Literature DB >> 9114097

An approach for the estimation of effective radiation dose at CT in pediatric patients.

W Huda1, J V Atherton, D E Ware, W A Cumming.   

Abstract

PURPOSE: To estimate the effective radiation dose to pediatric and adult patients at head and abdomen computed tomography (CT).
MATERIALS AND METHODS: Cylindrical water-equivalent phantoms were modeled for patients aged newborn to adult, and the energy imparted per unit axial exposure was computed. To determine the energy imparted to the simulated patients of different ages undergoing head and abdomen CT examinations, x-ray technique factors were combined with measured CT axial exposures. Body-region-specific ratios were calculated for effective dose per unit energy imparted, and these ratios were corrected for patient mass to obtain the effective dose to simulated patients.
RESULTS: With use of standard techniques, the energy imparted to simulated patients at CT always increased with patient size, but the effective dose was higher in children than in adults. At CT in the head and abdomen, effective doses were highest in newborns. Effective doses ranged from 1.5 to 6.0 mSv in head CT examinations and from 3.1 to 5.3 mSv in abdomen CT examinations.
CONCLUSION: The values for energy imparted at CT in pediatric patients were generally lower than in adults. The smaller mass of children, however, caused the corresponding effective doses to be higher than those in adults undergoing similar CT examinations.

Entities:  

Mesh:

Year:  1997        PMID: 9114097     DOI: 10.1148/radiology.203.2.9114097

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  45 in total

1.  Estimating cancer risks from pediatric CT: going from the qualitative to the quantitative.

Authors:  David J Brenner
Journal:  Pediatr Radiol       Date:  2002-03-07

2.  Management of patient dose and image noise in routine pediatric CT abdominal examinations.

Authors:  Francis R Verdun; D Lepori; Pascal Monnin; Jean-François Valley; Pierre Schnyder; F Gudinchet
Journal:  Eur Radiol       Date:  2004-01-13       Impact factor: 5.315

3.  Comparison of doses for pulmonary embolism detection with helical CT and pulmonary angiography.

Authors:  Arnaud Resten; Franck Mausoleo; M Valero; Dominique Musset
Journal:  Eur Radiol       Date:  2002-07-04       Impact factor: 5.315

Review 4.  Exposing the thyroid to radiation: a review of its current extent, risks, and implications.

Authors:  Bridget Sinnott; Elaine Ron; Arthur B Schneider
Journal:  Endocr Rev       Date:  2010-07-21       Impact factor: 19.871

5.  Third-generation dual-source 70-kVp chest CT angiography with advanced iterative reconstruction in young children: image quality and radiation dose reduction.

Authors:  Oliver Rompel; Martin Glöckler; Rolf Janka; Sven Dittrich; Robert Cesnjevar; Michael M Lell; Michael Uder; Matthias Hammon
Journal:  Pediatr Radiol       Date:  2016-01-06

6.  Evaluation of chronic infectious interstitial pulmonary disease in children by low-dose CT-guided transthoracic lung biopsy.

Authors:  Christoph M Heyer; Stefan P Lemburg; Thomas Kagel; Klaus-Michael Mueller; Thomas G Nuesslein; Christian H L Rieger; Volkmar Nicolas
Journal:  Eur Radiol       Date:  2005-01-26       Impact factor: 5.315

7.  Converting dose-length product to effective dose at CT.

Authors:  Walter Huda; Kent M Ogden; Mohammad R Khorasani
Journal:  Radiology       Date:  2008-09       Impact factor: 11.105

8.  Age-specific effective doses for pediatric MSCT examinations at a large children's hospital using DLP conversion coefficients: a simple estimation method.

Authors:  Karen E Thomas; Bo Wang
Journal:  Pediatr Radiol       Date:  2008-04-08

9.  Patient-based estimation of organ dose for a population of 58 adult patients across 13 protocol categories.

Authors:  Pooyan Sahbaee; W Paul Segars; Ehsan Samei
Journal:  Med Phys       Date:  2014-07       Impact factor: 4.071

Review 10.  [Low dose multislice CT in the pediatric patient].

Authors:  A Hojreh; F Kainberger; S Puig
Journal:  Radiologe       Date:  2003-12       Impact factor: 0.635

View more