Quantification of metal artifacts on cone beam computed tomography images.

Ruben Pauwels1, Harry Stamatakis, Hilde Bosmans, Ria Bogaerts, Reinhilde Jacobs, Keith Horner, Kostas Tsiklakis.   

Abstract

OBJECTIVES: To quantify metal artifacts obtained from a wide range of cone beam computed tomography (CBCT) devices and exposure protocols, to compare their tolerance to metals of different densities, and to provide insights regarding the possible implementation of metal artifact analysis into a QC protocol for CBCT.
MATERIALS AND METHODS: A customized polymethyl methacrylate (PMMA) phantom, containing titanium and lead rods, was fabricated. It was scanned on 13 CBCT devices and one multi-slice computed tomography (MSCT) device, including high-dose and low-dose exposure protocols. Artifacts from the rods were assessed by two observers by measuring the standard deviation of voxel values in the vicinity of the rods, and normalizing this value to the percentage of the theoretical maximum standard deviation.
RESULTS: For CBCT datasets, artifact values ranged between 6.1% and 27.4% for titanium, and between 10.% and 43.7% for lead. Most CBCT devices performed worse than MSCT for titanium artifacts, but all of them performed better for lead artifacts. In general, no clear improvement of metal artifacts was seen for high-dose protocols, although certain devices showed some artifact reduction for large FOV or high exposure protocols.
CONCLUSIONS: Regions in the vicinity of the metal rods were moderately or gravely affected, particularly in the area between the rods. In practice, the CBCT user has very limited possibilities to reduce artifacts. Researchers and manufacturers need to combine their efforts in optimizing exposure factors and implementing metal artifact reduction algorithms.
© 2011 John Wiley & Sons A/S.

Entities:  

Keywords:  cone beam computed tomography; image analysis; metal artifacts; titanium

Mesh:

Substances:

Year:  2011        PMID: 22168574     DOI: 10.1111/j.1600-0501.2011.02382.x

Source DB:  PubMed          Journal:  Clin Oral Implants Res        ISSN: 0905-7161            Impact factor:   5.977


  64 in total

1.  Image quality assessment of three cone beam CT machines using the SEDENTEXCT CT phantom.

Authors:  J Bamba; K Araki; A Endo; T Okano
Journal:  Dentomaxillofac Radiol       Date:  2013       Impact factor: 2.419

2.  Automated implant segmentation in cone-beam CT using edge detection and particle counting.

Authors:  Ruben Pauwels; Reinhilde Jacobs; Hilde Bosmans; Pisha Pittayapat; Pasupen Kosalagood; Onanong Silkosessak; Soontra Panmekiate
Journal:  Int J Comput Assist Radiol Surg       Date:  2014-07       Impact factor: 2.924

3.  Artifacts in orthodontic bracket systems in cone-beam computed tomography and multislice computed tomography.

Authors:  V Hirschinger; S Hanke; U Hirschfelder; E Hofmann
Journal:  J Orofac Orthop       Date:  2015-03       Impact factor: 1.938

4.  Do the number of basis images and metal artifact reduction affect the production of artifacts near and far from zirconium dental implants in CBCT?

Authors:  Eduarda Helena Leandro Nascimento; Hugo Gaêta-Araujo; Rocharles Cavalcante Fontenele; Nicolly Oliveira-Santos; Christiano Oliveira-Santos; Deborah Queiroz Freitas
Journal:  Clin Oral Investig       Date:  2021-02-24       Impact factor: 3.573

5.  Image quality optimization using a narrow vertical detector dental cone-beam CT.

Authors:  Danieli Moura Brasil; Ruben Pauwels; Wim Coucke; Francisco Haiter-Neto; Reinhilde Jacobs
Journal:  Dentomaxillofac Radiol       Date:  2019-01-31       Impact factor: 2.419

6.  Metal and motion artifacts by cone beam computed tomography (CBCT) in dental and maxillofacial study.

Authors:  Cosimo Nardi; Claudia Borri; Francesco Regini; Linda Calistri; Alessandro Castellani; Chiara Lorini; Stefano Colagrande
Journal:  Radiol Med       Date:  2015-01-30       Impact factor: 3.469

7.  Comparative study of image quality and radiation dose of cone beam and low-dose multislice computed tomography--an in-vitro investigation.

Authors:  Elisabeth Hofmann; Matthias Schmid; Martin Sedlmair; Rosemarie Banckwitz; Ursula Hirschfelder; Michael Lell
Journal:  Clin Oral Investig       Date:  2013-03-05       Impact factor: 3.573

8.  Assessment of metal artefact reduction around dental titanium implants in cone beam CT.

Authors:  A Parsa; N Ibrahim; B Hassan; K Syriopoulos; P van der Stelt
Journal:  Dentomaxillofac Radiol       Date:  2014-08-19       Impact factor: 2.419

9.  Variability of dental cone beam CT grey values for density estimations.

Authors:  R Pauwels; O Nackaerts; N Bellaiche; H Stamatakis; K Tsiklakis; A Walker; H Bosmans; R Bogaerts; R Jacobs; K Horner
Journal:  Br J Radiol       Date:  2013-01       Impact factor: 3.039

10.  Bisphosphonate-induced osteonecrosis of the jaw: comparison of disease extent on contrast-enhanced MR imaging, [18F] fluoride PET/CT, and conebeam CT imaging.

Authors:  R Guggenberger; D R Fischer; P Metzler; G Andreisek; D Nanz; C Jacobsen; D T Schmid
Journal:  AJNR Am J Neuroradiol       Date:  2012-12-06       Impact factor: 3.825

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.