Literature DB >> 8923081

Impedance cardiography. Importance of the equation and the electrode configuration.

B J van der Meer1, H H Woltjer, A M Sousman, W O Schreuder, E R Bulder, M A Huybregts, P M de Vries.   

Abstract

OBJECTIVE: Electrical impedance cardiography (EIC) has been suggested as a non-invasive method to measure cardiac output. In several studies it proved to be a reliable method, although there were some restrictions. In 1966 Kubicek et al. developed an impedance cardiac output system based upon electrodes and a specific stroke volume formula. In 1983 Sramek et al. developed a new electrode configuration, and a new equation to calculate stroke volume, an equation that was adjusted by Bernstein in 1986. Since then these two methods have been used in clinical medicine. The purpose of the present study was to compare both electrode configurations and both stroke volume calculation equations with each other. The cardiac output (CO) values obtained by means of EIC are compared with CO values obtained by means of thermodilution.
DESIGN: Prospective study.
SETTING: Surgical intensive care unit of a university hospital. PATIENTS: 20 mechanically ventilated patients after cardiac surgery. MEASUREMENTS AND
RESULTS: Simultaneous measurement of CO by means of electrical impedance cardiography (COEIC) and thermodilution (COTD) was performed. COEIC was obtained using the lateral spot electrode configuration (LS) and an adjusted circular electrode configuration (SC). The formulas of Sramek (S), Sramek-Bernstein (SB), Kubicek (K) and an adjusted Kubicek formula (aK) were employed. Using the LS electrode configuration, significant differences were found between COEIC and COTD with the S formula (p < 0.005), the K formula (p < 0.001), and the aK formula (p < 0.05). Using the SC electrode configuration, significant differences between COEIC and COTD were found with the K formula (p < 0.005), the S formula (p < 0.01), and the SB formula (p < 0.05). No significant differences was found between EIC and TD using the LS electrode configuration together with the SB formula or using the SC electrode configuration with the aK formula. In both cases a good correlation was found between COEIC and COTD (r = 0.86, p < 0.001 and r = 0.79, p < 0.001, respectively). The mean difference between EIC and TD was 0.15 +/- 0.96 1/min and 0.19 +/- 1.19 1/min, respectively.

Entities:  

Mesh:

Year:  1996        PMID: 8923081

Source DB:  PubMed          Journal:  Intensive Care Med        ISSN: 0342-4642            Impact factor:   17.440


  10 in total

1.  Determination of cardiac output during positive end-expiratory pressure--noninvasive electrical bioimpedance compared with standard thermodilution.

Authors:  G Castor; G Molter; J Helms; I Niedermark; P Altmayer
Journal:  Crit Care Med       Date:  1990-05       Impact factor: 7.598

2.  Relationship of bioimpedance to thermodilution and echocardiographic measurements of cardiac function.

Authors:  F G Spinale; A C Smith; F A Crawford
Journal:  Crit Care Med       Date:  1990-04       Impact factor: 7.598

3.  Comparison of two impedance cardiographic techniques for measuring cardiac output in critically ill patients.

Authors:  R W Gotshall; V C Wood; D S Miles
Journal:  Crit Care Med       Date:  1989-08       Impact factor: 7.598

4.  A new stroke volume equation for thoracic electrical bioimpedance: theory and rationale.

Authors:  D P Bernstein
Journal:  Crit Care Med       Date:  1986-10       Impact factor: 7.598

5.  Comparison of measurements of cardiac output by bioimpedance and thermodilution in severely ill surgical patients.

Authors:  P L Appel; H B Kram; J Mackabee; A W Fleming; W C Shoemaker
Journal:  Crit Care Med       Date:  1986-11       Impact factor: 7.598

6.  Development and evaluation of an impedance cardiac output system.

Authors:  W G Kubicek; J N Karnegis; R P Patterson; D A Witsoe; R H Mattson
Journal:  Aerosp Med       Date:  1966-12

7.  Comparison between spot and band electrodes and between two equations for calculations of stroke volume by means of impedance cardiography.

Authors:  H H Woltjer; B J van der Meer; H J Bogaard; P M de Vries
Journal:  Med Biol Eng Comput       Date:  1995-05       Impact factor: 2.602

8.  Noninvasive assessment of cardiac performance by impedance cardiography: disagreement between two equations to estimate stroke volume.

Authors:  C de Mey; D Enterling
Journal:  Aviat Space Environ Med       Date:  1988-01

9.  Thoracic resistivity for stroke volume calculation in impedance cardiography.

Authors:  A W Quail; F M Traugott; W L Porges; S W White
Journal:  J Appl Physiol Respir Environ Exerc Physiol       Date:  1981-01

10.  Continuous noninvasive real-time monitoring of stroke volume and cardiac output by thoracic electrical bioimpedance.

Authors:  D P Bernstein
Journal:  Crit Care Med       Date:  1986-10       Impact factor: 7.598

  10 in total
  7 in total

1.  Descending aortic flow contribution to intrathoracic impedance-development and preliminary testing of a dual impedance model.

Authors:  A Barry Baker; Chris N McLeod; Alastair J Roxburgh; Paul Bannister
Journal:  J Clin Monit Comput       Date:  2007-11-15       Impact factor: 2.502

Review 2.  Impedance cardiography: more questions than answers.

Authors:  David J Wang; Stephen S Gottlieb
Journal:  Curr Heart Fail Rep       Date:  2006-09

3.  Stroke volume equation for impedance cardiography.

Authors:  D P Bernstein; H J M Lemmens
Journal:  Med Biol Eng Comput       Date:  2005-07       Impact factor: 2.602

4.  Lack of agreement between thermodilution and electrical velocimetry cardiac output measurements.

Authors:  Matthias Heringlake; Ulrich Handke; Thorsten Hanke; Frank Eberhardt; Jan Schumacher; Hartmut Gehring; Hermann Heinze
Journal:  Intensive Care Med       Date:  2007-08-10       Impact factor: 17.440

Review 5.  Impedance cardiography: more questions than answers.

Authors:  David J Wang; Stephen S Gottlieb
Journal:  Curr Cardiol Rep       Date:  2006-05       Impact factor: 2.931

Review 6.  Noninvasive continuous hemodynamic monitoring.

Authors:  Jasper Truijen; Johannes J van Lieshout; Wilbert A Wesselink; Berend E Westerhof
Journal:  J Clin Monit Comput       Date:  2012-06-14       Impact factor: 2.502

7.  Accuracy, Precision, and Trending Ability of Electrical Cardiometry Cardiac Index versus Continuous Pulmonary Artery Thermodilution Method: A Prospective, Observational Study.

Authors:  P B W Cox; A M den Ouden; M Theunissen; L J Montenij; A G H Kessels; M D Lancé; W F F A Buhre; M A E Marcus
Journal:  Biomed Res Int       Date:  2017-10-09       Impact factor: 3.411

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.