Validation of arterial blood pressures observed from the patient monitor; a tool for prehospital research.

Sandy Muecke1, Andrew Bersten, John Plummer.   

Abstract

OBJECTIVES: For some time, the inaccuracies of non-invasive blood pressure measurement in critically ill patients have been recognised. Measurement difficulties can occur even in optimal conditions, but in prehospital transportation vehicles, problems are exacerbated. Intra-arterial pressures must be used as the reference against which to compare the performance of non-invasive methods in the critically ill patient population. Intra-arterial manometer data observed from the patient monitor has frequently been used as the reference against which to assess the accuracy of noninvasive devices in the emergency setting. To test this method's validity, this study aimed to determine whether numerical monitor pressures can be considered interchangeable with independently sampled intra-arterial pressures.
METHODS: Intensive Care Unit nurses were asked to document arterial systolic, diastolic and mean pressures numerically displayed on the patient monitor. Observed pressures were compared to reference intra-arterial pressures independently recorded to a computer following analogue to digital conversion. Differences between observed and recorded pressures were evaluated using the Association for the Advancement of Medical Instrumentation (AAMI) protocol. Additionally, two-level linear mixed effects analyses and Bland-Altman comparisons were undertaken.
RESULTS: Systolic, diastolic and integrated mean pressures observed during 60 data collection sessions (n = 600) fulfilled AAMI protocol criteria. Integrated mean pressures were the most robust. For these pressures, mean error (reference minus observed) was 0.5 mm Hg (SD 1.4 mm Hg); 95% CI (two-level linear mixed effects analysis) 0.4-0.6 mm Hg; P < 0.001. Bland-Altman plots demonstrated tight 95% limits of agreement (-2.3 to 3.2 mm Hg), and uniform agreement across the range of mean blood pressures.
CONCLUSIONS: Integrated mean arterial pressures observed from a well maintained patient monitor can be considered interchangeable with independently sampled intra-arterial pressures and may be confidently used as the reference against which to test the accuracy of non-invasive blood pressure measuring methods in the prehospital or emergency setting.

Entities:  

Mesh:

Year:  2009        PMID: 20020187     DOI: 10.1007/s10877-009-9215-7

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  28 in total

Review 1.  Measuring agreement in method comparison studies.

Authors:  J M Bland; D G Altman
Journal:  Stat Methods Med Res       Date:  1999-06       Impact factor: 3.021

2.  STUDIES IN CLINICAL SHOCK AND HYPOTENSION; THE VALUE OF BEDSIDE HEMODYNAMIC OBSERVATIONS.

Authors:  J N COHN; M H LURIA
Journal:  JAMA       Date:  1964-12-07       Impact factor: 56.272

3.  Comparison of indirect and direct methods of measuring arterial blood pressure.

Authors:  F H VAN BERGEN; D S WEATHERHEAD; A E TRELOAR; A B DOBKIN; J J BUCKLEY
Journal:  Circulation       Date:  1954-10       Impact factor: 29.690

4.  Accuracy of blood pressure measurements made aboard helicopters.

Authors:  R B Low; D Martin
Journal:  Ann Emerg Med       Date:  1988-06       Impact factor: 5.721

5.  Direct versus indirect blood pressure measurements in critically ill patients.

Authors:  B Venus; M Mathru; R A Smith; C G Pham
Journal:  Heart Lung       Date:  1985-05       Impact factor: 2.210

6.  Analysis and correction of pressure wave distortion in fluid-filled catheter systems.

Authors:  H L Falsetti; R E Mates; R J Carroll; R L Gupta; A C Bell
Journal:  Circulation       Date:  1974-01       Impact factor: 29.690

7.  Blood pressure measurement in shock. Mechanism of inaccuracy in ausculatory and palpatory methods.

Authors:  J N Cohn
Journal:  JAMA       Date:  1967-03-27       Impact factor: 56.272

8.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

9.  Determination of frequency response from step response: application to fluid-filled catheters.

Authors:  S A Glantz; J V Tyberg
Journal:  Am J Physiol       Date:  1979-02

10.  Oscillometry and direct blood pressure: a comparative clinical study during deliberate hypotension.

Authors:  M Gourdeau; R Martin; Y Lamarche; L Tétreault
Journal:  Can Anaesth Soc J       Date:  1986-05
View more
  1 in total

1.  The mean prehospital machine; accurate prehospital non-invasive blood pressure measurement in the critically ill patient.

Authors:  Sandy Muecke; Andrew Bersten; John Plummer
Journal:  J Clin Monit Comput       Date:  2010-06-08       Impact factor: 2.502

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.