Dosimetric impact of different multileaf collimators on prostate intensity modulated treatment planning.

Lucia Clara Orlandini1, Margherita Betti1, Christian Fulcheri1, Marianna Coppola2, Luca Cionini2.   

Abstract

AIM: The main purpose of this study is to perform a dosimetric comparison on target volumes and organs at risks (OARs) between prostate intensity modulated treatment plans (IMRT) optimized with different multileaf collimators (MLCs).
BACKGROUND: The use of MLCs with a small leaf width in the IMRT optimization may improve conformity around the tumor target whilst reducing the dose to normal tissues.
MATERIALS AND METHODS: Two linacs mounting MLCs with 5 and 10 mm leaf-width, respectively, implemented in Pinnacle(3) treatment planning system were used for this work. Nineteen patients with prostate carcinoma undergoing a radiotherapy treatment were enrolled. Treatment planning with different setup arrangements (7 and 5 beams) were performed for each patient and each machine. Dose volume histograms (DVHs) cut-off points were used in the treatment planning comparison.
RESULTS: Comparable planning target volume (PTV) coverage was obtained with 7- and 5-beam configuration (both with 5 and 10 mm MLC leaf-width). The comparison of bladder and rectum DVH cut-off points for the 5-beam arrangement shows that 52.6% of the plans optimized with a larger leaf-width did not satisfy at least one of the OARs' constraints. This percentage is reduced to 10.5% for the smaller leaf-width. If a 7-beam arrangement is used the value of 52.6% decreases to 21.1% while the value of 10.5% remains unchanged.
CONCLUSION: MLCs collimators with different widths and number of leaves lead to a comparable prostate treatment planning if a proper adjustment is made of the number of gantry angles.

Entities:  

Keywords:  Intensity modulated radiation therapy (IMRT); Multileaf collimator; Prostate treatment

Year:  2015        PMID: 26549993      PMCID: PMC4597085          DOI: 10.1016/j.rpor.2015.06.005

Source DB:  PubMed          Journal:  Rep Pract Oncol Radiother        ISSN: 1507-1367


  11 in total

1.  A quantitative evaluation of IMRT dose distributions: refinement and clinical assessment of the gamma evaluation.

Authors:  Tom Depuydt; Ann Van Esch; Dominique Pierre Huyskens
Journal:  Radiother Oncol       Date:  2002-03       Impact factor: 6.280

2.  Evaluation of MatriXX for IMRT and VMAT dose verifications in peripheral dose regions.

Authors:  Zhaohui Han; Sook Kien Ng; Mandar S Bhagwat; Yulia Lyatskaya; Piotr Zygmanski
Journal:  Med Phys       Date:  2010-07       Impact factor: 4.071

3.  The lessons of QUANTEC: recommendations for reporting and gathering data on dose-volume dependencies of treatment outcome.

Authors:  Andrew Jackson; Lawrence B Marks; Søren M Bentzen; Avraham Eisbruch; Ellen D Yorke; Randal K Ten Haken; Louis S Constine; Joseph O Deasy
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-03-01       Impact factor: 7.038

4.  Two-dimensional ionization chamber arrays for IMRT plan verification.

Authors:  Bjoern Poppe; Arne Blechschmidt; Armand Djouguela; Ralf Kollhoff; Antje Rubach; Kay C Willborn; Dietrich Harder
Journal:  Med Phys       Date:  2006-04       Impact factor: 4.071

5.  Comparison of dose distributions in IMRT planning using the gamma function.

Authors:  C Andenna; M Benassi; B Caccia; S Marzi; M Pedrini; C Zicari
Journal:  J Exp Clin Cancer Res       Date:  2006-06

6.  6 MV dosimetric characterization of the 160 MLC, the new Siemens multileaf collimator.

Authors:  Martin B Tacke; Simeon Nill; Peter Häring; Uwe Oelfke
Journal:  Med Phys       Date:  2008-05       Impact factor: 4.071

Review 7.  Evidence behind use of intensity-modulated radiotherapy: a systematic review of comparative clinical studies.

Authors:  Liv Veldeman; Indira Madani; Frank Hulstaert; Gert De Meerleer; Marc Mareel; Wilfried De Neve
Journal:  Lancet Oncol       Date:  2008-04       Impact factor: 41.316

Review 8.  Planning and delivery of intensity-modulated radiation therapy.

Authors:  Cedric X Yu; Christopher J Amies; Michelle Svatos
Journal:  Med Phys       Date:  2008-12       Impact factor: 4.071

9.  Improvement of I'mRT MatriXX in terms of spatial resolution and large area acquisition for patient-specific intensity-modulated radiotherapy verification.

Authors:  Arun S Oinam; Lakhwant Singh; S C Sharma; Pradeep Goswami
Journal:  J Med Phys       Date:  2009-07

10.  Dose linearity and uniformity of Siemens ONCOR impression plus linear accelerator designed for step-and-shoot intensity-modulated radiation therapy.

Authors:  Janhavi R Bhangle; V K Sathiya Narayanan; Shrikant A Deshpande
Journal:  J Med Phys       Date:  2007-07
View more
  3 in total

1.  A comparison of a moderately hypofractionated IMRT planning technique used in a randomised UK external beam radiotherapy trial with an in-house technique for localised prostate cancer.

Authors:  Ian Gleeson
Journal:  Rep Pract Oncol Radiother       Date:  2020-03-19

2.  A strategy to determine off-axis dosimetric leaf gap using OSLD and EPID.

Authors:  Janahiraman Divyalakshmi; K Mohamathu Rafic; Ebenezer Suman Babu; Timothy Peace Balasingh; Amalan Sebastin; Christopher J Sujith; L Jose Solomon Raj
Journal:  Rep Pract Oncol Radiother       Date:  2021-12-30

3.  Dosimetric sensitivity of leaf width on volumetric modulated arc therapy plan quality: an objective approach.

Authors:  Ghulam Murtaza; Muhammad Shamshad; Munir Ahmed; Shahid Mehmood; Ehsan Ullah Khan
Journal:  Rep Pract Oncol Radiother       Date:  2022-03-22
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.