Andre Luiz Ferreira Costa1, Karolina A C Fardim2, Jennifer M Mantoani2, Ana Lucia Franco Ricardo1, Maria Aparecida N Jardini2, Kaan Orhan3, Sérgio Lúcio Pereira de Castro Lopes2.
Abstract
Objective: To evaluate a postprocessing filter of a new imaging-processing software for analysis of metal artifact reduction.Entities:
Mesh:
Substances:
Year: 2022 PMID: 35295958 PMCID: PMC8920685 DOI: 10.1155/2022/1362473
Source DB: PubMed Journal: Biomed Res Int Impact factor: 3.411
Figure 1Schematic figure showing phantom positioned on the paper template with laser light markers.
Figure 2Selection of axial sections from CBCT images.
Figure 3Example of artifacts generated by titanium and zirconium dioxide implants in images of teeth with application of the four filters and unfiltered images.
Quantification mean (standard deviation) values of the artifact with titanium and zirconium dioxide implants.
| Titanium implant | Zirconium dioxide | ||||
|---|---|---|---|---|---|
| Filter | Mean | SD | Filter | Mean | SD |
| 43 | 43 | ||||
| Filter 1x | 6.4 | 0.3 | Filter 1x | 9.2 | 0.0 |
| Filter 2x | 6.3 | 0.2 | Filter 2x | 9.1 | 0.1 |
| Filter BAR | 5.7 | 0.0 | Filter BAR | 7.4 | 0.1 |
| Multi-CDT NR | 6.0 | 0.0 | Multi-CDT NR | 8.4 | 0.1 |
| Nonfiltered | 6.7 | 0.0 | Nonfiltered | 9.5 | 0.0 |
| 41 | 41 | ||||
| Filter 1x | 6.6 | 0.1 | Filter 1x | 9.3 | 0.2 |
| Filter 2x | 6.5 | 0.2 | Filter 2x | 9.2 | 0.0 |
| Filter BAR | 5.7 | 0.1 | Filter BAR | 6.7 | 0.3 |
| Multi-CDT NR | 6.3 | 0.1 | Multi-CDT NR | 7.6 | 0.0 |
| Nonfiltered | 6.9 | 0.2 | Nonfiltered | 9.8 | 0.1 |
| 44 | 44 | ||||
| Filter 1x | 6.6 | 0.2 | Filter 1x | 9.2 | 0.2 |
| Filter 2x | 6.5 | 0.2 | Filter 2x | 8.9 | 0.2 |
| Filter BAR | 5.1 | 0.0 | Filter BAR | 7.2 | 0.1 |
| Multi-CDT NR | 5.4 | 0.1 | Multi-CDT NR | 8.4 | 0.2 |
| Nonfiltered | 6.5 | 0.1 | Nonfiltered | 9.1 | 0.0 |
| 46 | 46 | ||||
| Filter 1x | 6.4 | 0.3 | Filter 1x | 8.9 | 0.0 |
| Filter 2x | 6.0 | 0.2 | Filter 2x | 8.8 | 0.0 |
| Filter BAR | 5.1 | 0.1 | Filter BAR | 6.6 | 0.2 |
| Multi-CDT NR | 5.9 | 0.2 | Multi-CDT NR | 7.2 | 0.1 |
| Nonfiltered | 6.2 | 0.2 | Nonfiltered | 9.1 | 0.2 |
N: number of scans; SD: standard deviation.
Comparison between filters/teeth for titanium and zirconium dioxide implant.
| Titanium implant | Zirconium dioxide | ||||||
|---|---|---|---|---|---|---|---|
| Filter | Median | Result∗ | SD | Filter | Median | Result∗ | SD |
| 41 | 41 | ||||||
| BAR | 7.4 | C | 0.1 | BAR | 6.7 | D | 0.3 |
| Multi-CDT NR | 8.4 | B | 0.1 | Multi-CDT NR | 7.6 | C | 0.0 |
| 2x | 9.1 | AB | 0.2 | 2x | 9.2 | B | 0.0 |
| 1x | 9.2 | AB | 0.1 | 1x | 9.3 | B | 0.2 |
| Nonfiltered | 9.5 | A | 0.2 | Nonfiltered | 9.8 | A | 0.1 |
| 43 | 43 | ||||||
| BAR | 5.7 | D | 0.0 | BAR | 7.4 | D | 0.1 |
| Multi-CDT NR | 6.0 | CD | 0.0 | Multi-CDT NR | 8.4 | C | 0.1 |
| 2x | 6.3 | BC | 0.2 | 2x | 9.1 | B | 0.1 |
| 1x | 6.4 | AB | 0.3 | 1x | 9.2 | AB | 0.0 |
| Nonfiltered | 6.7 | A | 0.0 | Nonfiltered | 9.5 | A | 0.0 |
| 44 | 44 | ||||||
| BAR | 6.7 | B | 0.0 | BAR | 7.2 | C | 0.1 |
| Multi-CDT NR | 7.6 | B | 0.1 | Multi-CDT NR | 8.4 | B | 0.2 |
| 2x | 9.2 | A | 0.2 | 2x | 8.9 | A | 0.2 |
| Nonfiltered | 9.3 | A | 0.1 | Nonfiltered | 9.1 | A | 0.0 |
| 1x | 9.8 | A | 0.2 | 1x | 9.2 | A | 0.2 |
| 46 | 46 | ||||||
| BAR | 5.1 | D | 0.1 | BAR | 6.6 | C | 0.2 |
| Multi-CDT NR | 5.9 | C | 0.1 | Multi-CDT NR | 7.2 | B | 0.1 |
| 2x | 6.0 | BC | 0.2 | 2x | 8.8 | A | 0.0 |
| Nonfiltered | 6.2 | AB | 0.1 | 1x | 8.9 | A | 0.0 |
| 1x | 6.4 | A | 0.3 | Nonfiltered | 9.1 | A | 0.2 |
∗Medians followed by the same letters do not differ significantly from each other.