Beam hardening artefacts occur in dental implant scans with the NewTom cone beam CT but not with the dental 4-row multidetector CT.

F G Draenert1, E Coppenrath, P Herzog, S Müller, U G Mueller-Lisse.   

Abstract

OBJECTIVES: The aim of this study was to examine beam hardening artefacts of the NewTom 9000 cone-beam computed tomography (CBCT) device compared with the Philips MX 8000 (4-row multidetector CT (MDCT)).
METHODS: We modified a SawBone skull to become a standardized model for our study. The skull was used for scans with the NewTom 9000 CBCT device and a standard dental multi-detector CT (MDCT) at a comparable reconstruction resolution with a standard Straumann ITI 4.1 mm implant in four implant positions in the maxilla (first permanent premolar in the right maxilla region, second permanent molar in the right maxilla region, first permanent premolar in the left maxilla region and second permanent molar in the left maxilla region). Results were compared with construction data of the dental implant. An image quality assessment of the images from both devices was performed with four experienced physicians and statistically analysed with the two-tailed Wilcoxon test.
RESULTS: Scans with the NewTom 9000 CBCT showed strong beam hardening artefacts in the form of a radiation beam shadow in all reconstructions compared with the MDCT. These imaging artefacts became stronger with greater distance from the centre of the scanned volume. These differences in the imaging quality were proved as significant in a quality evaluation by four experienced physicians (P<0.05).
CONCLUSIONS: Visual spatial resolution of the NewTom 9000 CBCT was less accurate than the Philips MX 8000 MDCT in the imaging of metallic dental implants.

Mesh:

Substances:

Year:  2007        PMID: 17536086     DOI: 10.1259/dmfr/32579161

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  45 in total

1.  Evaluation of a cone beam CT artefact reduction algorithm.

Authors:  B Bechara; C A McMahan; H Geha; M Noujeim
Journal:  Dentomaxillofac Radiol       Date:  2012-02-23       Impact factor: 2.419

2.  A comparative study of accuracy of detection of surface osseous changes in the temporomandibular joint using multidetector CT and cone beam CT.

Authors:  E H Zain-Alabdeen; R I Alsadhan
Journal:  Dentomaxillofac Radiol       Date:  2012-03       Impact factor: 2.419

3.  Accuracy of CBCT measurements of a human skull.

Authors:  Kivanç Kamburoğlu; Eray Kolsuz; Hakan Kurt; Cenk Kiliç; Tuncer Özen; Candan Semra Paksoy
Journal:  J Digit Imaging       Date:  2011-10       Impact factor: 4.056

4.  Bone density changes around teeth during orthodontic treatment.

Authors:  Jui-Ting Hsu; Hsun-Wen Chang; Heng-Li Huang; Jian-Hong Yu; Yu-Fen Li; Ming-Gene Tu
Journal:  Clin Oral Investig       Date:  2011-08       Impact factor: 3.573

5.  Ex vivo evaluation of new 2D and 3D dental radiographic technology for detecting caries.

Authors:  Laurence Gaalaas; Donald Tyndall; André Mol; Eric T Everett; Ananta Bangdiwala
Journal:  Dentomaxillofac Radiol       Date:  2015-12-16       Impact factor: 2.419

6.  Accuracy of linear intraoral measurements using cone beam CT and multidetector CT: a tale of two CTs.

Authors:  R Patcas; G Markic; L Müller; O Ullrich; T Peltomäki; C J Kellenberger; C A Karlo
Journal:  Dentomaxillofac Radiol       Date:  2012-05-03       Impact factor: 2.419

7.  Assessment of buccal marginal alveolar peri-implant and periodontal defects using a cone beam CT system with and without the application of metal artefact reduction mode.

Authors:  K Kamburoglu; E Kolsuz; S Murat; H Eren; S Yüksel; C S Paksoy
Journal:  Dentomaxillofac Radiol       Date:  2013       Impact factor: 2.419

8.  Influence of object location in different FOVs on trabecular bone microstructure measurements of human mandible: a cone beam CT study.

Authors:  N Ibrahim; A Parsa; B Hassan; P van der Stelt; I H A Aartman; P Nambiar
Journal:  Dentomaxillofac Radiol       Date:  2013-11-21       Impact factor: 2.419

9.  Resorption of labial bone in maxillary anterior implant.

Authors:  Young-Bum Cho; Seung-Jin Moon; Chae-Heon Chung; Hee-Jung Kim
Journal:  J Adv Prosthodont       Date:  2011-06-30       Impact factor: 1.904

10.  Cone beam CT image artefacts related to head motion simulated by a robot skull: visual characteristics and impact on image quality.

Authors:  R Spin-Neto; J Mudrak; L H Matzen; J Christensen; E Gotfredsen; A Wenzel
Journal:  Dentomaxillofac Radiol       Date:  2012-07-27       Impact factor: 2.419

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.