Janahiraman Divyalakshmi1, K Mohamathu Rafic1,2, Ebenezer Suman Babu1,2, Timothy Peace Balasingh1, Amalan Sebastin1, Christopher J Sujith1, L Jose Solomon Raj1.
Abstract
BACKGROUND: The aim of the study was to investigate the dosimetric feasibility of using optically stimulated luminescence dosimeters (OSLD) and an electronic portal imaging device (EPID) for central axis (CA X) and off-axis (OAX) dosimetric leaf gap (DLG) measurement.Entities:
Keywords: dosimetric leaf gap; electronic portal imaging device; optically stimulated luminescence dosimeter; rounded leaf-end transmission
Year: 2021 PMID: 34992876 PMCID: PMC8726435 DOI: 10.5603/RPOR.a2021.0117
Source DB: PubMed Journal: Rep Pract Oncol Radiother ISSN: 1507-1367
Figure 1Schematic of dosimetric leaf gap (DLG) measurement using optically stimulated luminescence dosimeters (OSLD). A — gantry; B — MLC; C — 1D OSLD array; D — 10 cm Perspex Backscatter; E and G — narrow groove and tongue joint in OSLD; F — active OSLD phosphor; SLP — superior leaf pair; CLP — central leaf pair; ILP — inferior leaf pair
Figure 2A workflow algorithm illustrating the generation of 2D DLG map from the portal images of the EPID
Dosimetric leaf gap (DLG) measured at three positions using ionization chamber dosimetry (ICD) and optically stimulated luminescence dosimeters (OSLD) using two normalization methods
| Parameter | Global normalization | Local normalization | ||||
|---|---|---|---|---|---|---|
| Superior | Centre | Inferior | Superior | Centre | Inferior | |
| DLGICD [mm] | 1.27 ± 0.02 | 1.04 ± 0.02 | 1.05 ± 0.01 | 1.28 ± 0.02 | 1.04 ± 0.02 | 1.05 ± 0.01 |
| DLGOSLD [mm] | 1.25 ± 0.07 | 1.08 ± 0.03 | 1.08 ± 0.09 | 1.24 ± 0.04 | 1.08 ± 0.04 | 1.08 ± 0.06 |
DLGEPID measured for ROI corresponding to ionization chamber dosimetry (ICD) and optically stimulated luminescence dosimeters (OSLD)
| Parameter | Superior | Centre | Inferior |
|---|---|---|---|
| DLGEPID, ICD [mm] | 1.28 ± 0.02 (1.06 ± 0.02) | 1.05 ± 0.02 (0.90 ± 0.03) | 1.04 ± 0.02 (0.87 ± 0.03) |
| DLGEPID, OSLD [mm] | 1.28 ± 0.03 (1.09 ± 0.02) | 1.05 ± 0.02 (0.91 ± 0.02) | 1.04 ± 0.01 (0.88 ± 0.01) |
Uncorrected DLGEPID values are given within parenthesis
Figure 3Sliding aperture-specific EPID response correction
Figure 4Analysis of the 2D DLG map. A. The central 7 × 7 cm2 (180 × 180 pixels) region of the 2D DLG map selected for analysis; B. An adjacent leaf pair (width of 26 pixels) consisting of two adjoining MLC leaves (blue regions) that encompass an inter-leaf region in between (yellow region); C. The 2D map portraying the averaged leaf pair-specific DLG values across the field. The 3rd, 4th and 5th leaf pairs correspond to the superior, central and inferior leaf pairs
Figure 5Leaf pair-specific dosimetric leaf gap (DLG) values obtained from the 2D DLGEPID map, optically stimulated luminescence dosimeters (OSLD) and ionization chamber dosimetry (ICD); DLG-EPID WCF & DLG-EPID WoCF — DLGEPID with and without correction factors. 3rd, 4th & 5th MLC leaf pairs correspond to superior, central and inferior leaf pairs
Figure 6DLGEPID profile across the field for CA X and ±1 cm OAX leaf pairs. SLP — superior leaf pair; CLP — central leaf pair; INP — inferior leaf pair