Michael Barnes1,2, Dennis Pomare1, Marcus Doebrich1, Therese S Standen1, Joshua Wolf1,3, Peter Greer1,2, John Simpson1,2.
Abstract
PURPOSE: Multi-leaf-collimator (MLC) leaf position accuracy is important for accurate dynamic radiotherapy treatment plan delivery. Machine log files have become widely utilized for quality assurance (QA) of such dynamic treatments. The primary aim is to test the sensitivity of machine log files in comparison to electronic portal imaging device (EPID)-based measurements to MLC position errors caused by leaf backlash. The secondary aim is to investigate the effect of MLC leaf backlash on MLC leaf motion during clinical dynamic plan delivery.Entities:
Keywords: EPID; MLC; log files; quality assurance (QA)
Mesh:
Year: 2022 PMID: 35678793 PMCID: PMC9512360 DOI: 10.1002/acm2.13660
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.243
FIGURE 1Schematic diagram of a Varian multi‐leaf‐collimator (MLC) leaf showing the mechanical relationship between the MLC motor, lead screw, T‐nut, and MLC leaf (not to scale). (a) Without modification, (b) after modification via sandpapering the top of the “T” as performed in this study.
FIGURE 2Example Stakitt test plan image.
Measured backlash and micrometer results for the modified leaves
| Distance at isocenter plane (mm) | ||||
|---|---|---|---|---|
| MLC leaf | Total backlash after | Backlash change | Micrometer change | Difference micrometer change—backlash change |
| A20 | 0.55 | 0.31 | 0.39 | 0.08 |
| A24 | 0.76 | 0.57 | 0.56 | −0.01 |
| A28 | 0.59 | 0.33 | 0.40 | 0.07 |
| A34 | 0.78 | 0.67 | 0.62 | −0.05 |
| A38 | 0.74 | 0.59 | 0.60 | 0.01 |
| A42 | 0.57 | 0.37 | 0.49 | 0.12 |
| B22 | 0.71 | 0.49 | 0.46 | −0.03 |
| B26 | 0.80 | 0.59 | 0.63 | 0.04 |
| B30 | 0.67 | 0.51 | 0.44 | −0.07 |
| B32 | 0.86 | 0.61 | 0.66 | 0.05 |
| B36 | 0.67 | 0.43 | 0.46 | 0.03 |
| B40 | 0.80 | 0.55 | 0.60 | 0.05 |
aMulti‐leaf‐collimator.
FIGURE 3Stakitt test measured results. Difference (mm) between before and after modification for each of the modified leaves alongside a combined set of five unmodified control leaves. Results normalized to the mean of the control leaves. The boxplot whiskers cover 95% of the data with any outliers plotted individually. Differences measured by the micrometer included (red crosses) for comparison. Differences between the mean Stakitt measured value and the micrometer result presented in the accompanying tables for each leaf. a) Gantry 0 degrees for MLC bank A, b) Gantry 0 degrees for MLC Bank B, c) Gantry 90 degrees for MLC Bank A, d) Gantry 90 degrees for MLC bank B, e) Gantry 270 degrees for MLC Bank A, f) Gantry 270 degrees for MLC Bank B
FIGURE 4Stakitt measured leaf position differences before/after modification for each of the modified leaves plotted per picket number. A40 is included as an example control leaf.
Mean QuAVER measured multi‐leaf‐collimator (MLC) deviations from plan for the test plan compared to the built‐in backlash test results
| Measured mean leaf position error (abs, mm, both arcs combined) | |||||||||
|---|---|---|---|---|---|---|---|---|---|
| Before T‐nut modification | After T‐nut modification | ||||||||
| Bank | Leaf | Measurement: all gantry angles | Difference (measurement—backlash) | Measurement: all gantry angles | Difference (measurement—backlash) | Measurement: negative gantry angles | Difference (measurement—backlash) | Measurement: positive gantry angles | Difference (measurement—backlash) |
| A | 20 | 0.28 | 0.04 | 0.39 | −0.16 | 0.60 | 0.05 | 0.20 | −0.35 |
| 24 | 0.12 | −0.08 | 0.39 | −0.37 | 0.71 | −0.06 | 0.07 | −0.69 | |
| 28 | 0.17 | −0.09 | 0.23 | −0.35 | 0.40 | −0.19 | 0.07 | −0.52 | |
| 34 | 0.00 | −0.12 | 0.23 | −0.56 | 0.60 | −0.18 | 0.14 | −0.65 | |
| 38 | 0.05 | −0.11 | 0.25 | −0.50 | 0.57 | −0.18 | 0.07 | −0.67 | |
| 42 | 0.05 | −0.14 | 0.03 | −0.54 | 0.31 | −0.26 | 0.24 | −0.33 | |
| B | 22 | 0.19 | −0.02 | 0.02 | −0.68 | 0.12 | −0.58 | 0.17 | −0.54 |
| 26 | 0.07 | −0.15 | 0.26 | −0.54 | 0.03 | −0.77 | 0.55 | −0.25 | |
| 30 | 0.03 | −0.13 | 0.22 | −0.44 | 0.04 | −0.63 | 0.43 | −0.24 | |
| 32 | 0.06 | −0.19 | 0.35 | −0.51 | 0.01 | −0.85 | 0.69 | −0.17 | |
| 36 | 0.02 | −0.22 | 0.23 | −0.43 | 0.08 | −0.58 | 0.54 | −0.13 | |
| 40 | 0.13 | −0.12 | 0.43 | −0.37 | 0.03 | −0.77 | 0.76 | −0.05 | |
FIGURE 5QuAVER result histograms for measured deviations to the plan for Leaf A24 for Test Plan 9, including both arcs. Top left = electronic portal imaging device (EPID) measurements before T‐nut modification, top right = log file measurement before T‐nut modification, bottom left = EPID measurement after modification, bottom right = log file measurement after modification. The EPID‐after histogram has been presented split into different colors for the positive gantry angles (toward G90) and the negative gantry angles (toward G270).
FIGURE 6Multi‐leaf‐collimator (MLC) leaf A24 trajectory during CCW arc in Test Plan 9 as measured with QuAVER electronic portal imaging device (EPID) before and after T‐nut modification and with log file.