Commissioning and quality assurance for VMAT delivery systems: An efficient time-resolved system using real-time EPID imaging.

Benjamin J Zwan1,2, Michael P Barnes3,4, Jonathan Hindmarsh1, Seng B Lim5, Dale M Lovelock5, Todsaporn Fuangrod3, Daryl J O'Connor2, Paul J Keall6, Peter B Greer2,3.   

Abstract

PURPOSE: An ideal commissioning and quality assurance (QA) program for Volumetric Modulated Arc Therapy (VMAT) delivery systems should assess the performance of each individual dynamic component as a function of gantry angle. Procedures within such a program should also be time-efficient, independent of the delivery system and be sensitive to all types of errors. The purpose of this work is to develop a system for automated time-resolved commissioning and QA of VMAT control systems which meets these criteria.
METHODS: The procedures developed within this work rely solely on images obtained, using an electronic portal imaging device (EPID) without the presence of a phantom. During the delivery of specially designed VMAT test plans, EPID frames were acquired at 9.5 Hz, using a frame grabber. The set of test plans was developed to individually assess the performance of the dose delivery and multileaf collimator (MLC) control systems under varying levels of delivery complexities. An in-house software tool was developed to automatically extract features from the EPID images and evaluate the following characteristics as a function of gantry angle: dose delivery accuracy, dose rate constancy, beam profile constancy, gantry speed constancy, dynamic MLC positioning accuracy, MLC speed and acceleration constancy, and synchronization between gantry angle, MLC positioning and dose rate. Machine log files were also acquired during each delivery and subsequently compared to information extracted from EPID image frames.
RESULTS: The largest difference between measured and planned dose at any gantry angle was 0.8% which correlated with rapid changes in dose rate and gantry speed. For all other test plans, the dose delivered was within 0.25% of the planned dose for all gantry angles. Profile constancy was not found to vary with gantry angle for tests where gantry speed and dose rate were constant, however, for tests with varying dose rate and gantry speed, segments with lower dose rate and higher gantry speed exhibited less profile stability. MLC positional accuracy was not observed to be dependent on the degree of interdigitation. MLC speed was measured for each individual leaf and slower leaf speeds were shown to be compensated for by lower dose rates. The test procedures were found to be sensitive to 1 mm systematic MLC errors, 1 mm random MLC errors, 0.4 mm MLC gap errors and synchronization errors between the MLC, dose rate and gantry angle controls systems of 1°. In general, parameters measured by both EPID and log files agreed with the plan, however, a greater average departure from the plan was evidenced by the EPID measurements.
CONCLUSION: QA test plans and analysis methods have been developed to assess the performance of each dynamic component of VMAT deliveries individually and as a function of gantry angle. This methodology relies solely on time-resolved EPID imaging without the presence of a phantom and has been shown to be sensitive to a range of delivery errors. The procedures developed in this work are both comprehensive and time-efficient and can be used for streamlined commissioning and QA of VMAT delivery systems.
© 2017 American Association of Physicists in Medicine.

Entities:  

Keywords:  EPID; MLC; VMAT; commissioning; quality assurance

Mesh:

Year:  2017        PMID: 28564208     DOI: 10.1002/mp.12387

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  5 in total

1.  A novel quality assurance procedure for trajectory log validation using phantom-less real-time latency corrected EPID images.

Authors:  Seng Boh Lim; Benjamin J Zwan; Danny Lee; Peter B Greer; Dale Michael Lovelock
Journal:  J Appl Clin Med Phys       Date:  2021-02-26       Impact factor: 2.102

2.  Quality assurance of VMAT on flattened and flattening filter-free accelerators using a high spatial resolution detector.

Authors:  F S Matar; D Wilkinson; J Davis; G Biasi; T Causer; I Fuduli; O Brace; N Stansook; M Carolan; A B Rosenfeld; Marco Petasecca
Journal:  J Appl Clin Med Phys       Date:  2020-04-11       Impact factor: 2.102

3.  Efficient quality assurance method with automated data acquisition of a single phantom setup to determine radiation and imaging isocenter congruence.

Authors:  Hyejoo Kang; Rakesh Patel; John C Roeske
Journal:  J Appl Clin Med Phys       Date:  2019-09-19       Impact factor: 2.102

4.  Insensitivity of machine log files to MLC leaf backlash and effect of MLC backlash on clinical dynamic MLC motion: An experimental investigation.

Authors:  Michael Barnes; Dennis Pomare; Marcus Doebrich; Therese S Standen; Joshua Wolf; Peter Greer; John Simpson
Journal:  J Appl Clin Med Phys       Date:  2022-06-09       Impact factor: 2.243

5.  Quality assurance of geometric accuracy based on an electronic portal imaging device and log data analysis for Dynamic WaveArc irradiation.

Authors:  Hideaki Hirashima; Yuki Miyabe; Mitsuhiro Nakamura; Nobutaka Mukumoto; Takashi Mizowaki; Masahiro Hiraoka
Journal:  J Appl Clin Med Phys       Date:  2018-04-06       Impact factor: 2.102

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.