| Literature DB >> 35204474 |
Kasama Sukapirom1,2, Somrat Matchua3, Charin Thepthai4, Narinee Srimark1, Ladawan Khowawisetsut5, Kovit Pattanapanyasat1,2.
Abstract
HIV viral load is more reliable tool for monitoring treatment throughout the course of HIV/AIDS, but the test may be expensive in resource-limited settings. Therefore, enumeration of CD4 T-lymphocyte count remains important in these settings. This study evaluated the performance of BDFACSPresto, a near-patient CD4 counter planned to be used in primary healthcare clinics in Thailand. Results of percent, absolute CD4 count and hemoglobin (Hb) on the FACSPresto were compared with the TriTEST/TruCOUNT/BDFACSCalibur method and a Sysmex hematology analyzer. Phase I of the study was performed in an ISO15189 laboratory. Both percentage and absolute values showed Passing-Bablok slopes within 0.98-1.06 and 0.97-1.13, mean Bland-Altman biases of +1.2% and +20.5 cells/µL, respectively. In phase II, venous and some capillary blood samples were analyzed in four primary healthcare clinics. The results showed good correlation between capillary and venous blood. For venous blood samples, regression lines showed slopes of 1.01-1.05 and 1.01-1.07 for all percentage and absolute values. The overall mean biases were +0.9% and +17.0 cells/µL. For Hb, Passing-Bablok regression result gave slope within 1.01-1.07 and mean bias of -0.06 g/dL. Thus, CD4 enumeration in blood by the FACSPresto is reliable and can be performed to an identical standard at primary healthcare clinics.Entities:
Keywords: AIDS; CD4; FACSPresto; HIV; flow cytometry; point-of-care testing
Year: 2022 PMID: 35204474 PMCID: PMC8871446 DOI: 10.3390/diagnostics12020382
Source DB: PubMed Journal: Diagnostics (Basel) ISSN: 2075-4418
Reproducibility of the BD FACSCalibur and the BD FACSPresto system determined by analysis of 8 replicates from one fresh HIV-1 infected blood sample and 8 replicates from a stabilized CD-Check Plus Low blood.
| Variation | Mean % CV ± SD | |
|---|---|---|
| FACSCalibur/TriTest/TruCount | FACSPresto | |
| Within-run | 2.6 ± 1.2 | 3.8 ± 2.0 |
| Between-run | 7.8 ± 3.7 | 8.3 ± 4.6 |
Comparison analysis of percent and absolute CD4 T-lymphocyte values between the standard TriTEST/TruCount BD FACSCalibur and the BD FACSPresto near-patient CD4 counter at the ISO15189-accredited clinical laboratory by using Passing and Bablok regression and Bland–Altman mean bias analysis.
| Parameter | CD4 (%) | Absolute CD4 (cells/μL) |
|---|---|---|
| Regression | y = 0.75 + 1.02x | y = 0.92 + 1.07x |
| 95% * CI for intercept | +0.06 to 1.82 | −32.17 to 36.20 |
| 95% CI for slope | 0.98 to 1.06 | 0.97 to 1.13 |
| Bland–Altman mean bias | +1.2 | +20.5 |
| ** LOA (mean difference ± 1.96SD) | 1.0 to 3.3% | −136.1 to +177.2 |
* CI; Confidence Interval; ** LOA; Limits of Agreement.
Comparison analysis of percent and absolute CD4 T-lymphocyte values between capillary and venous blood of the same 66 HIV-infected blood samples on BD FACSPresto near-patient CD4 counter at the primary healthcare clinics, and on BD FACSCalibur at the ISO15189-accredited clinical laboratory by using Passing and Bablok regression and Bland–Altman mean bias analysis.
|
| ||
|
|
|
|
| Mean | ||
| Venous blood | 28.23 ± 8.18% | 874 ± 617 |
| Capillary blood | 27.05 ± 8.39% | 928 ± 656 |
| Regression | y = −1.34 + 1.01x | y = −15.84 + 1.91x |
| 95% * CI for intercept | −2.67 to 0.03 | −48.77 to 9.46 |
| 95% CI for slope | 0.97 to 1.06 | 1.05 to 1.14 |
| Bland–Altman mean bias | −1.2% | +54 |
| ** LOA (mean difference ± 1.96SD) | −4.6 to 2.2% | −148.3 to +256.2 |
|
| ||
|
|
|
|
| Regression | y = −0.62 + 0.97x | y = −34.41 + 1.09x |
| 95% * CI for intercept | −2.11 to 0.57 | −73.26 to 4.82 |
| 95% CI for slope | 0.93 to 1.04 | 1.03 to 1.15 |
| Bland–Altman mean bias | −1.2% | +40.4 |
| ** LOA (mean difference ± 1.96SD) | −5.6 to 3.2% | −273.6 to +354.4 |
* CI; Confidence Interval; ** LOA; Limits of Agreement.
Figure 1(a,b) Passing and Bablok regression analyses; (c,d) Bland–Altman bias plots; and (e,f) percent similarity plots of percent CD4 T-lymphocytes (top panel) and absolute CD4 T-lymphocyte counts (lower panel) of all 180 HIV-1 infected blood samples between the standard TriTEST/TruCount BD FACSCalibur and the BD FACSPresto near-patient CD4 counter. Passing and Bablok plots showing regression line of measured data and agreement of fitted regression line and identity line. In Bland–Altman bias plots, the horizontal lines at the center indicate the mean bias; the lower and upper lines represent LOA ± 1.96SD.
Figure 2(a) Passing and Bablok regression analysis; (b) Bland–Altman bias plot; and (c) percent similarity plot of Hb measurements from BD FACSPresto near-patient CD4 counter and Sysmex XT2000i hematology analyzer. Passing and Bablok plot showing regression line of measured data and agreement of fitted regression line and identity line. In Bland–Altman bias plot, the horizontal line at the center indicates the mean bias; the lower and upper lines represent LOA ± 1.96SD.
Figure 3(a,c) Bland–Altman bias plots and (b,d) percent similarity plots of absolute CD4 T-lymphocyte counts of <500 cells/µL (top panel) and absolute CD4 T-lymphocyte counts > 500 cells/µL (lower panel) of HIV-1 infected blood samples between the standard TriTEST/TruCount BD FACSCalibur and the BD FACSPresto near-patient CD4 counter. In the Bland–Altman bias plots, the horizontal lines at the center indicate the mean bias; the lower and upper lines represent LOA ± 1.96SD.
Figure 4(a,b) Passing and Bablok regression analysis; (c,d) Bland–Altman bias plot and (e,f) percent similarity plot of percent CD4 T-lymphocytes (top panel) and absolute CD4 T-lymphocyte counts (lower panel) at the threshold of 200 cells/µL of HIV-1 infected blood samples between the standard TriTEST/TruCount BD FACSCalibur and the BD FACSPresto near-patient CD4 counter. Passing and Bablok plot showing regression line of measured data and agreement of fitted regression line and identity line. In Bland–Altman bias plot, the horizontal line at the center indicates the mean bias; the lower and upper lines represent LOA ± 1.96SD.
Comparison analysis of values from the present study with previously published data for the performance evaluation of BD FACSPresto near-patient CD4 counter with the standard flow cytometer using venous blood samples.
| Parameter | Sample Size | Flow Cytometer | Method | Bland–Altman | |
|---|---|---|---|---|---|
| % CD4 | Absolute CD4 (cells/µL) | ||||
| Present study | 180 | BD FACSCalibur | TriTEST/ | +0.9 | +17.0 |
| South Africa [ | 214 | BC Coulter XL | FlowCare PanLeuco-gate | +1.41 | +40.39 |
| Kenya/USA [ | 189 | BD FACSCalibur | TriTEST/ | +3.19 | −0.78 |
| Tanzania/Belgium [ | 200 | BD FACSCalibur | TriTEST/ | +0.6 | +2.5 |
| India/China/Kenya/USA/Thailand [ | 720 | BD FACSCalibur | TriTEST/ | +0.75 | +3.10 |
| Ghana [ | 53 | BD FACSCount | new CD4 reagent | +6.9 | +44 |