B Lees1, S W Garland, C Walton, J C Stevenson.
Abstract
The European Spine Phantom (ESP) has recently been developed as a universal standard for instruments measuring bone density. The ESP is composed of three semi-anthropomorphic hydroxyapatite vertebrae of varying densities surrounded by soft tissue equivalent plastic designed to resemble human bone and soft tissue when scanned on bone densitometers. In multi-centre studies it is particularly important to verify that each participating bone densitometer is performing in a stable and linear fashion. The aim of this study was to evaluate the ESP within the context of a multi-centre clinical trial. Eighteen centres in the UK and Canada with dual-energy X-ray absorptiometry (DXA) instruments (Lunar DPX, DPX-alpha and DPX-L) participated in the study. The ESP was scanned 10 times on each instrument without repositioning using standardized protocols. The precision of the bone mineral density (BMD) measurements (LI-3) expressed as a coefficient of variation ranged from 0.4% to 1.1% (mean 0.7%). The mean BMD of each instrument was expressed as a percentage difference from the overall mean and ranged from -1.33% to 1.33%. Linear regression analysis showed that all instruments behaved in a linear fashion across the range of densities with correlation coefficients all > or = 0.999 and standard errors of the estimate < 1.5% of the mean BMD ESP value. The data from this study demonstrate that the ESP is a useful phantom for assessing the linearity, stability and differences between DXA instruments from one manufacturer.Entities:
Mesh:
Year: 1997 PMID: 9604054 DOI: 10.1007/bf02652564
Source DB: PubMed Journal: Osteoporos Int ISSN: 0937-941X Impact factor: 4.507