Literature DB >> 28643751

Quick, efficient and effective patient-specific intensity-modulated radiation therapy quality assurance using log file and electronic portal imaging device.

Rajesh Kumar1, H I Amols2, M Lovelock3, S D Sharma1, D Datta1.   

Abstract

AIM: The aim of work is to explore a quick, efficient, and effective patient-specific intensity-modulated radiation therapy (IMRT) quality assurance (QA).
MATERIALS AND METHODS: Software tools were developed to extract and analyze the multi-leaf collimator (MLC) leaf positions (LPs) from electronic portal imaging device (EPID) images for Varian C-series machine and TrueBeam, to extract useful data from MLC log file of C-series linear accelerator (LINAC), to extract useful information from the trajectory log binary file of TrueBeam LINAC, to compare LPs derived from EPID images with log file/trajectory log data, and to analyze IMRT treatment files using the MATLAB programming language. The difference in LP determined from the trajectory log and EPID images was proposed for patient-specific QA.
RESULTS: It was found that the differences in LP for regular radiation fields generated using stationary leaves are <0.5 mm for all the field sizes while for regular radiation fields generated using the moving leaves are more but <2 mm. The differences in LPs for IMRT field were also determined and found to be <2 mm.
CONCLUSIONS: The methodology demonstrated can be used for establishing the accuracy of trajectory log data and for independent routine IMRT QA by generating single number like gamma index to indicate pass or fail of an IMRT treatment plan. The QA indices such as numbers of occurrences of ≥2 mm error in LPS are found more than 5% of total number of occurrences; the dosimetric review of planned treatment is advisable.

Entities:  

Mesh:

Year:  2017        PMID: 28643751     DOI: 10.4103/jcrt.JCRT_1045_16

Source DB:  PubMed          Journal:  J Cancer Res Ther        ISSN: 1998-4138            Impact factor:   1.805


  2 in total

1.  Trajectory log file sensitivity: A critical analysis using DVH and EPID.

Authors:  Wui Ann Woon; Paul B Ravindran; Piyasiri Ekayanake; Subramani Vikraman; Siti Amirah; Yivonne Y F Lim; Christopher H S Vun; Jamsari Khalid
Journal:  Rep Pract Oncol Radiother       Date:  2018-08-13

2.  Insensitivity of machine log files to MLC leaf backlash and effect of MLC backlash on clinical dynamic MLC motion: An experimental investigation.

Authors:  Michael Barnes; Dennis Pomare; Marcus Doebrich; Therese S Standen; Joshua Wolf; Peter Greer; John Simpson
Journal:  J Appl Clin Med Phys       Date:  2022-06-09       Impact factor: 2.243

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.