Clinical comparison of two- and three-wavelength systems for continuous measurement of venous oxygen saturation.

M Chulay1, J Palmer, J Neblett, A R Nancherla, D Tripodi, D Caden.   

Abstract

OBJECTIVE: To evaluate clinically the accuracy of continuous SvO2 systems to reflect reference SvO2 values over a 24-hour period.
DESIGN: Randomized clinical trial.
SETTING: Six-bed cardiac surgical intensive care unit of a 540-bed federal facility. POPULATION: Sixty postoperative cardiac patients.
INTERVENTIONS: Random assignment to a two- or three-wavelength continuous SvO2 catheter for postoperative SvO2 monitoring. At 4-hour intervals over a 24-hour period, mixed venous blood samples were analyzed with a reference cooximeter and compared with the monitor value of the SvO2 catheter. MAIN OUTCOME MEASURES: A reference cooximeter method to measure SvO2 in mixed venous blood; SvO2 as measured by the in-line, SvO2 catheter system.
RESULTS: SvO2 measured by the three-wavelength system did not differ significantly from the reference SvO2 measurement. In contrast, SvO2 measured by the two-wavelength system was significantly lower than the reference SvO2 measurement within 4 hours of admission to the critical care unit and remained significantly lower throughout the 24-hour study period.
CONCLUSIONS: The results of this clinical study confirmed a previous study in dogs, showing that SvO2 is measured more accurately by the three-wavelength continuous monitoring system.

Entities:  

Mesh:

Substances:

Year:  1992        PMID: 1307880

Source DB:  PubMed          Journal:  Am J Crit Care        ISSN: 1062-3264            Impact factor:   2.228


  2 in total

1.  Validation of oxygen saturation monitoring in neonates.

Authors:  Shyang-Yun Pamela K Shiao; Ching-Nan Ou
Journal:  Am J Crit Care       Date:  2007-03       Impact factor: 2.228

2.  Simultaneous in vivo comparison of two-versus three-wavelength mixed venous (Svo2) oximetry catheters.

Authors:  F Bongard; T S Lee; T Leighton; S Y Liu
Journal:  J Clin Monit       Date:  1995-09
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.