Effects of motion, ambient light, and hypoperfusion on pulse oximeter function.

N S Trivedi1, A F Ghouri, N K Shah, E Lai, S J Barker.   

Abstract

STUDY
OBJECTIVE: To compare the performance of five pulse oximeters during hypoperfusion, probe motion, and exposure to ambient light interference.
DESIGN: Prospective study.
SETTING: Laboratory facility at a university medical center. PATIENTS: 8 unanesthetized, ASA physical status I volunteers.
INTERVENTIONS: We evaluated five common pulse oximeters with respect to three scenarios: (1) an operating room light was shone on oximeter probes, (2) a motion generator was used to generate 2 Hz and 4 Hz hand motion, and (3) a pneumatic compression device overlying the brachial artery was used to simulate hypoperfusion. Electrocardiographic (ECG) and arterial blood gas values were considered gold standards for heart rate (HR) and oxygen saturation (SpO2) respectively. SpO2 nondisplay and values greater than 4% from simultaneous arterial SaO2-oximeter values were defined as errors. Nondisplay of HR, or HR greater than 5% from ECG values, were also considered errors.
MEASUREMENTS AND MAIN RESULTS: The Ohmeda and Nellcor N200 with finger probe had the highest total failure rates with respect to both SpO2 and HR due to ambient light interference (p < 0.05). The Nellcor N200 with finger probe and N200 with C lock were the most accurate with regard to SpO2 during 2 Hz and 4 Hz motion (p < 0.05). However, all oximeters failed dramatically during 4 Hz motion when measuring HR. In the hypoperfusion model, the Nellcor N200 with finger probe and the Nellcor C Lock oximeters performed significantly better than all others in terms of both HR and SpO2 (P < 0.05), while the Criticare oximeter failed 100% of the time.
CONCLUSION: There are significant differences in the accuracy of commercially available pulse oximeters during nonideal circumstances, with failure rates varying from approximately 5% to 50% depending on the oximeter and source of interference. Furthermore, no single oximeter performed the best under all conditions.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9172022     DOI: 10.1016/s0952-8180(97)00039-1

Source DB:  PubMed          Journal:  J Clin Anesth        ISSN: 0952-8180            Impact factor:   9.452


  22 in total

1.  Decreased accuracy of pulse oximetry measurements during low perfusion caused by sepsis: Is the perfusion index of any value?

Authors:  Helmut D Hummler; Anja Engelmann; Frank Pohlandt; Josef Högel; Axel R Franz
Journal:  Intensive Care Med       Date:  2006-06-30       Impact factor: 17.440

2.  Validation of oxygen saturation monitoring in neonates.

Authors:  Shyang-Yun Pamela K Shiao; Ching-Nan Ou
Journal:  Am J Crit Care       Date:  2007-03       Impact factor: 2.228

Review 3.  Intermittent Hypoxemia in Preterm Infants.

Authors:  Juliann M Di Fiore; Peter M MacFarlane; Richard J Martin
Journal:  Clin Perinatol       Date:  2019-06-15       Impact factor: 3.430

4.  Esophageal pulse oximetry is more accurate and detects hypoxemia earlier than conventional pulse oximetry during general anesthesia.

Authors:  Guo Chen; Zhaoqiong Zhu; Jin Liu; Wei Wei
Journal:  Front Med       Date:  2012-10-02       Impact factor: 4.592

Review 5.  Improving cardiac surgical care: a work systems approach.

Authors:  Douglas A Wiegmann; Ashley A Eggman; Andrew W Elbardissi; Sarah Henrickson Parker; Thoralf M Sundt
Journal:  Appl Ergon       Date:  2010-03-03       Impact factor: 3.661

6.  A comparison of finger and forehead pulse oximeters in heart failure patients during maximal exercise.

Authors:  Kevin L Kelly; Alex R Carlson; Thomas G Allison; Bruce D Johnson
Journal:  Heart Lung       Date:  2019-11-14       Impact factor: 2.210

7.  The analysis of transesophageal oxygen saturation photoplethysmography from different signal sources.

Authors:  Ling Mou; Quan Gong; Wei Wei; Bo Gao
Journal:  J Clin Monit Comput       Date:  2013-03-10       Impact factor: 2.502

8.  Accuracy of pulse oximetry readings in an animal model of low perfusion caused by emerging pneumonia and sepsis.

Authors:  Helmut D Hummler; Anja Engelmann; Frank Pohlandt; Josef Högel; Axel R Franz
Journal:  Intensive Care Med       Date:  2004-01-13       Impact factor: 17.440

9.  Remote plethysmographic imaging using ambient light.

Authors:  Wim Verkruysse; Lars O Svaasand; J Stuart Nelson
Journal:  Opt Express       Date:  2008-12-22       Impact factor: 3.894

10.  Forehead reflectance oximetry: a clinical comparison with conventional digit sensors during laparotomic and laparoscopic abdominal surgery.

Authors:  Andrea Casati; Grazia Squicciarini; Marco Baciarello; Marta Putzu; Alessandra Salvadori; Guido Fanelli
Journal:  J Clin Monit Comput       Date:  2007-08-16       Impact factor: 2.502

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.