Fetal hemoglobin quantitations using the Hemocue system are rapid and accurate.

S M Berry1, M P Dombrowski, W B Blessed, J A Bichalski, T B Jones, D B Cotton.   

Abstract

OBJECTIVES: To determine whether hemoglobin quantitations using the Hemocue system, a rapid and portable hemoglobin photometer, on fetal blood obtained via funipuncture were accurate compared to the Coulter S-Plus IV. We also examined whether gestational age or extremes in hemoglobin levels significantly affected the accuracy of the Hemocue system.
METHODS: We performed fetal hemoglobin quantitations using both systems on 58 specimens obtained between 18-38 weeks' gestation. Correlation between values by both systems was determined by linear regression analysis. The effects of gestational age and hemoglobin extremes on the accuracy of the Hemocue system were evaluated by stepwise regression.
RESULTS: The mean Hemocue value was 12.0 +/- 2.4 g/dL (range 3.3-16.4); the mean Coulter value was 11.7 +/- 2.3 g/dL (range 3.6-16.2). The regression equation for Hemocue (y) versus Coulter (x) values was y = 0.72 + 0.97x; r = 0.94 (P < .0001). Neither gestational age nor hemoglobin extremes significantly affected the accuracy of the Hemocue system.
CONCLUSION: The Hemocue system is rapid and accurate for fetal hemoglobin quantitation between 18-38 weeks' gestation regardless of extremes in gestational age or hemoglobin levels.

Mesh:

Substances:

Year:  1993        PMID: 7679787

Source DB:  PubMed          Journal:  Obstet Gynecol        ISSN: 0029-7844            Impact factor:   7.661


  5 in total

1.  Evaluation of the HemoCue compared with the Coulter STKS for measurement of neonatal haemoglobin.

Authors:  I J Rechner; A Twigg; A F Davies; S Imong
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2002-05       Impact factor: 5.747

2.  Hemocue, an accurate bedside method of hemoglobin measurement?

Authors:  C E Rippmann; P C Nett; D Popovic; B Seifert; T Pasch; D R Spahn
Journal:  J Clin Monit       Date:  1997-11

3.  Bedside hemoglobin measurements: sensitivity to changes in serum protein and electrolytes.

Authors:  S E McNulty; S J Sharkey; H Schieren
Journal:  J Clin Monit       Date:  1994-11

Review 4.  Methods and analyzers for hemoglobin measurement in clinical laboratories and field settings.

Authors:  Ralph D Whitehead; Zuguo Mei; Carine Mapango; Maria Elena D Jefferds
Journal:  Ann N Y Acad Sci       Date:  2019-06-04       Impact factor: 5.691

5.  Accuracy of Capillary Hemoglobin Measurements for the Detection of Anemia among U.S. Low-Income Toddlers and Pregnant Women.

Authors:  Safia Boghani; Zuguo Mei; Geraldine S Perry; Gary M Brittenham; Mary E Cogswell
Journal:  Nutrients       Date:  2017-03-09       Impact factor: 5.717

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.