Evaluation of two prototype devices producing noninvasive, pulsatile, calibrated blood pressure measurement from a finger.

N T Smith, K H Wesseling, B de Wit.   

Abstract

We evaluated two prototype instruments that measure pulsatile blood pressure continuously and noninvasively and compared the mean arterial pressure obtained from these devices with that obtained invasively in 17 male surgical patients. Each prototype consisted of an infrared photoplethysmograph mounted inside a finger cuff. The cuff was connected to a pressure control valve, which rapidly changed the cuff pressure so as to maintain a null pressure difference across the finger arterial wall. The resultant cuff pressure rapidly tracked the pulsatile intraarterial pressure. The prototypes reproduced absolute pressure, as well as pressure changes, accurately and linearly over a wide range of mean arterial pressures (from 2 to 164 mm Hg), with an average offset error of 0.8 mm Hg (SD +/- 3.8; range, -4.6 to 7.9), a mean scatter error of 5.3 mm Hg (range, 3.6 to 8.6), a mean regression slope of 0.97 (range, 0.79 to 1.22) and a mean correlation coefficient of the regression of 0.96 (range, 0.89 to 0.98). Both prototypes worked satisfactorily on all 17 patients, but not all the time on all patients. In 7 patients, probable arterial spasm prevented measurement of finger blood pressure 12.1% of the time, or 5.4% of the time for all patients. Ninety-six percent of the lost samples occurred with prototype 2, suggesting an instrument-related cause, rather than one related to the principle itself. The prototypes were simple to use and were almost free from artifact. Continuous monitoring for up to 7 hours on a single finger caused no harm to the finger.

Entities:  

Mesh:

Year:  1985        PMID: 4093786     DOI: 10.1007/BF02832685

Source DB:  PubMed          Journal:  J Clin Monit        ISSN: 0748-1977


  13 in total

1.  Comparison of indirect and direct methods of measuring arterial blood pressure.

Authors:  F H VAN BERGEN; D S WEATHERHEAD; A E TRELOAR; A B DOBKIN; J J BUCKLEY
Journal:  Circulation       Date:  1954-10       Impact factor: 29.690

2.  [Contribution to the continuous indirect blood pressure measurement].

Authors:  J Penáz; A Voigt; W Teichmann
Journal:  Z Gesamte Inn Med       Date:  1976-12-15

Review 3.  Comparison of indirect pressure measurements (Korotkoff) with simultaneous direct brachial artery pressure distal to the cuff.

Authors:  S B London; R E London
Journal:  Adv Intern Med       Date:  1967

4.  Noninvasive automatic determination of mean arterial pressure.

Authors:  M Ramsey
Journal:  Med Biol Eng Comput       Date:  1979-01       Impact factor: 2.602

5.  Noninvasive automatic monitoring of instantaneous arterial blood pressure using the vascular unloading technique.

Authors:  K Yamakoshi; A Kamiya; H Shimazu; H Ito; T Togawa
Journal:  Med Biol Eng Comput       Date:  1983-09       Impact factor: 2.602

6.  Can we trust the direct radial artery pressure immediately following cardiopulmonary bypass?

Authors:  D H Stern; J I Gerson; F B Allen; F B Parker
Journal:  Anesthesiology       Date:  1985-05       Impact factor: 7.892

7.  Indirect measurement of instantaneous arterial blood pressure in the human finger by the vascular unloading technique.

Authors:  K I Yamakoshi; H Shimazu; T Togawa
Journal:  IEEE Trans Biomed Eng       Date:  1980-03       Impact factor: 4.538

8.  Comparison of direct and indirect methods of measuring arterial blood pressure, part II.

Authors:  J M Bruner; L J Krenis; J M Kunsman; A P Sherman
Journal:  Med Instrum       Date:  1981 Mar-Apr

9.  Indirect measurement of instantaneous arterial blood pressure in the rat.

Authors:  K I Yamakoshi; H Shimazu; T Togawa
Journal:  Am J Physiol       Date:  1979-11

10.  Evaluating indirect blood pressure measurement techniques: a comparison of three systems in infants and children.

Authors:  R F Reder; I Dimich; M L Cohen; L Steinfeld
Journal:  Pediatrics       Date:  1978-09       Impact factor: 7.124

View more
  26 in total

1.  Evaluation of noninvasive blood pressure recording by photoplethysmography in clinical studies using angiotensin challenges.

Authors:  T Buclin; C Buchwalder-Csajka; H R Brunner; J Biollaz
Journal:  Br J Clin Pharmacol       Date:  1999-10       Impact factor: 4.335

2.  Venous responses to rhythmic exercise in contralateral forearm and calf.

Authors:  D A Duprez; M De Buyzere; J M De Sutter; S A Deman; N Y De Pue; D L Clement
Journal:  Eur J Appl Physiol Occup Physiol       Date:  1992

3.  Blood pressure levels and variance assessed by ambulatory monitoring: optimal parameters.

Authors:  F E Yates; L A Benton
Journal:  Ann Biomed Eng       Date:  1990       Impact factor: 3.934

4.  Haemodynamic effects of changes in atrioventricular and interventricular delay in cardiac resynchronisation therapy show a consistent pattern: analysis of shape, magnitude and relative importance of atrioventricular and interventricular delay.

Authors:  Z I Whinnett; J E R Davies; K Willson; C H Manisty; A W Chow; R A Foale; D Wyn Davies; A D Hughes; J Mayet; D P Francis
Journal:  Heart       Date:  2006-05-18       Impact factor: 5.994

5.  Continuous finger arterial pressure: utility in the cardiovascular laboratory.

Authors:  B P Imholz; W Wieling; G J Langewouters; G A van Montfrans
Journal:  Clin Auton Res       Date:  1991-03       Impact factor: 4.435

6.  Continuous noninvasive arterial monitoring.

Authors:  C Salter
Journal:  West J Med       Date:  1988-07

7.  The relative accuracies of two automated noninvasive arterial pressure measurement devices.

Authors:  M S Gorback; T J Quill; M L Lavine
Journal:  J Clin Monit       Date:  1991-01

8.  Initial cardiovascular response on change of posture from squatting to standing.

Authors:  F Rossberg; J Penaz
Journal:  Eur J Appl Physiol Occup Physiol       Date:  1988

9.  Identification of sleep disruption and sleep disordered breathing from the systolic blood pressure profile.

Authors:  R J Davies; K Vardi-Visy; M Clarke; J R Stradling
Journal:  Thorax       Date:  1993-12       Impact factor: 9.139

10.  Non-invasive beat to beat arterial blood pressure during non-REM sleep in obstructive sleep apnoea and snoring.

Authors:  R J Davies; J Crosby; K Vardi-Visy; M Clarke; J R Stradling
Journal:  Thorax       Date:  1994-04       Impact factor: 9.139

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.