Charles Kirkby1,2,3, Esmaeel Ghasroddashti1,2,3, Crystal Plume Angers4, Grace Zeng5, Erin Barnett6.
Abstract
The Canadian Organization of Medical Physicists (COMP), in close partnership with the Canadian Partnership for Quality Radiotherapy (CPQR) has developed a series of Technical Quality Control (TQC) guidelines for radiation treatment equipment. These guidelines outline the performance objectives that equipment should meet in order to ensure an acceptable level of radiation treatment quality. The TQC guidelines have been rigorously reviewed and field tested in a variety of Canadian radiation treatment facilities. The development process enables rapid review and update to keep the guidelines current with changes in technology (the most updated version of this guideline can be found on the CPQR website). This particular TQC details recommended quality control testing for medical linear accelerators and multileaf collimators.Entities:
Keywords: linear accelerator; multileaf collimator; radiotherapy quality assurance
Mesh:
Year: 2017 PMID: 29205837 PMCID: PMC5849832 DOI: 10.1002/acm2.12236
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Daily quality control tests
| Designator | Test | Performance | |
|---|---|---|---|
| Tolerance | Action | ||
| Daily | |||
| DL1 | Motion interlock | Functional | |
| DL2 | Couch brakes | Functional | |
| DL3 | Beam interrupt/counters | Functional | |
| DL4 | Lasers/crosshairs | 1 mm | 2 mm |
| DL5 | Optical distance indicator | 1 mm | 2 mm |
| DL6 | Optical back pointer | 2 mm | 3 mm |
| DL7 | Field definition: Jaws/MLC leaves | 1 mm | 2 mm |
| DL8 | Output constancy – photons | 2% | 3% |
| DL9 | Output constancy – electrons | 2% | 3% |
| DL10 | Dynamic (Varian), Virtual (Siemens) or Universal (Elekta) wedge factors | 2% | 3% |
Monthly quality control tests
| Designator | Test | Performance | |
|---|---|---|---|
| Tolerance | Action | ||
| Monthly | |||
| ML1 | Wedge, tray, cone, interlocks | Functional | |
| ML2 | Accessories integrity and centering | Functional | |
| ML3 | Gantry angle readouts | 0.5° | 1.0° |
| ML4 | Collimator angle readouts | 0.5° | 1.0° |
| ML5 | Crosshairs centering/collimator rotation isocenter (mechanical) | 1 mm | 2 mm |
| ML6 | Couch position readouts | 1 mm | 2 mm |
| ML7 | Couch rotation isocenter (mechanical) | 1 mm | 2 mm |
| ML8 | Couch isocentric angle | 0.5° | 1.0° |
| ML9 | Optical distance indicator | 1 mm | 2 mm |
| ML10 | Relative dosimetry | 2% | 3% |
| ML11 | Central axis depth dose reproducibility | 1%/2 mm | 2%/3 mm |
| ML12 | Beam profile constancy | 2% | 3% |
| ML13 | Light/radiation coincidence | 1 mm | 2 mm |
| ML14 | Jaw position accuracy | 1 mm | 2 mm |
| ML15 | Backup jaw position accuracy (Elekta) | 1 mm | 2 mm |
| ML16 | MLC leaf position accuracy | 1 mm | 2 mm |
| ML17 | Dynamic leaf position accuracy (picket fence) | 0.5 mm | 1 mm |
| ML18 | Dynamic MLC fluence delivery | 95% ≤ 3%/3 mm | 95% ≤ 5%/3 mm |
| ML19 | Variation in dose rate, gantry speed, MLC leaf speed and position during arc delivery | See note: ML19 | See note: ML19 |
| ML20 | Records | Complete | |
Annual quality control tests
| Designator | Test | Performance | |
|---|---|---|---|
| Tolerance | Action | ||
| Annual | |||
| AL1 | Profile reproducibility | 2% | 3% |
| AL2 | Depth dose reproducibility | 1% | 2% |
| AL3 | Reference dosimetry | 1% | 2% |
| AL4 | Relative output factor reproducibility | 1% | 2% |
| AL5 | Wedge transmission factor reproducibility | 1% | 2% |
| AL6 | Accessory transmission factor reproducibility | 1% | 2% |
| AL7 | Wedge profile reproducibility | 1% | 2% |
| AL8 | Profile and output reproducibility versus gantry angle | 1% | 2% |
| AL9 | Monitor chamber linearity | 1%/1 MU | 2%/2 MU |
| AL10 | End monitor effect | 0.5 MU | 1 MU |
| AL11 | Collimator rotation isocenter (radiation) | 1 mm | 2 mm |
| AL12 | Gantry rotation isocenter (mechanical and radiation) | 1 mm | 2 mm |
| AL13 | Couch rotation isocenter (radiation) | 1 mm | 2 mm |
| AL14 | Coincidence of radiation and mechanical isocenters | 1 mm | 2 mm |
| AL15 | Coincidence of axes of rotation | 1 mm | 2 mm |
| AL16 | Couch deflection | 3 mm | 5 mm |
| AL17 | Leaf transmission (all energies) | 1% | 2% |
| AL18 | Leakage between leaves (all energies) | 2% | 3% |
| AL19 | Transmission through abutting leaves | 2% | 3% |
| A20 | MLC leaf alignment with jaws | 0.5° | 1° |
| A21 | Dosimetric leaf gap | 0.2 mm | 0.3 mm |
| AL22 | Independent quality control review | Complete | |
| DL1 | This test establishes that motion‐enabling features on the linac (e.g., those that allow the gantry to rotate only under desired conditions) are operational. These include functionality tests of couch and hand‐pendant controls and the proper engagement of collision interlocks when touch guards are engaged |
| DL2 | A functional test is performed to establish that brakes on the treatment couch engage when desired and prevent the couch from floating freely or moving when a small force is applied |
| DL3 | This test demonstrates (when applicable): the key interlock prevents the linac from irradiating; the nonemergency beam interruption system stops the beam; upon beam interruption the monitor unitis recorded are in agreement with the backup monitor unit counter and/or timer if applicable |
| DL4 | This test establishes the alignment of crosshairs with appropriate lasers are within the specified limits |
| DL5 | At gantry angle 0°, the test demonstrates that the optical distance indicator identifies the isocenter plane within the specified limits |
| DL6 | This test verifies the performance accuracy of the optical back pointer for applicable units |
| DL7 | Gantry angle 0°, 100 cm source‐axis distance (SAD). This test demonstrates the field edges are accurately defined by jaws and/or MLC leaves. It is sufficient to confirm a predefined field shape using the projected light field at isocenter. Tolerance and action levels apply to each edge of a rectangular field at isocenter as defined by the jaws/MLC leaves. Note that systems with a tertiary collimation MLC system will require both jaw and MLC leaf positions to be verified |
| DL8 | Output constancy must be verified for all photon energies in use on the particular treatment day. Measurement is to be conducted using standard local geometry using a dosimetry system calibrated against the local secondary standard system |
| DL9 | Output constancy must be verified for all electron energies in use on the particular treatment day. Measurement is to be conducted using standard local geometry using a dosimetry system calibrated against the local secondary standard system |
| DL10 | Wedge factors for a representative set of dynamic or virtual soft wedges in use on a particular treatment day must be verified. Machine design characteristics must be considered when determining the representative set. Alternatively, a test cycle designed to test the full range of wedges over multiple days may be considered. Daily wedge factors for universal wedges are required to ensure functionality and position reproducibility |
| ML1 | Verify the functionality of latching interlocks (includes verification that electron beams cannot be turned on unless the MLC leaves are retracted) |
| ML2 | Verify the physical integrity and centering of accessories, including wedges, trays, and cones, as appropriate |
| ML3 | The accuracy of the digital and mechanical (if used clinically) gantry angle readouts must be verified for at least 0°, 90°, 180°, and 270°. The coordinate system convention should also be verified |
| ML4 | The accuracy of the digital and mechanical (if used clinically) collimator angle readouts must be verified for at least 0°, 90°, 180°, and 270°. The coordinate system convention should also be verified |
| ML5 | This test establishes the correct centering of the crosshairs as well as the mechanical axis of rotation of the collimator. Tolerance and action levels refer to the maximum diameter of the mechanical isocenter and the maximum displacement of the crosshairs projection from the center of the mechanical isocenter circle |
| ML6 | Mechanical and digital couch position readouts must be verified over an appropriate clinical range in the directions of the three cardinal axes. Also verify coordinate system convention |
| ML7 | Isocentric rotation of the couch about the collimator rotation axis must be verified. Similar to ML5, the tolerance and action levels refer to the maximum displacement of crosshairs projection from the initial position in the isocenter plane |
| ML8 | Mechanical and digital couch isocentric rotation angle readouts must be verified over the applicable clinical range. Also verify coordinate system convention |
| ML9 | A mechanical device, calibrated against the true radiation isocenter, is used to provide the base reading for the check of the optical distance indicator. The standards stated in the Table apply at the isocenter. The optical distance indicator should be checked over a clinically relevant range of source‐to‐skin distances (SSDs) and gantry angles. The tolerance and action levels may be twice as large (i.e., 2 mm and 4 mm) at the clinical limits of the optical distance indicator's range |
| ML10 | Using a dosimetry system calibrated against the local secondary standard, the output of all clinical beams is checked against yearly reference dosimetry |
| ML11 | Measurements are made to confirm that the depth dose has not changed since commissioning the unit. Tolerance and action levels are specified in percentages for photon beams and in millimetres for electron beams. A single ratio of doses taken at clinically relevant depths is sufficient for these measurements. Alternatively, a tissue‐phantom ratio (TPR) measurement or a check of profile constancy at a shallow depth could be used, and the tolerance and action levels adjusted appropriately |
| ML12 | This test replaces testing of flatness and symmetry and is intended to be consistent with the testing suggested in American Association of Physicists in Medicine (AAPM) protocol TG‐142. |
| ML13 | Geometric alignment of the radiation and optical field edges must be established over a range of field sizes. Tolerance and action levels apply to each edge of a rectangular field |
| ML14 | Accuracy of the radiation field edge of the jaw must be established over a range of jaw positions. The number of positions tested shall be determined from the jaw calibration method. In conjunction with this test it is important to establish acceptable dose profiles for abutting fields at the 0 position. Here, the 2 mm action level for each jaw is generally not sufficient since in principle, abutting fields could have a difference of up to 4 mm between field edges, which can lead to unacceptable peaks or valleys in dose distributions. A tolerance of 5% and an action level of 10% in dose profile deviations for abutting fields are suggested |
| ML15 | Accuracy of the radiation field edge of the backup jaw must be established over a range of positions, if applicable. The number of positions tested shall be determined from the jaw calibration method |
| ML16 | Accuracy of the radiation field edge of the MLC leaf edges must be established over a range of MLC positions. The number of MLC positions tested shall be determined from the MLC calibration method. For some MLC designs this test may be accomplished by evaluating the radiation position of each leaf relative to a reference leaf |
| ML17 | For dynamic MLC IMRT, leaf gap accuracy for all leaf pairs is verified via inspection of a two‐dimensional dose map of a picket fence pattern delivered at gantry angle of 0° |
| ML18 | Specific to IMRT, this test demonstrates that the interplay of leaf velocity, gap width, gap position, and beam holds combine to deliver a planar dose map consistent with the prediction of the treatment planning system. A test plan should consider extreme conditions (e.g., the highest levels of modulation used clinically for each leaf pair). An acceptable alternative to this test is the regular (more than once per month) measurement of patient‐specific, dynamic MLC IMRT fields. Tolerance and action levels are defined via the gamma metric comparing dose map differences (plan versus measurement). Dose maps are defined with region of interest threshold of 10% of the maximum dose. Dose differences are global (i.e., with respect to maximum dose). |
| ML19 | The synchronicity of all dynamic parameters during arc delivery is verified. Parameters may be evaluated independently, using a subset of the tests described by Ling et al. |
| ML20 | Documentation relating to the daily quality control checks, preventive maintenance, service calls, and subsequent checks must be complete, legible, and the operator identified |
| AL1 | This test establishes that an appropriate subset of the cross‐plane and in‐plane profiles at gantry angle 0° are consistent with water‐tank measurements made at the time of commissioning. Tolerance and action levels refer to differences from commissioning or baseline. Measurements should be made for all clinically operable beams. |
| AL2 | Depth dose scans necessary for calibration protocols (alternatively TPR measurements) are also made and used to verify consistency with commissioning/baseline water‐tank measurements. Tolerance and action levels refer to differences from commissioning or baseline. Measurements should be made for all clinically operable beams |
| AL3 | A full absolute dosimetry output calibration based on an internationally accepted protocol (e.g., AAPM TG‐51) |
| AL4 | An appropriate subset of relative output factors are confirmed to be consistent with commissioning measurements |
| AL5 | The wedge transmission factors (if applicable) are confirmed to be consistent with commissioning measurements |
| AL6 | Transmission factors are confirmed to be consistent with commissioning measurements. Discretion may be used. Devices where the physical composition/dimension can be confirmed not to have changed since a previous measurement need not be measured again |
| AL7 | This test applies to moving jaw (dynamic and virtual) and universal (Elekta) wedges. This test confirms that wedged fields produce profiles that are consistent with baseline data through the central 80% of the field for all clinically used wedge angles |
| AL8 | This test establishes the independence of output with gantry angle. It requires that output be measured under identical conditions (e.g., dosimeter under the same amount of buildup material in each position) and that the difference from the gantry at 0° position be within the specified limits. In addition to central axis output, beam profiles must be measured at three cardinal gantry angles: 0°, 90°, and 270°. Measurements should be made for all clinically operable beams |
| AL9, 10 | From a series of radiation measurements with different monitor units the linearity and the end monitor effect are determined. |
| AL11 | Commonly measured using a star shot technique, |
| AL12 | This test determines the diameter of both the mechanical and the radiation isocenter defined by gantry rotation through its full clinical range of motion. Each diameter must be within specifications |
| AL13 | This test determines the diameter of the radiation isocenter defined by couch rotation through its full clinical range of motion. The diameter must be within specifications |
| AL14 | The coincidence of radiation and mechanical isocenters is established for the collimator, gantry and couch. Coincidence must meet the specified limits |
| AL15 | The three axes of rotation (the collimator/MLC, the couch, and the gantry) must meet within a sphere of the specified diameter. |
| AL16 | Couch deflection is measured as a difference in surface position (load versus no load) of the couch extended longitudinally at least 30 cm through isocenter. Under “load” is considered as a typical patient mass (approximately 70 kg) distributed over the couch or placed at the center. The difference is the couch deflection. Tolerance and action levels are defined relative to the deflection measured at the time of commissioning |
| AL17 | The average and maximum MLC leaf transmission is verified in this test for all photon energies and compared with the values established at the time of commissioning or the values adopted in the treatment planning system. Tolerance and action levels refer to changes from the commissioning measurements |
| AL18 | The average and maximum leakage between adjacent, closed MLC leaves is verified in this test for all photon energies and compared with the values established at the time of commissioning or the values adopted in the treatment planning system. Tolerance and action levels refer to changes from the commissioning measurements |
| AL19 | The average and maximum leakage between abutting closed MLC leaves is verified in this test for all photon energies and compared with the values established at the time of commissioning or the values adopted in the treatment planning system. Tolerance and action levels refer to changes from the commissioning measurements |
| AL20 | Use a leaf pattern where one leaf from each leaf bank protrudes well into the field. Confirm the leaf edge parallelism with the collimator or solid jaw edge |
| AL21 | A dynamic leaf gap test (sometimes referred to as a dosimetric leaf gap test) is performed to confirm consistency with baseline measurements. The minimum standard is to establish this using a single detector (e.g., an ion chamber) method, although methods that calculate separate factors for each leaf pair may be employed. The value must be consistent within tolerance for all four cardinal gantry angles |
| AL22 | To ensure redundancy and adequate monitoring, a second qualified medical physicist must independently verify the implementation, analysis, and interpretation of the quality control tests at least annually |