Optimal Transducer Level for Atrial and Pulmonary Arterial Pressure Measurement in Patients with Functional Single Ventricle.

Yong-Hee Park1, Da-Hye Yoo2, Eun-Hee Kim2, In-Kyung Song2, Ji-Hyun Lee2, Hee-Soo Kim2, Woong-Han Kim3, Jin-Tae Kim4.   

Abstract

This study aimed to investigate the optimal transducer level for accurate measurement of atrial and pulmonary arterial pressures in the supine position for patients with functional single ventricle. Contrast-enhanced chest computed tomographic images of 108 patients who underwent either the bidirectional cavopulmonary shunt (BCPS) placement or the Fontan procedure were reviewed. Vertical distances from the skin of the back to the uppermost levels of fluid in the single atrium or the pulmonary artery confluence and their ratios to the greatest anteroposterior (AP) diameter of the thoracic cage were determined. In patients who underwent BCPS, the ratios of the uppermost levels of blood in the atrium and pulmonary artery confluence to the greatest AP diameter of the thorax were 76.0 ± 8.1 and 56.3 ± 5.5 %, respectively. The distance (mm) between these two levels was calculated as 24.2 + 0.31 × age (years) (r 2 = 0.08, P < 0.03). In patients who underwent the Fontan procedure, the ratios were 79.3 ± 10.0 and 58.3 ± 5.8 %, respectively. The distance (mm) between these two levels was calculated as 31.1 + 0.44 × age (years) (r 2 = 0.05, P < 0.11). The optimal transducer levels for measuring atrial and pulmonary arterial pressures in the supine position are 75-80 and 55-60 % of the AP diameter of the thorax, respectively, in patients with functional single ventricle. We should consider the difference of the pressure when atrial and pulmonary arterial pressures were measured with the same level of transducers.

Entities:  

Keywords:  Bidirectional cavopulmonary shunt; Central venous pressure; Congenital heart disease; Fontan procedure; Transducers, Pressure

Mesh:

Substances:

Year:  2016        PMID: 27696307     DOI: 10.1007/s00246-016-1481-9

Source DB:  PubMed          Journal:  Pediatr Cardiol        ISSN: 0172-0643            Impact factor:   1.655


  15 in total

1.  Outcomes after bidirectional cavopulmonary shunt in infants less than 6 months old.

Authors:  V M Reddy; D B McElhinney; P Moore; G S Haas; F L Hanley
Journal:  J Am Coll Cardiol       Date:  1997-05       Impact factor: 24.094

2.  [Effect of posture and positive end expiratory pressure on central venous pressure in patients with mechanical ventilation].

Authors:  Bo Wang; Yan Kang; Xiao-dong Jin; Zhi-cheng Qian; Liang Dong
Journal:  Zhongguo Wei Zhong Bing Ji Jiu Yi Xue       Date:  2007-02

3.  Uppermost blood levels of the right and left atria in the supine position: implication for measuring central venous pressure and pulmonary artery wedge pressure.

Authors:  Jeong-Hwa Seo; Chul-Woo Jung; Jae-Hyon Bahk
Journal:  Anesthesiology       Date:  2007-08       Impact factor: 7.892

4.  Factors associated with prolonged recovery after the fontan operation.

Authors:  Joshua W Salvin; Mark A Scheurer; Peter C Laussen; John E Mayer; Pedro J Del Nido; Frank A Pigula; Emile A Bacha; Ravi R Thiagarajan
Journal:  Circulation       Date:  2008-09-30       Impact factor: 29.690

Review 5.  ACC expert consensus document. Present use of bedside right heart catheterization in patients with cardiac disease. American College of Cardiology.

Authors:  H S Mueller; K Chatterjee; K B Davis; M A Fifer; C Franklin; M A Greenberg; A J Labovitz; P K Shah; K J Tuman; M H Weil; W S Weintraub
Journal:  J Am Coll Cardiol       Date:  1998-09       Impact factor: 24.094

6.  Monitoring pulmonary artery pressure in children undergoing the Fontan procedure.

Authors:  Sumit Vasdev; Sandeep Chauhan; Sachin Talwar; Anil Pandey; Usha Kiran
Journal:  Asian Cardiovasc Thorac Ann       Date:  2013-06

7.  Optimal level of the reference transducer for central venous pressure and pulmonary artery occlusion pressure monitoring in supine, prone, and sitting position.

Authors:  In-Kyung Song; Han-Seul Park; Ji-Hyun Lee; Eun-Hee Kim; Hee-Soo Kim; Jae-Hyon Bahk; Jin-Tae Kim
Journal:  J Clin Monit Comput       Date:  2016-03-21       Impact factor: 2.502

8.  Echocardiographic determination of valid zero reference levels in supine and lateral positions.

Authors:  L L Kee; J S Simonson; N A Stotts; P Skov; N B Schiller
Journal:  Am J Crit Care       Date:  1993-01       Impact factor: 2.228

9.  Comparison of situs ambiguous patterns between heterotaxy syndromes with polysplenia and asplenia.

Authors:  Hiroko Yoneyama; Chisato Kondo; Aki Yamasaki; Toshio Nakanishi; Shuji Sakai
Journal:  Eur J Radiol       Date:  2015-08-05       Impact factor: 3.528

10.  The Fontan procedure: contemporary techniques have improved long-term outcomes.

Authors:  Yves d'Udekem; Ajay J Iyengar; Andrew D Cochrane; Leeanne E Grigg; James M Ramsay; Gavin R Wheaton; Dan J Penny; Christian P Brizard
Journal:  Circulation       Date:  2007-09-11       Impact factor: 29.690

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.