The effect of pulsating arteries on reflectance pulse oximetry: measurements in adults and neonates.

R Nijland1, H W Jongsma, P P van den Berg, J G Nijhuis, B Oeseburg.   

Abstract

OBJECTIVE: The objective of our study was to describe the results from human experiments during normoxia that demonstrate the effect of pulsating arteries on the measured arterial oxygen saturation (SpO2) using a reflectance pulse oximeter sensor.
METHODS: In 6 healthy adults and 7 healthy neonates, a Nellcor reflection sensor (FS-10 oxisensor, Nellcor, Inc., Pleasanton, CA) was placed in three different positions: (1) on the forehead, (2) on the temporal area, with the photodiode placed over the superficial temporal artery, and (3) on the temporal area, with the light-emitting diodes (LEDs) placed over the superficial temporal artery.
RESULTS: Placement of the sensor in position 2 resulted in a significantly lower SpO2 reading, compared to sensor position 1: 5.8% (p < 0.01) lower for adults and 7.5% (p < 0.01) lower for neonates. Placement of the sensor in position 3 resulted in significantly larger plethysmographic signals, compared to sensor position 1; but, the Spo2 readings were alike.
CONCLUSIONS: Pulsating arteries can affect the reliability of reflection pulse oximetry. Depending on the position of the sensor, a falsely low Spo2 value can be observed.

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 7760084     DOI: 10.1007/BF01617733

Source DB:  PubMed          Journal:  J Clin Monit        ISSN: 0748-1977


  12 in total

1.  Inappropriate sensor application in pulse oximetry.

Authors:  J O Gardosi; D Damianou; C M Schram
Journal:  Lancet       Date:  1992-10-10       Impact factor: 79.321

2.  In vitro investigation of the factors affecting pulse oximetry.

Authors:  J P de Kock; L Tarassenko
Journal:  J Biomed Eng       Date:  1991-01

Review 3.  Pulse oximetry.

Authors:  K K Tremper; S J Barker
Journal:  Anesthesiology       Date:  1989-01       Impact factor: 7.892

Review 4.  Pulse oximetry: analysis of theory, technology, and practice.

Authors:  M W Wukitsch; M T Petterson; D R Tobler; J A Pologe
Journal:  J Clin Monit       Date:  1988-10

5.  A comparison of a new reflectance oximeter with the Hewlett-Packard ear oximeter.

Authors:  M J Decker; D Dickensheets; J L Arnold; P W Cheung; K P Strohl
Journal:  Biomed Instrum Technol       Date:  1990 Mar-Apr

6.  Forehead pulse oximetry compared with finger pulse oximetry and arterial blood gas measurement.

Authors:  E Y Cheng; M B Hopwood; J Kay
Journal:  J Clin Monit       Date:  1988-07

7.  Noninvasive pulse oximetry utilizing skin reflectance photoplethysmography.

Authors:  Y Mendelson; B D Ochs
Journal:  IEEE Trans Biomed Eng       Date:  1988-10       Impact factor: 4.538

8.  Near-infrared monitoring of cerebral oxygen sufficiency. I. Spectra of cytochrome c oxidase.

Authors:  F F Jöbsis-VanderVliet; C A Piantadosi; A L Sylvia; S K Lucas; H H Keizer
Journal:  Neurol Res       Date:  1988-03       Impact factor: 2.448

9.  Errors in 14 pulse oximeters during profound hypoxia.

Authors:  J W Severinghaus; K H Naifeh; S O Koh
Journal:  J Clin Monit       Date:  1989-04

10.  Preliminary experience with intrapartum fetal pulse oximetry in humans.

Authors:  G A Dildy; S L Clark; C A Loucks
Journal:  Obstet Gynecol       Date:  1993-04       Impact factor: 7.661

View more
  4 in total

Review 1.  Factors influencing fetal pulse oximetry performance.

Authors:  James L Reuss
Journal:  J Clin Monit Comput       Date:  2004-02       Impact factor: 2.502

2.  The influence of larger subcutaneous blood vessels on pulse oximetry.

Authors:  Paul D Mannheimer; Neil Michael P O'; Ewald Konecny
Journal:  J Clin Monit Comput       Date:  2004-06       Impact factor: 2.502

3.  The pulse in reflectance pulse oximetry: modeling and experimental studies.

Authors:  James L Reuss; Daniel Siker
Journal:  J Clin Monit Comput       Date:  2004-08       Impact factor: 2.502

4.  Reflectance pulse oximetry in fetal lambs: subcutaneous vessels and vasoconstriction affect its reliability.

Authors:  R Nijland; H W Jongsma; I M Verbruggen; J G Nijhuis
Journal:  J Clin Monit       Date:  1996-05
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.