An evaluation of the HemoCue for correcting the haemoglobin value of lipaemic samples.

P D Nicholls1.   

Abstract

HemoCue, an instrument for measuring B-haemoglobin, is described as correcting for the turbidity caused by high white cell counts or hyperlipaemia. It was evaluated for its ability to correct haemoglobin measurement for hyperlipaemia against a previously published method and found to perform adequately, but with a small constant positive bias.

Mesh:

Substances:

Year:  1990        PMID: 2402201

Source DB:  PubMed          Journal:  Med Lab Sci        ISSN: 0308-3616


  3 in total

1.  Bedside hemoglobin measurements: sensitivity to changes in serum protein and electrolytes.

Authors:  S E McNulty; S J Sharkey; H Schieren
Journal:  J Clin Monit       Date:  1994-11

Review 2.  Methods and analyzers for hemoglobin measurement in clinical laboratories and field settings.

Authors:  Ralph D Whitehead; Zuguo Mei; Carine Mapango; Maria Elena D Jefferds
Journal:  Ann N Y Acad Sci       Date:  2019-06-04       Impact factor: 5.691

3.  Perioperative on-site haemoglobin determination: as accurate as laboratory values?

Authors:  M Jaeger; T Ashbury; M Adams; P Duncan
Journal:  Can J Anaesth       Date:  1996-08       Impact factor: 5.063