Reflectance pulse oximetry in fetal lambs: subcutaneous vessels and vasoconstriction affect its reliability.

R Nijland1, H W Jongsma, I M Verbruggen, J G Nijhuis.   

Abstract

OBJECTIVE: Reflectance pulse oximetry (RPOX) has been introduced for intrapartum fetal surveillance. The purpose of this study was to describe two possible effects on the reliability of RPOX, namely the effect of the presence of a subcutaneous vein and the effect of vasoconstriction by adrenaline, both at fetal SaO2 levels.
METHODS: In four anesthetized fetal lambs, a prototype 660/890 nm reflectance sensor (Nellcor Inc.) was placed on the fetal head, with the photodiode of the sensor precisely over a superficial subcutaneous vein. Measurements were made before and after coagulation of the vein. In five anesthetized fetal lambs, one or two reflectance sensors were placed on the fetal head and/or neck and adrenaline was administered in doses of 0.02 to 0.04 mg via a brachial artery. Pulse oximeter saturation readings (SpO2) were compared with continuous arterial oxygen saturation (SaO2) values obtained using a fiberoptic catheter (Opticath, Abbott) in the carotid artery.
RESULTS: When the sensor was placed over the vein, the pulse oximeter read 18% to 24% too low at a SaO2 level of 20% to 50%. After coagulation of the vein, SpO2 readings were in agreement with fiberoptic SaO2 values. Administration of adrenaline resulted in a large overestimation of the SaO2 in 6 of the 7 measurements.
CONCLUSIONS: Subcutaneous veins and vasoconstriction can affect the reliability of reflectance pulse oximetry. As comparable situations may occur during labor, SpO2 readings should be interpreted with caution when this kind or comparable types of RPOX sensors are used at low SaO2 levels.

Entities:  

Mesh:

Substances:

Year:  1996        PMID: 8823646     DOI: 10.1007/bf00857643

Source DB:  PubMed          Journal:  J Clin Monit        ISSN: 0748-1977


  19 in total

1.  Reflectance pulse oximetry in fetal lambs.

Authors:  A C Dassel; R Graaff; J G Aarnoudse; J M Elstrodt; P Heida; M H Koelink; F F de Mul; J Greve
Journal:  Pediatr Res       Date:  1992-03       Impact factor: 3.756

Review 2.  Recent developments in pulse oximetry.

Authors:  J W Severinghaus; J F Kelleher
Journal:  Anesthesiology       Date:  1992-06       Impact factor: 7.892

3.  Adaptation of pulse oximetry for fetal monitoring during labour.

Authors:  J O Gardosi; C M Schram; E M Symonds
Journal:  Lancet       Date:  1991-05-25       Impact factor: 79.321

4.  The effect of caput succedaneum on oxygen saturation measurements.

Authors:  N Johnson; V A Johnson; J Bannister; R J Lilford
Journal:  Br J Obstet Gynaecol       Date:  1990-06

Review 5.  Pulse oximetry: analysis of theory, technology, and practice.

Authors:  M W Wukitsch; M T Petterson; D R Tobler; J A Pologe
Journal:  J Clin Monit       Date:  1988-10

6.  The effect of pulsating arteries on reflectance pulse oximetry: measurements in adults and neonates.

Authors:  R Nijland; H W Jongsma; P P van den Berg; J G Nijhuis; B Oeseburg
Journal:  J Clin Monit       Date:  1995-03

7.  Validation of arterial oxygen saturation measurements in utero using pulse oximetry.

Authors:  A P Harris; M J Sendak; D C Chung; C A Richardson
Journal:  Am J Perinatol       Date:  1993-05       Impact factor: 1.862

8.  Measurement of oxygen saturation by multiwavelength analyzer influenced by interspecies differences.

Authors:  R Nijland; B Ringnalda; H W Jongsma; B Oeseburg; W G Zijlstra
Journal:  Clin Chem       Date:  1994-10       Impact factor: 8.327

9.  Intrapartum fetal pulse oximetry: fetal oxygen saturation trends during labor and relation to delivery outcome.

Authors:  G A Dildy; P P van den Berg; M Katz; S L Clark; H W Jongsma; J G Nijhuis; C A Loucks
Journal:  Am J Obstet Gynecol       Date:  1994-09       Impact factor: 8.661

10.  Artifacts in fetal pulse oximetry: nonarterial pulsatile signals.

Authors:  C M Schram; J O Gardosi
Journal:  Am J Obstet Gynecol       Date:  1994-04       Impact factor: 8.661

View more
  4 in total

Review 1.  Factors influencing fetal pulse oximetry performance.

Authors:  James L Reuss
Journal:  J Clin Monit Comput       Date:  2004-02       Impact factor: 2.502

2.  The pulse in reflectance pulse oximetry: modeling and experimental studies.

Authors:  James L Reuss; Daniel Siker
Journal:  J Clin Monit Comput       Date:  2004-08       Impact factor: 2.502

3.  Validation of reflectance pulse oximetry: an evaluation of a new sensor in piglets.

Authors:  R Nijland; S Nierlich; H W Jongsma; J G Nijhuis; B Oeseburg; K Springer; P Mannheimer
Journal:  J Clin Monit       Date:  1997-01

Review 4.  Sources of Inaccuracy in Photoplethysmography for Continuous Cardiovascular Monitoring.

Authors:  Jesse Fine; Kimberly L Branan; Andres J Rodriguez; Tananant Boonya-Ananta; Jessica C Ramella-Roman; Michael J McShane; Gerard L Coté
Journal:  Biosensors (Basel)       Date:  2021-04-16