Intraarterial catheter diameter and dynamic response of arterial pressure monitoring system: a randomized controlled trial.

Hyongmin Oh1, Suk Hyung Choe1, Yoon Jung Kim1, Hyun-Kyu Yoon1, Hyung-Chul Lee1, Hee-Pyoung Park2.   

Abstract

The dynamic response (DR) of the arterial pressure monitoring system (APMS) may depend on the intraarterial catheter (IAC) diameter. We hypothesized that adequate DR would be more common when using a smaller IAC. We compared the DR of the AMPS (Auto Transducer™) between three IACs (BD Angiocath Plus™) with different diameters. 353 neurosurgical patients were randomized into three groups undergoing catheterization with a 20-, 22-, or 24-gauge IAC: 20G (n = 119), 22G (n = 117), and 24G (n = 117) groups, respectively. The DR, which depends on the natural frequency and damping coefficient, was divided into four types: adequate (primary outcome measure), underdamped, overdamped, and unacceptable. The frequency of intraoperative IAC malfunction was noted. Adequate DR was observed more frequently in the 22G and 24G groups than the 20G group (13.7% and 15.4% vs. 4.2%, P = 0.011 and 0.004, respectively). The frequency of underdamped DR was higher in the 20G group than the 24G group (86.6% vs. 69.2%, P = 0.001), whereas overdamped DR was more frequent in the 24G group than the 20G and 22G groups (6.0% vs. 0.0% and 0.0%, P = 0.007 and 0.014, respectively). IAC malfunctioned more frequently during surgery in the 24G group than the 20G and 22G groups (15.4% vs. 0.0% and 1.7%, P < 0.001 and P < 0.001, respectively). The frequency of adequate DR was low regardless of the IAC diameter. Nonetheless, in terms of DR and IAC malfunction, a 22-gauge BD Angiocath Plus™ was more suitable for invasive blood pressure monitoring with Auto Transducer™ than a 20- or 24-gauge BD Angiocath Plus™. Registration Registry: ClinicalTrials.gov. Registration number: NCT03642756. Date of Registration: July 27, 2018.
© 2021. The Author(s), under exclusive licence to Springer Nature B.V. part of Springer Nature.

Entities:  

Keywords:  Arterial pressure monitoring system; Dynamic response; Intraarterial catheter diameter; Invasive blood pressure monitoring

Mesh:

Year:  2021        PMID: 33527182     DOI: 10.1007/s10877-021-00663-7

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  27 in total

1.  Dynamic response of liquid-filled catheter systems for measurement of blood pressure: precision of measurements and reliability of the Pressure Recording Analytical Method with different disposable systems.

Authors:  Stefano Romagnoli; Salvatore Mario Romano; Sergio Bevilacqua; Chiara Lazzeri; Gian Franco Gensini; Carlo Pratesi; Diego Quattrone; Daniele Dini; Angelo Raffaele De Gaudio
Journal:  J Crit Care       Date:  2010-10-30       Impact factor: 3.425

2.  Evaluation of dynamic performance in liquid-filled catheter systems for measuring invasive blood pressure.

Authors:  M Todorovic; E W Jensen; C Thøgersen
Journal:  Int J Clin Monit Comput       Date:  1996-08

3.  Radial artery cannulation in 1,000 patients: precautions and complications.

Authors:  M A Mandel; P J Dauchot
Journal:  J Hand Surg Am       Date:  1977-11       Impact factor: 2.230

4.  Dynamic response of the ROSE damping device.

Authors:  B Kleinman; S Powell
Journal:  J Clin Monit       Date:  1989-04

5.  Accuracy and dynamic response of disposable pressure transducer-tubing systems.

Authors:  P Hunziker
Journal:  Can J Anaesth       Date:  1987-07       Impact factor: 5.063

6.  Direct blood pressure measurement--dynamic response requirements.

Authors:  R M Gardner
Journal:  Anesthesiology       Date:  1981-03       Impact factor: 7.892

7.  Effect of the dynamic response of transducer-tubing system on accuracy of direct blood pressure measurement in patients.

Authors:  A Boutros; S Albert
Journal:  Crit Care Med       Date:  1983-02       Impact factor: 7.598

Review 8.  Arterial cannulation: a critical review.

Authors:  Teresa R Cousins; John M O'Donnell
Journal:  AANA J       Date:  2004-08

9.  Peripheral arterial blood pressure versus central crterial blood pressure monitoring in critically ill patients after Cardio-pulmonary Bypass.

Authors:  Rana Altaf Ahmad; Suhail Ahmad; Anjum Naveed; Mirza Ahmad Raza Baig
Journal:  Pak J Med Sci       Date:  2017 Mar-Apr       Impact factor: 1.088

Review 10.  Clinical review: complications and risk factors of peripheral arterial catheters used for haemodynamic monitoring in anaesthesia and intensive care medicine.

Authors:  Bernd Scheer; Azriel Perel; Ulrich J Pfeiffer
Journal:  Crit Care       Date:  2002-04-18       Impact factor: 9.097

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.