Accuracy and dynamic response of disposable pressure transducer-tubing systems.

P Hunziker.   

Abstract

Accuracy and dynamic response of seven different disposable saline-filled transducer-tubing systems were evaluated in two configurations; a pole-mount version (60-inch tubing) and patient-mount version (12-inch tubing). Natural frequencies (Fn) and damping coefficients were measured as were pressure waveform recordings. A 0.03 ml air bubble was inserted into the different systems in close proximity to the transducer and measurements were repeated. There were dramatic changes in dynamic response with lengthening of tubing to a pole-mount configuration and air bubble additions. Only one transducer-tubing system had adequate accuracy and dynamic response in both the pole- and patient-mount version. Pressure waveforms in the pole-mount versions and with the air bubble produced systolic overshoots up to 35 mmHg. The performance characteristics dictate that the choice for a disposable transducer-tubing system must be made not only by design, but dynamic response which is directly related to clinical set-up.

Entities:  

Mesh:

Year:  1987        PMID: 3608063     DOI: 10.1007/BF03010146

Source DB:  PubMed          Journal:  Can J Anaesth        ISSN: 0832-610X            Impact factor:   5.063


  4 in total

1.  Direct blood pressure measurement--dynamic response requirements.

Authors:  R M Gardner
Journal:  Anesthesiology       Date:  1981-03       Impact factor: 7.892

2.  Use of a needle valve variable resistor to improve invasive blood pressure monitoring.

Authors:  J H Abrams; M L Olson; J A Marino; F B Cerra
Journal:  Crit Care Med       Date:  1984-11       Impact factor: 7.598

3.  Effect of the dynamic response of transducer-tubing system on accuracy of direct blood pressure measurement in patients.

Authors:  A Boutros; S Albert
Journal:  Crit Care Med       Date:  1983-02       Impact factor: 7.598

4.  The dynamic responses of liquid-filled catheter systems for direct measurements of blood pressure.

Authors:  T Shinozaki; R S Deane; J E Mazuzan
Journal:  Anesthesiology       Date:  1980-12       Impact factor: 7.892

  4 in total
  11 in total

1.  Development of a wireless intra-vaginal transducer for monitoring intra-abdominal pressure in women.

Authors:  Tanner J Coleman; Jens C Thomsen; Sean D Maass; Yvonne Hsu; Ingrid E Nygaard; Robert W Hitchcock
Journal:  Biomed Microdevices       Date:  2012-04       Impact factor: 2.838

2.  A simple solution to ringing phenomenon.

Authors:  Sameer Sethi; Virender Kumar Arya
Journal:  J Clin Monit Comput       Date:  2011-01-08       Impact factor: 2.502

3.  Frequency characteristics of pressure transducer kits with inserted pressure-resistant extension tubes.

Authors:  Shigeki Fujiwara; Satoshi Mori; Keiichi Tachihara; Takeshi Yamamoto; Chizuko Yokoe; Uno Imaizumi; Yoshinari Morimoto; Yoichiro Miki; Izumi Toyoguchi; Kazu-Ichi Yoshida; Takeshi Yokoyama
Journal:  J Clin Monit Comput       Date:  2016-03-05       Impact factor: 2.502

4.  Dynamic response of the ROSE damping device.

Authors:  B Kleinman; S Powell
Journal:  J Clin Monit       Date:  1989-04

5.  An evaluation of the N-CAT, a new arterial tonometer.

Authors:  N R Searle; J Gauthier; P Sahab
Journal:  Can J Anaesth       Date:  1995-06       Impact factor: 5.063

6.  Predictive Values for Time from Transducer Stopcock Closure to Accurate Intracranial Pressure Reading.

Authors:  Corrie Earthman; Kathrina Siaron; Jennifer Wilson; DaiWai M Olson
Journal:  Neurocrit Care       Date:  2022-08-24       Impact factor: 3.532

7.  Development of a novel intra-vaginal transducer with improved dynamic response.

Authors:  Paul J Johnson; Evan M Rosenbluth; Ingrid E Nygaard; Monir K Parikh; Robert W Hitchcock
Journal:  Biomed Microdevices       Date:  2009-12       Impact factor: 2.838

8.  Comparison of intraarterial with continuous noninvasive blood pressure measurement in postoperative pediatric patients.

Authors:  J K Triedman; J P Saul
Journal:  J Clin Monit       Date:  1994-01

9.  Intraarterial catheter diameter and dynamic response of arterial pressure monitoring system: a randomized controlled trial.

Authors:  Hyongmin Oh; Suk Hyung Choe; Yoon Jung Kim; Hyun-Kyu Yoon; Hyung-Chul Lee; Hee-Pyoung Park
Journal:  J Clin Monit Comput       Date:  2021-02-01       Impact factor: 2.502

10.  Correlation of invasive and non-invasive blood pressure: A must for management.

Authors:  Sameer Sethi
Journal:  Indian J Anaesth       Date:  2010-11
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.