The influence of larger subcutaneous blood vessels on pulse oximetry.

Paul D Mannheimer1, Neil Michael P O', Ewald Konecny.   

Abstract

OBJECTIVE: Recent studies have renewed interest in reflectance pulse oximetry, specifically for monitoring the patient's forehead. Blood circulation on the forehead immediately above the eyebrow is fed by arteries that branch from the internal carotid artery and lack the vasoconstrictor response present in more peripheral regions. Some investigators question, however, the reliability of monitoring SpO2 on the forehead due to prior reported inaccurate readings with reflectance sensors. The present study evaluates pulse oximetry accuracy when reflectance sensors are placed over potentially pulsing or moving larger arterial vessels, or over more homogeneous microvasculature devoid of larger subcutaneous vessels. METHODS. Ten healthy adult volunteers were fitted with reflectance pulse oximetry sensors and exposed to a controlled desaturation to 70%. Sensors were placed immediately above the left and right eyebrows as well as over the temple. Additionally, numerical modeling was used to simulate light signals and photon migration through a homogeneous tissue bed with an added static or dynamic artery.
RESULTS: Sensors placed above the eyebrows tracked one another with significantly better accuracy than when comparing temple with the brow placement (RMS of the Differences = 1.12% vs. 4.24%, respectively). Photon migration simulations indicate that the detected light bypasses the interior of larger vessels, while vessel presence affects the red and IR light pulse amplitudes independent of SaO2.
CONCLUSIONS: Placement of reflectance pulse oximetry sensors directly over larger cardio-synchronously pulsing or moving vasculature can significantly degrade SpO2 reading accuracy. Reflectance sensors placed low on the forehead directly over the eyebrow and slightly lateral to the iris appear to avoid such vasculature and provide consistent and accurate estimates of SaO2.

Entities:  

Mesh:

Year:  2004        PMID: 15562984     DOI: 10.1023/b:jocm.0000042924.16052.5e

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  13 in total

1.  Reflection and transmission pulse oximetry during compromised peripheral perfusion.

Authors:  H Pälve
Journal:  J Clin Monit       Date:  1992-01

2.  Effect of location of the sensor on reflectance pulse oximetry.

Authors:  A C Dassel; R Graaff; M Aardema; W G Zijlstra; J G Aarnoudse
Journal:  Br J Obstet Gynaecol       Date:  1997-08

3.  Absorption spectra of human fetal and adult oxyhemoglobin, de-oxyhemoglobin, carboxyhemoglobin, and methemoglobin.

Authors:  W G Zijlstra; A Buursma; W P Meeuwsen-van der Roest
Journal:  Clin Chem       Date:  1991-09       Impact factor: 8.327

Review 4.  Physio-optical considerations in the design of fetal pulse oximetry sensors.

Authors:  P D Mannheimer; M E Fein; J R Casciani
Journal:  Eur J Obstet Gynecol Reprod Biol       Date:  1997-03       Impact factor: 2.435

5.  Model for photon migration in turbid biological media.

Authors:  R F Bonner; R Nossal; S Havlin; G H Weiss
Journal:  J Opt Soc Am A       Date:  1987-03       Impact factor: 2.129

6.  Pulse oximeter failure thresholds in hypotension and vasoconstriction.

Authors:  J W Severinghaus; M J Spellman
Journal:  Anesthesiology       Date:  1990-09       Impact factor: 7.892

7.  The effect of pulsating arteries on reflectance pulse oximetry: measurements in adults and neonates.

Authors:  R Nijland; H W Jongsma; P P van den Berg; J G Nijhuis; B Oeseburg
Journal:  J Clin Monit       Date:  1995-03

8.  Design and validation of pulse oximetry for low saturation.

Authors:  Paul D Mannheimer
Journal:  Anesth Analg       Date:  2002-01       Impact factor: 5.108

9.  Influence of sensor site location on pulse oximetry kinetics in children.

Authors:  L M Reynolds; S C Nicolson; J M Steven; A Escobar; M E McGonigle; D R Jobes
Journal:  Anesth Analg       Date:  1993-04       Impact factor: 5.108

10.  Accuracy of response of six pulse oximeters to profound hypoxia.

Authors:  J W Severinghaus; K H Naifeh
Journal:  Anesthesiology       Date:  1987-10       Impact factor: 7.892

View more
  1 in total

1.  Pulse oximeter in a comparative test.

Authors:  Hartmut Gehring
Journal:  Intensive Care Med       Date:  2006-06-30       Impact factor: 17.440

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.