Bedside hemoglobin measurements: sensitivity to changes in serum protein and electrolytes.

S E McNulty1, S J Sharkey, H Schieren.   

Abstract

OBJECTIVE: Our objective was to compare the effect of protein and electrolyte changes associated with hemodilution on the accuracy of photometric and conductivity hemoglobin determination methods.
METHODS: Blood samples from 10 patients with normal preoperative serum electrolytes and total protein levels were studied. From an indwelling arterial line, 20 ml of blood were removed; hemoglobin values were measured pre-(Baseline) and postdilution by Coulter counter, conductivity, and photometric methods. Blood samples were diluted by placing 4 ml of blood into three test tubes, and adding 1 ml of either 25% albumin, 0.9% sodium chloride, or 5% dextrose in water.
RESULTS: Blood sample dilution resulted in a reported conductivity hemoglobin that was significantly different from the Coulter value (p = 0.0004) when 25% albumin, 0.9% sodium chloride, and 5% dextrose in water solution was used. Using the same dilutions, the photometric method accurately reflected Coulter hemoglobin values. The correlation between photometric and Coulter hemoglobin measurements was R2 = 0.97, p = 0.0001. Correcting the conductivity hemoglobin values for changes in total protein, chloride and sodium significantly improved correlation with Coulter hemoglobin (R2 of uncorrected versus corrected = 0.37 and 0.72; p = 0.0001).
CONCLUSIONS: In the range of electrolyte and protein concentrations found in this study, the photometric method of hemoglobin assessment was more accurate than either corrected or uncorrected conductivity hemoglobin determinations, as compared to Coulter-based measurements.

Entities:  

Mesh:

Substances:

Year:  1994        PMID: 7836972     DOI: 10.1007/bf01618418

Source DB:  PubMed          Journal:  J Clin Monit        ISSN: 0748-1977


  11 in total

1.  A clinical evaluation of the Hemocue haemoglobinometer using capillary, venous and arterial samples.

Authors:  P P Chen; T G Short; D H Leung; T E Oh
Journal:  Anaesth Intensive Care       Date:  1992-11       Impact factor: 1.669

2.  New electronic method for measuring hematocrit: clinical evaluation.

Authors:  J A KERNEN; H WURZEL; R OKADA
Journal:  J Lab Clin Med       Date:  1961-04

3.  A comparison of conductivity-based hematocrit determinations with conventional laboratory methods in autologous blood transfusions.

Authors:  D J McMahon; R L Carpenter
Journal:  Anesth Analg       Date:  1990-11       Impact factor: 5.108

4.  Hemocue--a new haemoglobinometer in the clinic.

Authors:  W K Be; H E Kerkkamp; L H Booij
Journal:  Eur J Anaesthesiol       Date:  1991-01       Impact factor: 4.330

5.  Measurement of the electric resistance of human blood; use in coagulation studies and cell volume determinations.

Authors:  R L ROSENTHAL; C W TOBIAS
Journal:  J Lab Clin Med       Date:  1948-09

6.  The electrical conductivity of blood. I: Relationship to erythrocyte concentration.

Authors:  F G HIRSCH; E C TEXTER; L A WOOD; W C BALLARD; F E HORNA; I S WRIGHT
Journal:  Blood       Date:  1950-11       Impact factor: 22.113

7.  An evaluation of the HemoCue for correcting the haemoglobin value of lipaemic samples.

Authors:  P D Nicholls
Journal:  Med Lab Sci       Date:  1990-07

8.  Fetal hemoglobin quantitations using the Hemocue system are rapid and accurate.

Authors:  S M Berry; M P Dombrowski; W B Blessed; J A Bichalski; T B Jones; D B Cotton
Journal:  Obstet Gynecol       Date:  1993-03       Impact factor: 7.661

9.  Evaluation of "HemoCue," a new device for determining hemoglobin.

Authors:  H von Schenck; M Falkensson; B Lundberg
Journal:  Clin Chem       Date:  1986-03       Impact factor: 8.327

10.  Evaluation of STAT-CRIT hematocrit determination in comparison to Coulter and centrifuge: the effects of isotonic hemodilution and albumin administration.

Authors:  S E McNulty; S J Sharkey; B Asam; J H Lee
Journal:  Anesth Analg       Date:  1993-04       Impact factor: 5.108

View more
  5 in total

1.  Conductivity-based hematocrit measurement during cardiopulmonary bypass.

Authors:  Jacoline Steinfelder-Visscher; Patrick W Weerwind; Steven Teerenstra; Gheorghe A M Pop; René M H J Brouwer
Journal:  J Clin Monit Comput       Date:  2006-11-04       Impact factor: 2.502

2.  Hemoglobin test result variability and cost analysis of eight different analyzers during open heart surgery.

Authors:  Kirti P Patel; Gary W Hay; Mahesh Keitheri Cheteri; David W Holt
Journal:  J Extra Corpor Technol       Date:  2007-03

3.  The use of the iSTAT portable analyzer in patients undergoing cardiopulmonary bypass.

Authors:  N R Connelly; M Magee; B Kiessling
Journal:  J Clin Monit       Date:  1996-07

4.  Clinical evaluation of the ABL-77 for point-of-care analysis in the cardiovascular operating room.

Authors:  Jack S Prichard; John S French; Nestor Alvar
Journal:  J Extra Corpor Technol       Date:  2006-06

5.  Evaluation of the HB-Quick: a portable hemoglobinometer.

Authors:  A K Gong; B Backenstose
Journal:  J Clin Monit Comput       Date:  1999-05       Impact factor: 2.502

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.