierre Schnyder; F Gudinchet
Journal:  Eur Radiol       Date:  2004-01-13       Impact factor: 5.315

2.  Left ventricle volume measurements in cardiac micro-CT: the impact of radiation dose and contrast agent.

Authors:  Cristian T Badea; Arthur W Wetzel; Nilesh Mistry; Stuart Pomerantz; Demian Nave; G Allan Johnson
Journal:  Comput Med Imaging Graph       Date:  2008-04       Impact factor: 4.790

Review 3.  Computed tomography dose optimisation in cystic fibrosis: A review.

Authors:  Helena Ferris; Maria Twomey; Fiachra Moloney; Siobhan B O'Neill; Kevin Murphy; Owen J O'Connor; Michael Maher
Journal:  World J Radiol       Date:  2016-04-28

4.  Improvement of image quality of multislice spiral CT scans of the head and neck region using a raw data-based multidimensional adaptive filtering (MAF) technique.

Authors:  Ulrich Baum; Katharina Anders; Gregor Steinbichler; Michael Lell; Holger Greess; Thomas Riedel; Marc Kachelriess; Willi A Kalender; Werner A Bautz
Journal:  Eur Radiol       Date:  2004-08-24       Impact factor: 5.315

5.  Fetal doses to pregnant patients from CT with tube current modulation calculated using Monte Carlo simulations and realistic phantoms.

Authors:  Jianwei Gu; X George Xu; Peter F Caracappa; Bob Liu
Journal:  Radiat Prot Dosimetry       Date:  2012-12-06       Impact factor: 0.972

6.  Low-dose cone-beam CT via raw counts domain low-signal correction schemes: Performance assessment and task-based parameter optimization (Part II. Task-based parameter optimization).

Authors:  Daniel Gomez-Cardona; John W Hayes; Ran Zhang; Ke Li; Juan Pablo Cruz-Bastida; Guang-Hong Chen
Journal:  Med Phys       Date:  2018-04-06       Impact factor: 4.071

7.  Evaluation of AAPM Reports 204 and 220: Estimation of effective diameter, water-equivalent diameter, and ellipticity ratios for chest, abdomen, pelvis, and head CT scans.

Authors:  Christiane S Burton; Timothy P Szczykutowicz
Journal:  J Appl Clin Med Phys       Date:  2017-11-27       Impact factor: 2.102