Evaluation of a new IOLMaster algorithm to measure axial length.

Warren Hill1, Raymund Angeles, Todd Otani.   

Abstract

PURPOSE: To evaluate the ability of the new IOLMaster with Advanced Technology version 5 software (Carl Zeiss Meditec, AG) to measure axial length (AL) in patients having preoperative cataract extraction evaluation.
SETTING: Private practice, Arizona, USA.
METHODS: This single-center study comprised 54 eyes (33 patients) scheduled for cataract extraction. Visual acuity, manifest refraction, and Lens Opacities Classification System III (LOCS III) scores were recorded. Twenty AL measurements per eye were acquired using an IOLMaster with Advanced Technology version 5 software. The AL measurements were analyzed and classified into 4 methods: standard, standard manipulated, composite-5, and composite-20.
RESULTS: Axial length measurements were successfully obtained in 55.6%, 94.4%, 92.6%, and 96.3% of eyes with the standard, standard manipulated, composite-5, and composite-20 methods, respectively. Axial length measurement availability (ability of the software to provide an AL measurement) was not statistically significantly affected by nuclear color, nuclear opalescence, or cortical changes with any method. However, posterior subcapsular changes statistically significantly affected the standard and composite methods (P< .05) and marginally affected the standard manipulated method in measuring the AL (P = .0868). Axial length availability was significantly reduced in eyes with posterior subcapsular cataract of LOCS III grade 5.0 or higher.
CONCLUSIONS: The 2 composite methods performed as well as the more time consuming, complicated standard manipulated method in measuring AL. More than 90% of eyes could be measured by the composite methods. Dense posterior subcapsular cataracts significantly reduced the ability of the composite and standard manipulated methods to successfully measure AL.

Entities:  

Mesh:

Year:  2008        PMID: 18498996     DOI: 10.1016/j.jcrs.2008.02.021

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  13 in total

Review 1.  Optimising biometry for best outcomes in cataract surgery.

Authors:  R Sheard
Journal:  Eye (Lond)       Date:  2013-12-06       Impact factor: 3.775

2.  Comparison of acquisition rate and agreement of axial length with two swept-source optical coherence tomographers and a partial coherence interferometer.

Authors:  Shi-Ming Cheng; Wen-Tao Yan; Jia-Sheng Zhang; Tian-Tian Li; Xi Li; A-Yong Yu
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2022-04-30       Impact factor: 3.535

3.  Comparison axial length measurements from three biometric instruments in high myopia.

Authors:  Xiao-Gang Wang; Jing Dong; Yu-Lan Pu; Hui-Jun Liu; Qiang Wu
Journal:  Int J Ophthalmol       Date:  2016-06-18       Impact factor: 1.779

4.  Reproducibility of a long-range swept-source optical coherence tomography ocular biometry system and comparison with clinical biometers.

Authors:  Ireneusz Grulkowski; Jonathan J Liu; Jason Y Zhang; Benjamin Potsaid; Vijaysekhar Jayaraman; Alex E Cable; Jay S Duker; James G Fujimoto
Journal:  Ophthalmology       Date:  2013-06-04       Impact factor: 12.079

5.  Axial Length Measurement Failure Rates with the IOLMaster and Lenstar LS 900 in Eyes with Cataract.

Authors:  Colm McAlinden; Qinmei Wang; Konrad Pesudovs; Xin Yang; Fangjun Bao; Ayong Yu; Shishi Lin; Yifan Feng; Jinhai Huang
Journal:  PLoS One       Date:  2015-06-10       Impact factor: 3.240

Review 6.  Optical correction of aphakia in children.

Authors:  Alireza Baradaran-Rafii; Ebrahim Shirzadeh; Medi Eslani; Mitra Akbari
Journal:  J Ophthalmic Vis Res       Date:  2014-01

7.  Comparison of ocular biometry measurements by applanation and immersion A-scan techniques.

Authors:  Dupe S Ademola-Popoola; Donald A Nzeh; Sadiat E Saka; Lateefat B Olokoba; Tokunbo S Obajolowo
Journal:  J Curr Ophthalmol       Date:  2016-02-09

8.  Comparison of Anterior Segment Measurements with Optical Low-coherence Reflectometry and Partial-coherence Interferometry Optical Biometers.

Authors:  Ertuğrul Can; Mustafa Duran; Tuğba Çetinkaya; Nurşen Arıtürk
Journal:  Middle East Afr J Ophthalmol       Date:  2016 Oct-Dec

9.  Axial length measurement acquisition rates of two optical biometers in cataractous eyes.

Authors:  Alice Epitropoulos
Journal:  Clin Ophthalmol       Date:  2014-07-22

10.  Evaluation and comparison of the new swept source OCT-based IOLMaster 700 with the IOLMaster 500.

Authors:  Ahmet Akman; Leyla Asena; Sirel Gür Güngör
Journal:  Br J Ophthalmol       Date:  2015-12-16       Impact factor: 4.638

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.