Anatomically and physiologically based reference level for measurement of intracardiac pressures.

M Courtois1, P G Fattal, S J Kovács, A J Tiefenbrunn, P A Ludbrook.   

Abstract

BACKGROUND: Many reference levels have been proposed for the measurement of intracardiac pressures, but none have met with universal acceptance. In the first part of our study, we evaluated 10 cardiologists' understanding of how hydrostatic pressure influences intracardiac pressures as measured with fluid-filled catheters. In the second part, we proposed and validated a new zero level (H): the uppermost blood level in the left ventricular (LV) chamber relative to the anterior chest wall for a patient in the supine position. A comparison was made of LV minimum diastolic pressure measured by reference to H versus measurements made with the zero level at midchest. METHODS AND
RESULTS: Using two-dimensional echocardiography, we determined H in the LVs of seven normal patients (five male, two female; age, 49 +/- 9 years) undergoing routine cardiac catheterization. H was determined from a left parasternal short-axis view and calculated as the average distance between end diastole and end systole of the endocardium of the uppermost segment of the LV anterior wall below the fourth or fifth intercostal space of the left sternal border on the anterior surface of the chest wall, with the patient in the supine position. A micromanometer/fluid-filled lumen catheter was then positioned in the LV, and we compared the micromanometer LV minimum pressure (LVPmin) obtained when the reference fluid-filled transducer was aligned at midchest with the LVPmin obtained when the reference fluid-filled transducer was aligned at H. LVPmin referenced to a midchest fluid-filled external transducer was measured as 5.1 +/- 1.6 mm Hg (range, 2.4 to 7.2 mm Hg) versus -0.6 +/- 0.6 mm Hg (range, -1.6 to 0.4 mm Hg) when referenced to H (P < .001). A significant linear relation was found to exist between patient anterior-posterior chest diameter and the magnitude of hydrostatic pressure influences related to pressure referenced at midchest (r = .88; P < .01).
CONCLUSIONS: External fluid-filled transducers should be used with the goal of removing hydrostatic pressure and other influences so that the presence of subatmospheric pressure during diastole in any of the cardiac chambers is accurately measured. To achieve this goal, intracardiac pressure should be referenced to an external fluid-filled transducer aligned with the uppermost blood level in the chamber in which pressure is to be measured. The current practice of referencing the zero level of LV diastolic pressure to an external fluid-filled transducer positioned at the midchest level results in systematic overestimation due to hydrostatic effects and produces physiologically significant error in the measurement of diastolic intracardiac pressure.

Entities:  

Mesh:

Year:  1995        PMID: 7671382     DOI: 10.1161/01.cir.92.7.1994

Source DB:  PubMed          Journal:  Circulation        ISSN: 0009-7322            Impact factor:   29.690


  14 in total

1.  Evaluation of haemodynamics by cardiac catheterisation: historical perspective and present practice.

Authors:  A Bergstra; A F M van den Heuvel; G A Mook; F Zijlstra
Journal:  Neth Heart J       Date:  2005-04       Impact factor: 2.380

2.  Impact of two formulas to calculate percentage diameter stenosis of coronary lesions: from stenosis models (phantom lesion model) to actual clinical lesions.

Authors:  Alexandre Hideo-Kajita; Samuel Wopperer; Solomon S Beyene; Yael F Meirovich; Gebremedhin D Melaku; Kayode O Kuku; Echo J Brathwaite; Yuichi Ozaki; Kazuhiro Dan; Rebecca Torguson; Ron Waksman; Hector M Garcia-Garcia
Journal:  Int J Cardiovasc Imaging       Date:  2019-07-27       Impact factor: 2.357

3.  The effect of variable arterial transducer level on the accuracy of pulse contour waveform-derived measurements in critically ill patients.

Authors:  Huai-Wu He; Da-Wei Liu; Yun Long; Xiao-Ting Wang; Mei-Ling Zhao; Xiao-Li Lai
Journal:  J Clin Monit Comput       Date:  2015-08-23       Impact factor: 2.502

4.  Optimal level of the reference transducer for central venous pressure and pulmonary artery occlusion pressure monitoring in supine, prone, and sitting position.

Authors:  In-Kyung Song; Han-Seul Park; Ji-Hyun Lee; Eun-Hee Kim; Hee-Soo Kim; Jae-Hyon Bahk; Jin-Tae Kim
Journal:  J Clin Monit Comput       Date:  2016-03-21       Impact factor: 2.502

5.  In vivo accuracy of two intraparenchymal intracranial pressure monitors.

Authors:  Thomas Lescot; Vincent Reina; Yannick Le Manach; Filippo Boroli; Dorian Chauvet; Anne-Laure Boch; Louis Puybasset
Journal:  Intensive Care Med       Date:  2011-02-26       Impact factor: 17.440

6.  Evaluation of external reference levels for central venous pressure measurements of severely obese patients in the supine position.

Authors:  Kazuyuki Mizunoya; Hitoshi Saito; Yuji Morimoto
Journal:  J Anesth       Date:  2018-05-28       Impact factor: 2.078

7.  Optimal Transducer Level for Atrial and Pulmonary Arterial Pressure Measurement in Patients with Functional Single Ventricle.

Authors:  Yong-Hee Park; Da-Hye Yoo; Eun-Hee Kim; In-Kyung Song; Ji-Hyun Lee; Hee-Soo Kim; Woong-Han Kim; Jin-Tae Kim
Journal:  Pediatr Cardiol       Date:  2016-10-01       Impact factor: 1.655

8.  Impaired adaptation to left atrial pressure increase in patients with atrial fibrillation.

Authors:  Gergely Ágoston; Judit Szilágyi; Gábor Bencsik; Cristina Tutuianu; Gergely Klausz; László Sághy; Albert Varga; Tamás Forster; Róbert Pap
Journal:  J Interv Card Electrophysiol       Date:  2015-06-30       Impact factor: 1.900

9.  Sympathetic predominance before tourniquet deflation is associated with a reduction in arterial blood pressure after tourniquet deflation during total knee arthroplasty.

Authors:  E Kim; M R Cho; S H Byun; J A Lim; S Chae; W K Choi; I Kim; J Kim
Journal:  Physiol Res       Date:  2021-05-12       Impact factor: 1.881

10.  Echocardiographic estimation of mean pulmonary artery pressure in critically ill patients.

Authors:  Russell D Laver; Ubbo F Wiersema; Andrew D Bersten
Journal:  Crit Ultrasound J       Date:  2014-07-02
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.