Abbas Shokri1, Fariborz Vafaee2, Leila Haghighat3, Shiva Shahabi4, Maryam Farhadian5, Mohammad Reza Jamalpour6.
Abstract
BACKGROUND: This study aimed to compare the amount of artifacts induced by the titanium and zirconium implants on cone-beam computed tomography (CBCT) and assess the effect of different exposure settings on the image quality for both materials.Entities:
Keywords: Artifacts; Cone-beam computed tomography; Titanium; Zirconium
Mesh:
Substances:
Year: 2022 PMID: 36057551 PMCID: PMC9440550 DOI: 10.1186/s12880-022-00884-5
Source DB: PubMed Journal: BMC Med Imaging ISSN: 1471-2342 Impact factor: 2.795
Fig. 1A Prepared bone blocks. B Bone block with an inserted implant. C Implant containing rims prepared for imaging
Fig. 2The evaluation of the amount of artifacts in different study groups
Comparison of the amount of artifacts induced by zirconium and titanium implants
| FOV | Resolution | Implant’s type | Gray value (mean ± SD) | Mean difference ± SE | |
|---|---|---|---|---|---|
| Small | High (133 µ voxel size) | Zirconium | 1566.10 ± 0.77 | 1098.65 ± 41.50 | < 0.001 |
| Titanium | 167.44 ± 143.76 | ||||
| Low (200 µ voxel size) | Zirconium | 1434.46 ± 0.9 | 1061.18 ± 33.96 | < 0.001 | |
| Titanium | 373.28 ± 117.65 | ||||
| Large | High (133 µ voxel size) | Zirconium | 1251.02 ± 62.74 | 928.86 ± 60.71 | < 0.001 |
| Titanium | 322.16 ± 200.75 | ||||
| Low (200 µ voxel size) | Zirconium | 1329.05 ± 36.43 | 1016.52 ± 49.40 | < 0.001 | |
| Titanium | 312.53 ± 167.23 |
*Independent t test, FOV field of view, SD standard deviation, SE standard error
Fig. 3Box diagrams comparing the amount of artifacts induced by zirconium and titanium implants in different fields of view and resolutions
Comparison of the amount of artifacts induced by study groups in large and small FOVs
| Implant’s type | Resolution | FOV | Gray value (mean ± SD) | Mean difference ± SE | |
|---|---|---|---|---|---|
| Zirconium | High (133 µ voxel size) | Small | 1566.10 ± 0.77 | 315.07 ± 18.11 | < 0.001 |
| Large | 1251.02 ± 62.74 | ||||
| Low (200 µ voxel size) | Small | 1434.46 ± 0.9 | 105.40 ± 10.52 | < 0.001 | |
| Large | 1329.05 ± 36.43 | ||||
| Titanium | High (133 µ voxel size) | Small | 467.44 ± 143.7 | 145.28 ± 71.28 | 0.212 |
| Large | 322.16 ± 200.75 | ||||
| Low (200 µ voxel size) | Small | 373.28 ± 117.65 | 60.75 ± 59.02 | 0.281 | |
| Large | 312.53 ± 167.23 |
*Paired t test, FOV field of view, SD standard deviation, SE standard error
Comparison of the amount of artifacts induced by study groups in high and low resolutions
| Implant’s type | Resolution | Gray value (mean ± SD) | Mean difference ± SE | |
|---|---|---|---|---|
| Zirconium | High (133 µ voxel size) | 1408.56 ± 166.67 | 26.80 ± 36.12 | < 0.001 |
| Low (200 µ voxel size) | 1381.76 ± 59.44 | |||
| Titanium | High (133 µ voxel size) | 394.80 ± 186.18 | 51.89 ± 48.14 | 0.573 |
| Low (200 µ voxel size) | 342.90 ± 144.77 |
*Paired t test, SD standard deviation, SE standard error