Influence of Cone Beam Computed Tomography Settings on Implant Artifact Production: Zirconia and Titanium.

Taruska Ventorini Vasconcelos, Eduarda Helena Leandro Nascimento, Boulos B Bechara, Deborah Queiroz Freitas, Marcel Noujeim.   

Abstract

PURPOSE: This in vitro study assessed the artifact production related to titanium and zirconia implants in cone beam computed tomography (CBCT) and compared the effect of different protocol settings on image quality for both materials.
MATERIALS AND METHODS: A titanium implant and a zirconia implant were placed in a dry mandible. CBCT scans were obtained separately for each implant using the ProMax 3D (Planmeca) unit; 20 protocols were tested with varying kilovoltage (70 to 90 kVp) and resolution (high and low), and with and without a metal artifact reduction tool. Standard deviation and contrast-noise ratio were calculated in regions of interest adjacent and distant to the implant.
RESULTS: The zirconia produced more artifacts and its images were more affected by the different protocols than titanium. High kVps and an activated metal artifact reduction tool decreased the standard deviation values related to both implants. Activation of the metal artifact reduction tool also increased contrast-noise ratio values for both implants, whereas increasing kVp improved them only on titanium images. The standard deviation and contrast-noise ratio were not affected by resolution.
CONCLUSION: The zirconia implant generated more image artifacts than the titanium implant. Increasing kVp and the metal artifact reduction tool are efficient in decreasing the CBCT artifacts for both implants, whereas resolution does not affect their production.

Entities:  

Year:  2019        PMID: 31528863     DOI: 10.11607/jomi.7129

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  8 in total

1.  Artefacts at different distances from titanium and zirconia implants in cone-beam computed tomography: effect of tube current and metal artefact reduction.

Authors:  Arthur Xavier Maseti Mancini; Matheus Urias Cruz Santos; Hugo Gaêta-Araujo; Camila Tirapelli; Ruben Pauwels; Christiano Oliveira-Santos
Journal:  Clin Oral Investig       Date:  2021-02-05       Impact factor: 3.573

Review 2.  Diagnostic accuracy of imaging examinations for peri-implant bone defects around titanium and zirconium dioxide implants: A systematic review and meta-analysis.

Authors:  Mariana Murai Chagas; Solange Kobayashi-Velasco; Thais Gimenez; Marcelo Gusmão Paraiso Cavalcanti
Journal:  Imaging Sci Dent       Date:  2021-11-18

Review 3.  The Impact of Cone-Beam Computed Tomography Exposure Parameters on Peri-Implant Artifacts: A Literature Review.

Authors:  Paweł Sawicki; Paweł J Zawadzki; Piotr Regulski
Journal:  Cureus       Date:  2022-03-10

4.  In Vitro Quantitative Evaluation of Postprocessing Filter for Metal Artifact Reduction in Cone Beam Computed Tomography Images of Titanium and Zirconium Dioxide Implants.

Authors:  Andre Luiz Ferreira Costa; Karolina A C Fardim; Jennifer M Mantoani; Ana Lucia Franco Ricardo; Maria Aparecida N Jardini; Kaan Orhan; Sérgio Lúcio Pereira de Castro Lopes
Journal:  Biomed Res Int       Date:  2022-02-24       Impact factor: 3.411

5.  Comparison of the amount of artifacts induced by zirconium and titanium implants in cone-beam computed tomography images.

Authors:  Abbas Shokri; Fariborz Vafaee; Leila Haghighat; Shiva Shahabi; Maryam Farhadian; Mohammad Reza Jamalpour
Journal:  BMC Med Imaging       Date:  2022-09-03       Impact factor: 2.795

6.  CBCT artefact-burden of zirconia-based as compared to titanium implants for different beam energies: an analytical approach.

Authors:  Ralf Schulze
Journal:  Sci Rep       Date:  2022-09-10       Impact factor: 4.996

7.  Influence of windowing and metal artefact reduction algorithms on the volumetric dimensions of five different high-density materials: a cone-beam CT study.

Authors:  Fernanda Coelho-Silva; Luciano Augusto Cano Martins; Daniela Azeredo Braga; Eliana Zandonade; Francisco Haiter-Neto; Sergio Lins de-Azevedo-Vaz
Journal:  Dentomaxillofac Radiol       Date:  2020-06-25       Impact factor: 2.419

8.  Cone-beam computed tomography artifacts in the presence of dental implants and associated factors: an integrative review.

Authors:  Bianca Rodrigues Terrabuio; Caroline Gomes Carvalho; Mariela Peralta-Mamani; Paulo Sérgio da Silva Santos; Izabel Regina Fischer Rubira-Bullen; Cássia Maria Fischer Rubira
Journal:  Imaging Sci Dent       Date:  2021-03-11
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.