Literature DB >> 21520847

Using an EPID for patient-specific VMAT quality assurance.

M Bakhtiari1, L Kumaraswamy, D W Bailey, S de Boer, H K Malhotra, M B Podgorsak.   

Abstract

PURPOSE: A patient-specific quality assurance (QA) method was developed to verify gantry-specific individual multileaf collimator (MLC) apertures (control points) in volumetric modulated arc therapy (VMAT) plans using an electronic portal imaging device (EPID).
METHODS: VMAT treatment plans were generated in an Eclipse treatment planning system (TPS). DICOM images from a Varian EPID (aS1000) acquired in continuous acquisition mode were used for pretreatment QA. Each cine image file contains the grayscale image of the MLC aperture related to its specific control point and the corresponding gantry angle information. The TPS MLC file of this RapidArc plan contains the leaf positions for all 177 control points (gantry angles). In-house software was developed that interpolates the measured images based on the gantry angle and overlays them with the MLC pattern for all control points. The 38% isointensity line was used to define the edge of the MLC leaves on the portal images. The software generates graphs and tables that provide analysis for the number of mismatched leaf positions for a chosen distance to agreement at each control point and the frequency in which each particular leaf mismatches for the entire arc.
RESULTS: Seven patients plans were analyzed using this method. The leaves with the highest mismatched rate were found to be treatment plan dependent.
CONCLUSIONS: This in-house software can be used to automatically verify the MLC leaf positions for all control points of VMAT plans using cine images acquired by an EPID.

Entities:  

Mesh:

Year:  2011        PMID: 21520847     DOI: 10.1118/1.3552925

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  25 in total

1.  Exit fluence analysis using portal dosimetry in volumetric modulated arc therapy.

Authors:  Prabakar Sukumar; Sriram Padmanaban; Dhanabalan Rajasekaran; Muniyappan Kannan; Vivekanandan Nagarajan
Journal:  Rep Pract Oncol Radiother       Date:  2012-07-15

2.  A comprehensive study of the mechanical performance of gantry, EPID and the MLC assembly in Elekta linacs during gantry rotation.

Authors:  P Rowshanfarzad; H L Riis; S J Zimmermann; M A Ebert
Journal:  Br J Radiol       Date:  2015-04-23       Impact factor: 3.039

3.  Development of an accurate EPID-based output measurement and dosimetric verification tool for electron beam therapy.

Authors:  Aiping Ding; Lei Xing; Bin Han
Journal:  Med Phys       Date:  2015-07       Impact factor: 4.071

4.  Penalization of aperture complexity in inversely planned volumetric modulated arc therapy.

Authors:  Kelly C Younge; Martha M Matuszak; Jean M Moran; Daniel L McShan; Benedick A Fraass; Donald A Roberts
Journal:  Med Phys       Date:  2012-11       Impact factor: 4.071

5.  Dosimetric verification for intensity-modulated arc therapy plans by use of 2D diode array, radiochromic film and radiosensitive polymer gel.

Authors:  Naoki Hayashi; Ryan L Malmin; Yoichi Watanabe
Journal:  J Radiat Res       Date:  2014-01-20       Impact factor: 2.724

6.  Variation in Elekta iView electronic portal imager pixel scale factor with gantry angle, and impact on multi-leaf collimator quality assurance.

Authors:  Simon K Goodall; Craig Norvill
Journal:  J Appl Clin Med Phys       Date:  2022-06-06       Impact factor: 2.243

7.  The impact of cine EPID image acquisition frame rate on markerless soft-tissue tracking.

Authors:  Stephen Yip; Joerg Rottmann; Ross Berbeco
Journal:  Med Phys       Date:  2014-06       Impact factor: 4.071

8.  Clinical experience with machine log file software for volumetric-modulated arc therapy techniques.

Authors:  Luis Alberto Vazquez-Quino; Claudia Ivette Huerta-Hernandez; Dharanipathy Rangaraj
Journal:  Proc (Bayl Univ Med Cent)       Date:  2017-07

9.  Evaluation of two-dimensional electronic portal imaging device using integrated images during volumetric modulated arc therapy for prostate cancer.

Authors:  Shoki Inui; Yoshihiro Ueda; Shunsuke Ono; Shingo Ohira; Masaru Isono; Yuya Nitta; Hikari Ueda; Masayoshi Miyazaki; Masahiko Koizumi; Teruki Teshima
Journal:  Rep Pract Oncol Radiother       Date:  2021-04-14

10.  Clinical translation of a new flat-panel detector for beam's-eye-view imaging.

Authors:  T C Harris; J Seco; D Ferguson; M Lehmann; P Huber; M Shi; M Jacobson; I Valencia Lozano; M Myronakis; P Baturin; R Fueglistaller; D Morf; R Berbeco
Journal:  Phys Med Biol       Date:  2020-12-07       Impact factor: 4.174