In vivo accuracy of two intraparenchymal intracranial pressure monitors.

Thomas Lescot1, Vincent Reina, Yannick Le Manach, Filippo Boroli, Dorian Chauvet, Anne-Laure Boch, Louis Puybasset.   

Abstract

PURPOSE: To evaluate the in vivo accuracy of the new Pressio(®) device for intraparenchymal monitoring of intracranial pressure (ICP) versus the Codman(®) device and intraventricular measurement external ventricular drainage (EVD).
METHODS: Data were collected retrospectively for 30 consecutive patients admitted into a 25-bed neurosurgical intensive care unit of a university hospital between January and December 2009. Patients received both intraventricular and intraparenchymal ICP monitoring with Pressio(®) (n = 15) or Codman(®) (n = 15).
RESULTS: We obtained 3,089 data points from the 30 patients. Mean difference between intraparenchymal and EVD pressure (bias) was -0.6 mmHg, and limits of agreement (1.96 SD of the bias) were -8.1 to 6.9 mmHg with Pressio(®) and 0.3 mmHg with limits of agreement of -6.7 to 7.1 mmHg with Codman(®) (NS). The temporal difference was -0.7 ± 1.6 mmHg/100 h of monitoring with Pressio(®) and 0.1 ± 1.6 mmHg/100 h of monitoring with Codman(®) over the study period (NS).
CONCLUSIONS: Intraparenchymal pressure measured with both transducers approximates intraventricular cerebrospinal fluid pressure with an accuracy of ±7 mmHg.

Entities:  

Mesh:

Year:  2011        PMID: 21359608     DOI: 10.1007/s00134-011-2182-8

Source DB:  PubMed          Journal:  Intensive Care Med        ISSN: 0342-4642            Impact factor:   17.440


  17 in total

1.  Refractory intracranial hypertension and "second-tier" therapies in traumatic brain injury.

Authors:  Nino Stocchetti; Clelia Zanaboni; Angelo Colombo; Giuseppe Citerio; Luigi Beretta; Laura Ghisoni; Elisa Roncati Zanier; Katia Canavesi
Journal:  Intensive Care Med       Date:  2007-12-08       Impact factor: 17.440

2.  Guidelines for the management of severe traumatic brain injury. VIII. Intracranial pressure thresholds.

Authors:  Susan L Bratton; Randall M Chestnut; Jamshid Ghajar; Flora F McConnell Hammond; Odette A Harris; Roger Hartl; Geoffrey T Manley; Andrew Nemecek; David W Newell; Guy Rosenthal; Joost Schouten; Lori Shutter; Shelly D Timmons; Jamie S Ullman; Walter Videtta; Jack E Wilberger; David W Wright
Journal:  J Neurotrauma       Date:  2007       Impact factor: 5.269

3.  Guidelines for the management of severe traumatic brain injury. VII. Intracranial pressure monitoring technology.

Authors:  Susan L Bratton; Randall M Chestnut; Jamshid Ghajar; Flora F McConnell Hammond; Odette A Harris; Roger Hartl; Geoffrey T Manley; Andrew Nemecek; David W Newell; Guy Rosenthal; Joost Schouten; Lori Shutter; Shelly D Timmons; Jamie S Ullman; Walter Videtta; Jack E Wilberger; David W Wright
Journal:  J Neurotrauma       Date:  2007       Impact factor: 5.269

4.  Assessment of zero drift in the Codman intracranial pressure monitor: a study from 2 neurointensive care units.

Authors:  Yahia Z Al-Tamimi; Adel Helmy; Seb Bavetta; Stephen J Price
Journal:  Neurosurgery       Date:  2009-01       Impact factor: 4.654

5.  A quantitative computed tomography assessment of brain weight, volume, and specific gravity in severe head trauma.

Authors:  Thomas Lescot; Marie-Pierre Bonnet; Abederrezak Zouaoui; Jean-Charles Muller; Catalin Fetita; Pierre Coriat; Louis Puybasset
Journal:  Intensive Care Med       Date:  2005-07-01       Impact factor: 17.440

6.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

7.  Clinical experience with the intraparenchymal intracranial pressure monitoring Codman MicroSensor system.

Authors:  Lars-Owe D Koskinen; Magnus Olivecrona
Journal:  Neurosurgery       Date:  2005-04       Impact factor: 4.654

8.  Prevention of external ventricular drain--related ventriculitis.

Authors:  A-M Korinek; M Reina; A L Boch; A O Rivera; D De Bels; L Puybasset
Journal:  Acta Neurochir (Wien)       Date:  2005-01       Impact factor: 2.216

9.  Clinical evaluation of a miniature strain-gauge transducer for monitoring intracranial pressure.

Authors:  S P Gopinath; C S Robertson; C F Contant; R K Narayan; R G Grossman
Journal:  Neurosurgery       Date:  1995-06       Impact factor: 4.654

10.  Continuous versus intermittent cerebrospinal fluid drainage after severe traumatic brain injury in children: effect on biochemical markers.

Authors:  Paul M Shore; Neal J Thomas; Robert S B Clark; P David Adelson; Stephen R Wisniewski; Keri L Janesko; Hülya Bayir; Edwin K Jackson; Patrick M Kochanek
Journal:  J Neurotrauma       Date:  2004-09       Impact factor: 5.269

View more
  14 in total

1.  Differentiate the Source and Site of Intracranial Pressure Measurements Using More Precise Nomenclature.

Authors:  DaiWai M Olson; Stefany Ortega Peréz; Jonathan Ramsay; Chethan P Venkatasubba Rao; Jose I Suarez; Molly McNett; Venkatesh Aiyagari
Journal:  Neurocrit Care       Date:  2019-04       Impact factor: 3.210

Review 2.  Brain Multimodality Monitoring: Updated Perspectives.

Authors:  David Roh; Soojin Park
Journal:  Curr Neurol Neurosci Rep       Date:  2016-06       Impact factor: 5.081

Review 3.  Intracranial pressure monitoring: fundamental considerations and rationale for monitoring.

Authors:  Randall Chesnut; Walter Videtta; Paul Vespa; Peter Le Roux
Journal:  Neurocrit Care       Date:  2014-12       Impact factor: 3.210

4.  Year in review in Intensive Care Medicine 2011: I. Nephrology, epidemiology, nutrition and therapeutics, neurology, ethical and legal issues, experimentals.

Authors:  Massimo Antonelli; Marc Bonten; Jean Chastre; Giuseppe Citerio; Giorgio Conti; J Randall Curtis; Daniel De Backer; Goran Hedenstierna; Michael Joannidis; Duncan Macrae; Jordi Mancebo; Salvatore M Maggiore; Alexandre Mebazaa; Jean-Charles Preiser; Patricia Rocco; Jean-François Timsit; Jan Wernerman; Haibo Zhang
Journal:  Intensive Care Med       Date:  2012-01-04       Impact factor: 17.440

5.  Intracranial Pressure Monitoring: Invasive versus Non-Invasive Methods-A Review.

Authors:  P H Raboel; J Bartek; M Andresen; B M Bellander; B Romner
Journal:  Crit Care Res Pract       Date:  2012-06-08

Review 6.  Optical Fibre Pressure Sensors in Medical Applications.

Authors:  Sven Poeggel; Daniele Tosi; DineshBabu Duraibabu; Gabriel Leen; Deirdre McGrath; Elfed Lewis
Journal:  Sensors (Basel)       Date:  2015-07-15       Impact factor: 3.576

7.  Comparison of parenchymal and ventricular intracranial pressure readings utilizing a novel multi-parameter intracranial access system.

Authors:  Tracey Berlin; Cristina Murray-Krezan; Howard Yonas
Journal:  Springerplus       Date:  2015-01-06

Review 8.  Accuracy of intracranial pressure monitoring: systematic review and meta-analysis.

Authors:  Lucia Zacchetti; Sandra Magnoni; Federica Di Corte; Elisa R Zanier; Nino Stocchetti
Journal:  Crit Care       Date:  2015-12-02       Impact factor: 9.097

9.  Intracranial pressure monitoring in normal dogs using subdural and intraparenchymal miniature strain-gauge transducers.

Authors:  Beverly K Sturges; Peter J Dickinson; Linda D Tripp; Irina Udaltsova; Richard A LeCouteur
Journal:  J Vet Intern Med       Date:  2018-12-21       Impact factor: 3.333

10.  Intracranial and Cerebral Perfusion Pressure Thresholds Associated With Inhospital Mortality Across Pediatric Neurocritical Care.

Authors:  Kendra S Woods; Christopher M Horvat; Sajel Kantawala; Dennis W Simon; Jaskaran Rakkar; Patrick M Kochanek; Robert S B Clark; Alicia K Au
Journal:  Pediatr Crit Care Med       Date:  2021-02-01       Impact factor: 3.971

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.