Literature DB >> 12578078

Monitoring of respiratory rate in postoperative care using a new photoplethysmographic technique.

L Nilsson1, A Johansson, S Kalman.   

Abstract

OBJECTIVE: Photoplethysmography (PPG) is a non-invasive optical technique that measures variations in skin blood volume and perfusion. The PPG signal contains components that are synchronous with respiratory and cardiac rhythms. We undertook this study to evaluate PPG for monitoring patients' respiratory rate in the postoperative care unit, using a new prototype device. We compared it with the established technique, transthoracic impedance (TTI).
METHODS: PPG signals from 16 patients (ASA classes 1-2, mean age 43 years) who were recovering from general anaesthesia after routine operations were recorded continuously for 60 minutes/patient. The respiratory synchronous part of the PPG signal was extracted by using a bandpass filter. Detection of breaths in the filtered PPG signals was done both visually and by using an automated algorithm. In both procedures, the detected breaths were compared with the breaths detected in the TTI reference.
RESULTS: A total of 10,661 breaths were recorded, and the mean +/- SD respiratory rate was 12.3 +/- 3.5 breaths/minute. When compared with TTI, the rates of false positive and false negative breaths detected by PPG (visual procedure) were 4.6 +/- 4.5% and 5.8 +/- 6.5%, respectively. When using the algorithm for breath detection from PPG, the rates of false positive and false negative breaths were 11.1 +/- 9.7% and 3.7 +/- 3.8%, respectively, when compared to TTI. Lower respiratory rates increased the occurrence of false-positive breaths that were detected by the PPG using visual identification (p < 0.05). The same tendency was seen with the automated PPG procedure (p < 0.10).
CONCLUSIONS: Our results indicate that PPG has the potential to be useful for monitoring respiratory rate in the postoperative period.

Entities:  

Mesh:

Year:  2000        PMID: 12578078     DOI: 10.1023/a:1011424732717

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  23 in total

Review 1.  Skin photoplethysmography--a review.

Authors:  A A Kamal; J B Harness; G Irving; A J Mearns
Journal:  Comput Methods Programs Biomed       Date:  1989-04       Impact factor: 5.428

2.  Relationship between hypoxaemic and cardiac ischaemic events in the perioperative period.

Authors:  N P Gill; B Wright; C S Reilly
Journal:  Br J Anaesth       Date:  1992-05       Impact factor: 9.166

3.  Long-term registration of cutaneous microcirculation during general anesthesia.

Authors:  S Ziege; H Schmid-Schönbein; R Grebe; E Martin
Journal:  Int J Microcirc Clin Exp       Date:  1997 Nov-Dec

4.  Periodic cardiovascular and ventilatory activity during midazolam sedation.

Authors:  D C Galletly; T B Williams; B J Robinson
Journal:  Br J Anaesth       Date:  1996-04       Impact factor: 9.166

5.  The respiratory inductive plethysmograph: a new non-invasive monitor of respiration.

Authors:  M A Cohn; A S Rao; M Broudy; S Birch; H Watson; N Atkins; B Davis; F D Stott; M A Sackner
Journal:  Bull Eur Physiopathol Respir       Date:  1982 Jul-Aug

6.  Thoracic impedance used for measuring chest wall movement in postoperative patients.

Authors:  G B Drummond; A F Nimmo; R A Elton
Journal:  Br J Anaesth       Date:  1996-09       Impact factor: 9.166

7.  Evaluation of a respiratory rate monitor in postsurgical patients.

Authors:  W Anderson; A J Brock-Utne; J G Brock-Utne; J B Brodsky
Journal:  J Clin Anesth       Date:  1992 Jul-Aug       Impact factor: 9.452

8.  Postoperative mental confusion--association with postoperative hypoxemia.

Authors:  J Rosenberg; H Kehlet
Journal:  Surgery       Date:  1993-07       Impact factor: 3.982

9.  Randomized evaluation of pulse oximetry in 20,802 patients: II. Perioperative events and postoperative complications.

Authors:  J T Moller; N W Johannessen; K Espersen; O Ravlo; B D Pedersen; P F Jensen; N H Rasmussen; L S Rasmussen; T Pedersen; J B Cooper
Journal:  Anesthesiology       Date:  1993-03       Impact factor: 7.892

10.  Postanesthesia monitoring revisited: frequency of true and false alarms from different monitoring devices.

Authors:  L Wiklund; B Hök; K Ståhl; A Jordeby-Jönsson
Journal:  J Clin Anesth       Date:  1994 May-Jun       Impact factor: 9.452

View more
  29 in total

1.  Influence of tidal volume and thoraco-abdominal separation on the respiratory induced variation of the photoplethysmogram.

Authors:  A Johansson; T Strömberg
Journal:  J Clin Monit Comput       Date:  2000       Impact factor: 2.502

2.  Neural network for photoplethysmographic respiratory rate monitoring.

Authors:  A Johansson
Journal:  Med Biol Eng Comput       Date:  2003-05       Impact factor: 2.602

3.  Respiratory variations in the reflection mode photoplethysmographic signal. Relationships to peripheral venous pressure.

Authors:  L Nilsson; A Johansson; S Kalman
Journal:  Med Biol Eng Comput       Date:  2003-05       Impact factor: 2.602

Review 4.  Critical review of non-invasive respiratory monitoring in medical care.

Authors:  M Folke; L Cernerud; M Ekström; B Hök
Journal:  Med Biol Eng Comput       Date:  2003-07       Impact factor: 2.602

5.  An algorithm for the detection of individual breaths from the pulse oximeter waveform.

Authors:  Paul Leonard; Neil R Grubb; Paul S Addison; David Clifton; James N Watson
Journal:  J Clin Monit Comput       Date:  2004-12       Impact factor: 2.502

6.  The use of joint time frequency analysis to quantify the effect of ventilation on the pulse oximeter waveform.

Authors:  Kirk H Shelley; Aymen A Awad; Robert G Stout; David G Silverman
Journal:  J Clin Monit Comput       Date:  2006-06-01       Impact factor: 2.502

7.  Using the morphology of photoplethysmogram peaks to detect changes in posture.

Authors:  Stephen P Linder; Suzanne M Wendelken; Edward Wei; Susan P McGrath
Journal:  J Clin Monit Comput       Date:  2006-05-11       Impact factor: 2.502

8.  Age and gender do not influence the ability to detect respiration by photoplethysmography.

Authors:  Lena Nilsson; Tomas Goscinski; Anders Johansson; Lars-Göran Lindberg; Sigga Kalman
Journal:  J Clin Monit Comput       Date:  2006-10-11       Impact factor: 2.502

9.  New concept using Passive Infrared (PIR) technology for a contactless detection of breathing movement: a pilot study involving a cohort of 169 adult patients.

Authors:  V Hers; D Corbugy; I Joslet; P Hermant; J Demarteau; B Delhougne; G Vandermoten; J P Hermanne
Journal:  J Clin Monit Comput       Date:  2013-04-03       Impact factor: 2.502

10.  Determination of saturation, heart rate, and respiratory rate at forearm using a Nellcor™ forehead SpO2-saturation sensor.

Authors:  Jarkko Harju; Antti Vehkaoja; Ville Lindroos; Pekka Kumpulainen; Sasu Liuhanen; Arvi Yli-Hankala; Niku Oksala
Journal:  J Clin Monit Comput       Date:  2016-10-17       Impact factor: 2.502

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.