Stroke volume equation for impedance cardiography.

D P Bernstein1, H J M Lemmens.   

Abstract

The study's goal was to determine if cardiac output (CO), obtained by impedance cardiography (ICG), would be improved by a new equation N, implementing a square root transformation for dZ/dtmax/Z0, and a variable magnitude, mass-based volume conductor Vc. Pulmonary artery catheterisation was performed on 106 cardiac surgery patients pre-operatively. Post-operatively, thermodilution cardiac output (TDCO) was simultaneously compared with ICG CO. dZ/dtmax/Z0 and Z0 were obtained from a proprietary bioimpedance device. The impedance variables, in addition to left ventricular ejection time TLVE and patient height and weight, were input using four stroke volume (SV) equations: Kubicek (K), Sramek (S), Sramek-Bernstein (SB), and a new equation N. CO was calculated as SV x heart rate. Data are presented as mean +/- SD. One way repeated measures of ANOVA followed by the Tukey test were used for inter-group comparisons. Bland-Altman methods were used to assess bias, precision and limits of agreement. P< 0.05 was considered statistically significant. CO implementing N (6.06 +/- 1.48 l min(-1)) was not different from TDCO (5.97 +/- 1.41 l min(-1)). By contrast, CO calculated using K (3.70 +/- 1.53 l min(-1)), S (4.16 +/- 1.83 l min(-1)) and SB (4.37 +/- 1.82 l min(-1)) was significantly less than TDCO. Bland-Altman analysis showed poor agreement between TDCO and K, S and SB, but not between TDCO and N. Compared with TDCO, equation N, using a square-root transformation for dZ/dtmax/Z0, and a mass-based Vc, was superior to existing transthoracic impedance techniques for SV and CO determination.

Entities:  

Mesh:

Year:  2005        PMID: 16255425     DOI: 10.1007/bf02344724

Source DB:  PubMed          Journal:  Med Biol Eng Comput        ISSN: 0140-0118            Impact factor:   2.602


  50 in total

Review 1.  The non-invasive assessment of stroke volume and cardiac output by impedance cardiography: a review.

Authors:  D G Newman; R Callister
Journal:  Aviat Space Environ Med       Date:  1999-08

Review 2.  Impedance cardiography for cardiac output estimation in pacemaker patients: review of the literature.

Authors:  I Ovsyshcher; S Furman
Journal:  Pacing Clin Electrophysiol       Date:  1993-07       Impact factor: 1.976

Review 3.  The technique of impedance cardiography.

Authors:  H H Woltjer; H J Bogaard; P M de Vries
Journal:  Eur Heart J       Date:  1997-09       Impact factor: 29.983

4.  Electric properties of flowing blood and impedance cardiography.

Authors:  K R Visser
Journal:  Ann Biomed Eng       Date:  1989       Impact factor: 3.934

5.  Interpretation of the HJ interval of the normal ballistocardiogram based on the principle of conservation of momentum and aortic ultrasonic Doppler velocity measurements during left ventricular ejection.

Authors:  W S Seitz; M B McIlroy
Journal:  Cardiovasc Res       Date:  1988-08       Impact factor: 10.787

6.  Impedance cardiography. Importance of the equation and the electrode configuration.

Authors:  B J van der Meer; H H Woltjer; A M Sousman; W O Schreuder; E R Bulder; M A Huybregts; P M de Vries
Journal:  Intensive Care Med       Date:  1996-10       Impact factor: 17.440

7.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

8.  The effect of lung injury and excessive lung fluid, on impedance cardiac output measurements, in the critically ill.

Authors:  L A Critchley; R M Calcroft; P Y Tan; J Kew; J A Critchley
Journal:  Intensive Care Med       Date:  2000-06       Impact factor: 17.440

9.  Gender and age differences in transthoracic bioimpedance.

Authors:  G Metry; B Wikström; T Linde; B G Danielson
Journal:  Acta Physiol Scand       Date:  1997-10

10.  Investigation of the parallel conductor model of impedance cardiography by means of exchange transfusion with stroma free haemoglobin solution in the dog.

Authors:  K R Visser; R Lamberts; W G Zijlstra
Journal:  Cardiovasc Res       Date:  1987-09       Impact factor: 10.787

View more
  35 in total

1.  Bernstein-Osypka stroke volume equation for impedance cardiography: citation correction.

Authors:  Donald P Bernstein
Journal:  Intensive Care Med       Date:  2007-03-27       Impact factor: 17.440

2.  Physiological quality assessment of stored whole blood by means of electrical measurements.

Authors:  Y Ulgen; M Sezdi
Journal:  Med Biol Eng Comput       Date:  2007-06-28       Impact factor: 2.602

3.  Descending aortic flow contribution to intrathoracic impedance-development and preliminary testing of a dual impedance model.

Authors:  A Barry Baker; Chris N McLeod; Alastair J Roxburgh; Paul Bannister
Journal:  J Clin Monit Comput       Date:  2007-11-15       Impact factor: 2.502

4.  Does obesity affect the non-invasive measurement of cardiac output performed by electrical cardiometry in children and adolescents?

Authors:  Luis Altamirano-Diaz; Eva Welisch; Ralf Rauch; Michael Miller; Teresa Sohee Park; Kambiz Norozi
Journal:  J Clin Monit Comput       Date:  2017-02-17       Impact factor: 2.502

5.  Bioimpedance analysis: a guide to simple design and implementation.

Authors:  Kevin R Aroom; Matthew T Harting; Charles S Cox; Ravi S Radharkrishnan; Carter Smith; Brijesh S Gill
Journal:  J Surg Res       Date:  2008-05-07       Impact factor: 2.192

6.  Accuracy and precision of non-invasive cardiac output monitoring by electrical cardiometry: a systematic review and meta-analysis.

Authors:  M Sanders; S Servaas; C Slagt
Journal:  J Clin Monit Comput       Date:  2019-06-07       Impact factor: 2.502

7.  Stroke volume and cardiac output evaluation by electrical cardiometry: accuracy and reference nomograms in hemodynamically stable preterm neonates.

Authors:  A Boet; G Jourdain; S Demontoux; D De Luca
Journal:  J Perinatol       Date:  2016-04-21       Impact factor: 2.521

8.  Continuous non-invasive cardiac output monitoring during exercise: validation of electrical cardiometry with Fick and thermodilution methods.

Authors:  Y H Liu; B P Dhakal; C Keesakul; R M Kacmarek; G D Lewis; Y Jiang
Journal:  Br J Anaesth       Date:  2016-07       Impact factor: 9.166

9.  Beat-to-beat estimation of stroke volume using impedance cardiography and artificial neural network.

Authors:  S M M Naidu; Prem C Pandey; Uttam R Bagal; Suhas P Hardas
Journal:  Med Biol Eng Comput       Date:  2017-11-18       Impact factor: 2.602

10.  Electrical velocimetry for non-invasive monitoring of the closure of the ductus arteriosus in preterm infants.

Authors:  Ana Rodríguez Sánchez de la Blanca; M Sánchez Luna; N González Pacheco; M Arriaga Redondo; N Navarro Patiño
Journal:  Eur J Pediatr       Date:  2017-12-08       Impact factor: 3.183

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.