Oscillometry and direct blood pressure: a comparative clinical study during deliberate hypotension.

M Gourdeau, R Martin, Y Lamarche, L Tétreault.   

Abstract

Oscillometry using an automatic monitor was compared with invasive blood pressure monitoring in 21 patients scheduled for surgery under general anaesthesia with deliberate hypotension. Six ranges of mean blood pressure measurements were studied, two of which were hypotensive. An excellent correlation was found between the two methods (systolic: r = 0.94; mean: r = 0.93; diastolic: r = 0.88) but there was a large variability among individual subjects. For systolic, diastolic and mean intra-arterial readings above an approximative value of 10.64 KPa (80 mmHg), the oscillometric monitor was found to underestimate blood pressure. Inversely, it was found to overestimate blood pressure for intra-arterial readings under the approximative value of 10.8 KPa (80 mmHg). We conclude that the non-invasive monitor represents a good trend estimation of the invasive radial blood pressure technique, but that wide inter-individual variability and the overestimation of blood pressure below an approximative value of 10.64 KPa (80 mmHg) precludes interchange of techniques when absolute values are considered, especially during controlled hypotension. However, oscillometry could represent a better estimate of central aortic pressure.

Entities:  

Mesh:

Year:  1986        PMID: 3719430     DOI: 10.1007/bf03010741

Source DB:  PubMed          Journal:  Can Anaesth Soc J        ISSN: 0008-2856


  11 in total

1.  Comparison of indirect and direct methods of measuring arterial blood pressure.

Authors:  F H VAN BERGEN; D S WEATHERHEAD; A E TRELOAR; A B DOBKIN; J J BUCKLEY
Journal:  Circulation       Date:  1954-10       Impact factor: 29.690

2.  The importance of cuff width in measurement of blood pressure indirectly.

Authors:  L A Geddes; R Tivey
Journal:  Cardiovasc Res Cent Bull       Date:  1976 Jan-Mar

3.  Disparities between aortic and peripheral pulse pressures induced by upright exercise and vasomotor changes in man.

Authors:  L B Rowell; G L Brengelmann; J R Blackmon; R A Bruce; J A Murray
Journal:  Circulation       Date:  1968-06       Impact factor: 29.690

4.  Direct blood pressure measurement--dynamic response requirements.

Authors:  R M Gardner
Journal:  Anesthesiology       Date:  1981-03       Impact factor: 7.892

5.  An assessment of the Dinamap 845.

Authors:  P Hutton; J Dye; C Prys-Roberts
Journal:  Anaesthesia       Date:  1984-03       Impact factor: 6.955

6.  Automatic blood pressure monitors. A clinical evaluation of five models in adults.

Authors:  C J Johnson; J H Kerr
Journal:  Anaesthesia       Date:  1985-05       Impact factor: 6.955

7.  New oscillometric method for indirect measurement of systolic and mean arterial pressure in the human finger. Part 1: model experiment.

Authors:  K Yamakoshi; H Shimazu; M Shibata; A Kamiya
Journal:  Med Biol Eng Comput       Date:  1982-05       Impact factor: 2.602

Review 8.  Comparison of direct and indirect measuring arterial blood pressure.

Authors:  J M Bruner; L J Krenis; J M Kunsman; A P Sherman
Journal:  Med Instrum       Date:  1981 Jan-Feb

9.  Clinical evaluation of Dinamap 845 automated blood pressure recorder.

Authors:  J H Silas; A T Barker; L E Ramsay
Journal:  Br Heart J       Date:  1980-02

10.  Noninvasive estimation of central aortic pressure using the oscillometric method for analyzing systemic artery pulsatile blood flow: comparative study of indirect systolic, diastolic, and mean brachial artery pressure with simultaneous direct ascending aortic pressure measurements.

Authors:  K M Borow; J W Newburger
Journal:  Am Heart J       Date:  1982-05       Impact factor: 4.749

View more
  5 in total

1.  Accuracy of four indirect methods of blood pressure measurement, with hemodynamic correlations.

Authors:  G P Gravlee; J K Brockschmidt
Journal:  J Clin Monit       Date:  1990-10

2.  The relative accuracies of two automated noninvasive arterial pressure measurement devices.

Authors:  M S Gorback; T J Quill; M L Lavine
Journal:  J Clin Monit       Date:  1991-01

3.  Validation of arterial blood pressures observed from the patient monitor; a tool for prehospital research.

Authors:  Sandy Muecke; Andrew Bersten; John Plummer
Journal:  J Clin Monit Comput       Date:  2009-12-18       Impact factor: 2.502

4.  The mean machine; accurate non-invasive blood pressure measurement in the critically ill patient.

Authors:  Sandy Muecke; Andrew Bersten; John Plummer
Journal:  J Clin Monit Comput       Date:  2009-08-12       Impact factor: 2.502

Review 5.  Brachial-ankle pulse wave velocity, cardio-ankle vascular index, and prognosis.

Authors:  Dai Ato
Journal:  Vasc Health Risk Manag       Date:  2018-10-24
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.