Literature DB >> 7595690

Simultaneous in vivo comparison of two-versus three-wavelength mixed venous (Svo2) oximetry catheters.

F Bongard1, T S Lee, T Leighton, S Y Liu.   

Abstract

OBJECTIVE: Venous oximetry catheters provide useful realtime information about mixed venous hemoglobin saturation (Svo2). Currently available systems utilize either two or three wavelengths of light to obtain these measurements. Previous animal and clinical studies have attempted to compare the accuracy of these two devices under similar circumstances. However, the relative accuracy of the two-wavelength versus three-wavelength systems has never been assessed under identical conditions. For this purpose, we designed an animal model for simultaneous measurement of Svo2, over a wide range of physiologic and pathologic states.
METHODS: Seven anesthetized swine underwent simultaneous placement of two- and three-wavelength catheters. Paired data points consisted of values obtained from a reference oximeter and from each of the catheters. Observations were obtained every 15 min during the following manipulations: (1) eucarbic hypoxia induced by reducing FiO2 to 0.18, 0.15, and 0.12 for 15 min each; (2) stimulated surgical manipulation; and (3) hypovolemic shock produced by hemorrhage to a mean arterial pressure of 50 torr for 1 hr. Data were analyzed by calculation of mean error (bias) and precision for each system in comparison with the oximeter.
RESULTS: The overall error of the two-wavelength system was +0.15%, with a precision of +/- 2.52%. The three-wavelength system had an overall error of +3.71%, with a precision of +/- 2.30%. Overall correlation between catheter Svo2 and oximeter values was the same for both devices (r = 0.99).
CONCLUSIONS: Both currently available in vivo spectrophotometric systems are capable of producing satisfactory results over wide ranges of Svo2. In contradistinction to older reports, we found that the two-wavelength Svo2 system produced results equivalent to those obtained from the three-wavelength device. In this regard, there is no detectable advantage in accuracy to measuring in vivo Svo2 with three rather than with two wavelengths.

Entities:  

Mesh:

Year:  1995        PMID: 7595690     DOI: 10.1007/BF01616992

Source DB:  PubMed          Journal:  J Clin Monit        ISSN: 0748-1977


  11 in total

1.  Clinical evaluation of the edwards laboratories and oximetrix mixed venous oxygen saturation catheters.

Authors:  J H Karis; P D Lumb
Journal:  J Cardiothorac Anesth       Date:  1988-08

2.  Clinical comparison of two- and three-wavelength systems for continuous measurement of venous oxygen saturation.

Authors:  M Chulay; J Palmer; J Neblett; A R Nancherla; D Tripodi; D Caden
Journal:  Am J Crit Care       Date:  1992-07       Impact factor: 2.228

3.  New methods for whole blood oximetry.

Authors:  J M Schmitt; F G Mihm; J D Meindl
Journal:  Ann Biomed Eng       Date:  1986       Impact factor: 3.934

4.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

5.  Continuous monitoring of mixed venous oxygen saturation.

Authors:  M B Divertie; J C McMichan
Journal:  Chest       Date:  1984-03       Impact factor: 9.410

6.  Continuous monitoring of mixed venous oxygen saturation in critically ill patients.

Authors:  P L Baele; J C McMichan; H M Marsh; J C Sill; P A Southorn
Journal:  Anesth Analg       Date:  1982-06       Impact factor: 5.108

7.  Continuous dual oximetry in surgical critical care. Indications and limitations.

Authors:  F S Bongard; T A Leighton
Journal:  Ann Surg       Date:  1992-07       Impact factor: 12.969

8.  Use of continuous monitoring of mixed venous saturation in the coronary care unit.

Authors:  J M Gore; K Sloan
Journal:  Chest       Date:  1984-11       Impact factor: 9.410

9.  In vivo comparison of two mixed venous saturation catheters.

Authors:  A Gettinger; M C DeTraglia; D D Glass
Journal:  Anesthesiology       Date:  1987-03       Impact factor: 7.892

10.  A comparison of two pulmonary artery mixed venous oxygen saturation catheters during the changing conditions of cardiac surgery.

Authors:  B R Hecker; D L Brown; D Wilson
Journal:  J Cardiothorac Anesth       Date:  1989-06
View more
  2 in total

1.  Oximetry with a multiple wavelength SLO.

Authors:  R A Ashman; F Reinholz; R H Eikelboom
Journal:  Int Ophthalmol       Date:  2001       Impact factor: 2.031

Review 2.  Central venous oxygenation: when physiology explains apparent discrepancies.

Authors:  Pierre Squara
Journal:  Crit Care       Date:  2014-11-10       Impact factor: 9.097

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.