Replacing DXA scanners: cross-calibration with phantoms may be misleading.

G M Blake1.   

Abstract

The high precision and stable calibration of dual X-ray absorptiometry (DXA) has led to its widespread use in clinical trials of new therapies for osteoporosis. Daily scanning of phantoms provides a check on the continuity of the bone mineral density (BMD) calibration and allows for correction of minor changes. However, because clinical trials may last up to 5 years, it is likely that the scanner will need replacing during the course of a study. We report the upgrade of a Hologic QDR-2000 to a QDR-2000plus. The new scanner was cross-calibrated with the old according to the manufacturer's instructions using the Hologic spine phantom. The accuracy of cross-calibration was checked by in vivo scans of patients in clinical trials and also using the European spine phantom (ESP). Daily QC scans with the Hologic phantom over the 6-week period of the in vivo study showed that the new scanner agreed within 0.2% with the old. However, in vivo scans at 10 sites in the spine, hip, total body, and forearm showed significant mismatch between the two systems with five sites differing by more than 2%. Results for the ESP phantom lay closer to the in vivo spine data than the Hologic phantom. The mismatch revealed was larger than anticipated and emphasized the importance of performing adequate in vivo cross-calibration studies whenever DXA systems are replaced. Typically, scans of 20-30 subjects are required at each scan site to achieve an accuracy of 1%. Such studies require careful planning and make significant demands on space, patient cooperation, scanning time, scientific resources, and, not least, the implementation of the findings.

Entities:  

Mesh:

Year:  1996        PMID: 8661987     DOI: 10.1007/s002239900075

Source DB:  PubMed          Journal:  Calcif Tissue Int        ISSN: 0171-967X            Impact factor:   4.333


  14 in total

1.  Longitudinal precision of dual-energy x-ray absorptiometry in a multicenter study. The Nafarelin/Bone Study Group.

Authors:  E S Orwoll; S K Oviatt
Journal:  J Bone Miner Res       Date:  1991-02       Impact factor: 6.741

2.  Calibration and standardization of bone mineral densitometers.

Authors:  T L Kelly; D M Slovik; R M Neer
Journal:  J Bone Miner Res       Date:  1989-10       Impact factor: 6.741

3.  Comparative performance in vitro and in vivo of Lunar DPX and Hologic QDR-1000 dual energy X-ray absorptiometers.

Authors:  M A Laskey; M E Flaxman; R W Barber; S Trafford; M P Hayball; K D Lyttle; A J Crisp; J E Compston
Journal:  Br J Radiol       Date:  1991-11       Impact factor: 3.039

4.  Comparison of dual-energy x-ray absorptiometry and dual photon absorptiometry for bone mineral measurements of the lumbar spine.

Authors:  H W Wahner; W L Dunn; M L Brown; R L Morin; B L Riggs
Journal:  Mayo Clin Proc       Date:  1988-11       Impact factor: 7.616

5.  The European Spine Phantom--a tool for standardization and quality control in spinal bone mineral measurements by DXA and QCT.

Authors:  W A Kalender; D Felsenberg; H K Genant; M Fischer; J Dequeker; J Reeve
Journal:  Eur J Radiol       Date:  1995-07       Impact factor: 3.528

6.  Dual X-ray absorptiometry: a comparison between fan beam and pencil beam scans.

Authors:  G M Blake; J C Parker; F M Buxton; I Fogelman
Journal:  Br J Radiol       Date:  1993-10       Impact factor: 3.039

7.  Switching from DXA pencil-beam to fan-beam. I: Studies in vitro at four centers.

Authors:  P Eiken; O Bärenholdt; L Bjørn Jensen; J Gram; S Pors Nielsen
Journal:  Bone       Date:  1994 Nov-Dec       Impact factor: 4.398

8.  Comparison of four methods for cross-calibrating dual-energy X-ray absorptiometers to eliminate systematic errors when upgrading equipment.

Authors:  J S Finkelstein; J P Butler; R L Cleary; R M Neer
Journal:  J Bone Miner Res       Date:  1994-12       Impact factor: 6.741

9.  Bone density at various sites for prediction of hip fractures. The Study of Osteoporotic Fractures Research Group.

Authors:  S R Cummings; D M Black; M C Nevitt; W Browner; J Cauley; K Ensrud; H K Genant; L Palermo; J Scott; T M Vogt
Journal:  Lancet       Date:  1993-01-09       Impact factor: 79.321

10.  Universal standardization for dual x-ray absorptiometry: patient and phantom cross-calibration results.

Authors:  H K Genant; S Grampp; C C Glüer; K G Faulkner; M Jergas; K Engelke; S Hagiwara; C Van Kuijk
Journal:  J Bone Miner Res       Date:  1994-10       Impact factor: 6.741

View more
  6 in total

1.  The effect of weight and weight change on the long-term precision of spine and hip DXA measurements.

Authors:  R Rajamanohara; J Robinson; J Rymer; R Patel; I Fogelman; G M Blake
Journal:  Osteoporos Int       Date:  2010-08-11       Impact factor: 4.507

2.  Pediatric in vivo cross-calibration between the GE Lunar Prodigy and DPX-L bone densitometers.

Authors:  Nicola J Crabtree; N J Shaw; C M Boivin; B Oldroyd; J G Truscott
Journal:  Osteoporos Int       Date:  2005-10-19       Impact factor: 4.507

3.  Evaluation of the European Spine Phantom in a multi-centre clinical trial.

Authors:  B Lees; S W Garland; C Walton; J C Stevenson
Journal:  Osteoporos Int       Date:  1997       Impact factor: 4.507

4.  Cross-Calibrated Dual-Energy X-Ray Absorptiometry Scanners Demonstrate Systematic Bias in Pediatric and Young Adult Females.

Authors:  Jodi N Dowthwaite; Kristen A Dunsmore; Dongliang Wang; Paula F Rosenbaum; Tamara A Scerpella
Journal:  J Clin Densitom       Date:  2017-02-28       Impact factor: 2.617

5.  Cross-calibration of two dual-energy X-ray absorptiometry devices for the measurement of body composition in young children.

Authors:  Jaz Lyons-Reid; Timothy Kenealy; Benjamin B Albert; Kate A Ward; Nicholas Harvey; Keith M Godfrey; Shiao-Yng Chan; Wayne S Cutfield
Journal:  Sci Rep       Date:  2022-08-16       Impact factor: 4.996

6.  Cross-calibration of pencil-beam (DPX-NT) and fan-beam (QDR-4500C) dual-energy X-ray absorptiometry for sarcopenia.

Authors:  Kenyu Ito; Kazuyo Tsushita; Akiko Muramoto; Hiroki Kanzaki; Takashi Nohara; Hitomi Shimizu; Tomoko Nakazawa; Atsushi Harada
Journal:  Nagoya J Med Sci       Date:  2015-11       Impact factor: 1.131

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.