A comparison of Icare PRO and Perkins tonometers in anesthetized children.

Massimiliano Serafino1, Edoardo Villani1, Andrea Lembo2, Giovanni Rabbiolo1, Claudia Specchia3, Rupal H Trivedi4, Paolo Nucci1.   

Abstract

AIM: To compare intraocular pressure (IOP) measurements obtained with the Perkins applanation tonometer and Icare PRO (ICP) rebound tonometer in anesthetized aphakic or strabismus children. Furthermore, intra-operator correlation and inter-operator correlation have been evaluated, along with the effects of central corneal thickness (CCT) on IOP measurements.
METHODS: Seventy children undergoing examination under anesthesia with sevoflurane for aphakic patients and for surgery for strabismus were included. IOP have been measured twice immediately after anesthesia induction with both Perkins applanation tonometer (PAT) and ICP in one eye and by two different operators with both devices in the fellow eye. Furthermore, CCT was measured with ultrasound pachymetry Pacline (Optikon). Agreement between the device measurements has been evaluated using Bland-Altman analyses. Repeatability and reproducibility of the device have been evaluated with intraclass correlation coefficient (ICC) with a value > 0.75 associated with excellent reliability. The relationship between IOP and CCT has been evaluated with Spearman's correlation coefficient r and determination coefficient r2.
RESULTS: Mean difference in IOP measurements between ICP and PAT was 1.97 mmHg ± 1.23 mmHg (p < 0.05). This difference appeared to be higher in aphakic patients (mean difference 2.15 ± 1.35) than in patients undergoing strabismus surgery (mean difference 1.83 mmHg ± 1.12). Intraclass correlation coefficient (ICC) is used to evaluate repeatability and reproducibility, which are both high for PAT (repeatability 0.96, reproducibility 0.76) compared with ICP (repeatability 0.81, reproducibility 0.70). Correlation coefficient between CCT and IOP is 0.66 for both ICP and PAT.
CONCLUSION: ICP tends to overestimate IOP compared to PAT. Repeatability and reproducibility are both high for PAT as compared to ICP. A significant correlation between IOP and CCT for both instruments has been demonstrated.

Entities:  

Keywords:  Icare; Intraocular pressure; Pediatric glaucoma; Perkins; Tonometer

Mesh:

Year:  2019        PMID: 31313069     DOI: 10.1007/s10792-019-01143-3

Source DB:  PubMed          Journal:  Int Ophthalmol        ISSN: 0165-5701            Impact factor:   2.031


  27 in total

1.  Accuracy of the new ICare rebound tonometer vs. other portable tonometers in healthy eyes.

Authors:  Carlos García-Resúa; José M González-Meijome; Jorge Gilino; Eva Yebra-Pimentel
Journal:  Optom Vis Sci       Date:  2006-02       Impact factor: 1.973

2.  Correlations between corneal biomechanical properties measured with the ocular response analyzer and ICare rebound tonometry.

Authors:  Jorge Manuel Martins Jorge; Jose M González-Méijome; Antonio Queirós; Paulo Fernandes; Manuel A Parafita
Journal:  J Glaucoma       Date:  2008-09       Impact factor: 2.503

3.  Clinical use of a new position-independent rebound tonometer.

Authors:  Kerstin Stephanie Jablonski; André Rosentreter; Stergiani Gaki; Alexandra Lappas; Thomas Stefan Dietlein
Journal:  J Glaucoma       Date:  2013-12       Impact factor: 2.503

4.  Effects of different sleeping postures on intraocular pressure and ocular perfusion pressure in healthy young subjects.

Authors:  Tae-Eun Lee; Chungkwon Yoo; Yong Yeon Kim
Journal:  Ophthalmology       Date:  2013-04-03       Impact factor: 12.079

5.  Evaluation of the ICare rebound tonometer as a home intraocular pressure monitoring device.

Authors:  Sanjay Asrani; Ashmita Chatterjee; David K Wallace; Cecilia Santiago-Turla; Sandra Stinnett
Journal:  J Glaucoma       Date:  2011-02       Impact factor: 2.503

6.  Clinical Evaluation of the New Rebound Tonometers Icare PRO and Icare ONE Compared With the Goldmann Tonometer.

Authors:  Javier Moreno-Montañés; José M Martínez-de-la-Casa; Alfonso L Sabater; Laura Morales-Fernandez; Cristina Sáenz; Julián Garcia-Feijoo
Journal:  J Glaucoma       Date:  2015-09       Impact factor: 2.503

7.  Comparison of ICare tonometer with Goldmann applanation tonometer in glaucoma patients.

Authors:  Paolo Brusini; Maria Letizia Salvetat; Marco Zeppieri; Claudia Tosoni; Lucia Parisi
Journal:  J Glaucoma       Date:  2006-06       Impact factor: 2.503

8.  Changes in central corneal thickness after congenital cataract surgery.

Authors:  Amir Faramarzi; Mohammad Ali Javadi; Mohammad Hossein Jabbarpoor Bonyadi; Mehdi Yaseri
Journal:  J Cataract Refract Surg       Date:  2010-12       Impact factor: 3.351

9.  Factors influencing intraocular pressure, corneal thickness and corneal biomechanics after congenital cataract surgery.

Authors:  Amir Faramarzi; Sepehr Feizi; Ali Maghsoodlou
Journal:  Br J Ophthalmol       Date:  2017-03-28       Impact factor: 4.638

10.  Central corneal thickness and glaucoma in aphakic and pseudophakic children.

Authors:  John W Simon; Martin R O'Malley; Sai B Gandham; Rajat Ghaiy; Jitka Zobal-Ratner; Steven T Simmons
Journal:  J AAPOS       Date:  2005-08       Impact factor: 1.220

View more
  2 in total

1.  Effect of general inhalational anesthesia on intraocular pressure measurements in normal and glaucomatous children.

Authors:  Engy Samy; Yasmine El Sayed; Ahmed Awadein; Maha Gamil
Journal:  Int Ophthalmol       Date:  2021-03-23       Impact factor: 2.031

2.  24-h intraocular pressure patterns measured by Icare PRO rebound in habitual position of open-angle glaucoma eyes.

Authors:  Zhaobin Fang; Xiaolei Wang; Siyu Qiu; Xinghuai Sun; Yuhong Chen; Ming Xiao
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2021-04-29       Impact factor: 3.117

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.