Small field electron beam dosimetry using MOSFET detector.

Md Nurul Amin1, Robert Heaton, Bern Norrlinger, Mohammad K Islam.   

Abstract

The dosimetry of very small electron fields can be challenging due to relative shifts in percent depth-dose curves, including the location of dmax, and lack of lateral electronic equilibrium in an ion chamber when placed in the beam. Conventionally a small parallel plate chamber or film is utilized to perform small field electron beam dosimetry. Since modern radiotherapy departments are becoming filmless in favor of electronic imaging, an alternate and readily available clinical dosimeter needs to be explored. We have studied the performance of MOSFET as a relative dosimeter in small field electron beams. The reproducibility, linearity and sensitivity of a high-sensitivity microMOSFET were investigated for clinical electron beams. In addition, the percent depth doses, output factors and profiles have been measured in a water tank with MOSFET and compared with those measured by an ion chamber for a range of field sizes from 1 cm diameter to 10 cm × 10 cm for 6, 12, 16 and 20 MeV beams. Similar comparative measurements were also per-formed with MOSFET and films in solid water phantom. The MOSFET sensitivity was found to be practically constant over the range of field sizes investigated. The dose response was found to be linear and reproducible (within ± 1% for 100 cGy). An excellent agreement was observed among the central axis depth dose curves measured using MOSFET, film and ion chamber. The output factors measured with MOSFET for small fields agreed to within 3% with those measured by film dosimetry. Overall results indicate that MOSFET can be utilized to perform dosimetry for small field electron beam.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 21330970      PMCID: PMC5718583          DOI: 10.1120/jacmp.v12i1.3267

Source DB:  PubMed          Journal:  J Appl Clin Med Phys        ISSN: 1526-9914            Impact factor:   2.102


I. INTRODUCTION

Small superficial cancerous lesions are typically treated with electrons. Depending upon the size, shape and location of the lesion, the field aperture, commonly referred to as cutout, is custom‐made using cerrobend. Accurate dosimetry of such small electron fields usually requires dosimetric measurement. Accurate dosimetry for such small fields is challenging due to the loss of lateral electronic equilibrium within the field. This can result in a shift of towards the surface and other modifications of the depth dose curve, as well as a change in the beam profile characteristics at depth. The use of such fields in the clinic thus requires careful and detailed measurements to characterize both the output and coverage achieved. Small field dosimetry is typically achieved using film dosimetry.( ) However, modern radiotherapy clinics have increasingly reduced their use of film, making it difficult to maintain development systems in a condition to support dosimetric measurements. Many new clinics have eliminated film processing completely from the clinical area so that dosimetric film processing is no longer an option. While radiochromic film can be used for these measurements,( – ) this dosimetry system requires specialized equipment which may not be available in some clinics, in addition to a careful measurement procedure to obtain accurate measures. Consequently, other dose measurement methods need to be considered, especially for use in routine clinical measurements which need to be assessed in time periods shorter than a day. Conventional dosimetry equipment, such as Farmer type ion chambers and standard parallel plate chambers, is difficult to use for small electron field measurements due to the loss of electronic equilibrium and partial volume irradiation effects.( , ) Small area detectors such as diodes can be used, but are typically used in water and cannot be easily positioned in the water‐equivalent plastics favored for routine clinical work. In recent years, a new type of solid state dosimeter, the MOSFET detector, has been introduced into the clinic, primarily for in vivo patient dosimetry. In this work, we have investigated extending the use of the MOSFET to small field electron dosimetry. MOSFET dosimeters have been widely studied for photon beam dosimetry( – ) and have been used extensively for in vivo clinical dose measurements,( , , – ) as well as for small field photon dosimetry.( ) MOSFET systems have recently been investigated for use in electron beam dosimetry,( – ) as well. In this study, the application of the mobileMOSFET dosimeter system (Best Medical Canada, Ottawa, ON) to small field electron dosimetry is characterized and compared to conventional film and ionization chamber systems for determining required critical parameters for patient treatments.

II. MATERIALS AND METHODS

The mobileMOSFET system (TN‐RD‐70‐W), along with a set of high‐sensitivity microMOSFET (TN‐1002RDM), was used for this investigation. The sensitive detector region has dimensions of , with a layer thickness of 0.5 μm. In all measurements described here, the detector was oriented with the smallest dimension aligned to the beam axis. MOSFETs were irradiated under high bias setting. Dose measurements were made on a Varian linear accelerator (Clinac 2100EX, Varian Medical Systems, Palo Alto, CA), which provides 6, 9, 12, 16 and 20 MeV electron beams. Small circular fields ranging from 1 to 6 cm in diameter were fabricated, using cerrobend, for the applicator. First, the reproducibility, linearity and sensitivity of MOSFET were evaluated for all electron energies using a field and by placing the MOSFET at the respective depth in the solid water phantom. Subsequently, extensive measurements have been made for small fields using MOSFET, film and ion chamber. Each MOSFET was separately calibrated at a normalization depth of 1.5 cm in solid water for 6 MeV, 2.0 cm for 9 MeV and 3.0 cm for 12, 16 and 20 MeV. The dose versus optical density calibration curve for the films (Kodak X‐Omat V films, Eastman Kodak Company, Rochester, NY) were determined for each of the electron energies. For point dosimetry, the films were read using a manual densitometer (Macbeth, model:TD932, Kollmorgen Instruments Corporation, New Windsor, NY). The relative dose profiles were read off the films using VXR‐16 DOSIMETRYPRO scanner (Vidar System Corporation, Herndon, VA) and analyzed by RIT113 commercial software (V4.4, Radiological Imaging Technology Inc., Colorado Springs, CO). The selection of appropriate measuring devices is particularly important for small field electron dosimetry. To illustrate this point, a range of equipment available to a typical radiotherapy clinic was used to perform dosimetry measurements. The relative output in phantom at a depth of the nominal dose maximum ( for field) was determined using MOSFET, film and ion chamber for 50 MU irradiations using various field sizes. A commonly available parallel plate ion chamber (Capintec Inc, Ramsey, NJ: active volume, 1.6 cm active volume diameter, 2 cm cavity diameter) was used. Measurements were performed in a solid water phantom with 6, 12, 16 and 20 MeV electron beams.

A. Percentage depth dose (PDD)

To investigate the performance of MOSFET at various depths, the percentage depth doses for 12 MeV electrons using field were measured and compared with those obtained by an ion chamber (NE2571, Thermo Electron Corporation, Cambridge, UK) in a water tank (Med‐Tec Inc., South Plainfield, NJ). Corresponding measurements were performed with film stacked in between solid water slab and positioned perpendicular to the beam axis. Film depths within the stack were corrected to account for film thickness. To take into account the effective point of measurement of the cylindrical ion chamber, the measured depth ionization data were shifted upstream by 0.5 times the radius of the chamber cavity. A special holder (Fig. 1) was designed to keep the MOSFET in the desired position inside the water tank.
Figure 1

Solid water MOSFET holder to keep MOSFET in the desired position.

Solid water MOSFET holder to keep MOSFET in the desired position.

B. BEAM PROFILE

To assess the performance of MOSFET in the crossbeam dimension of electron beams, profiles of 6, 12, 16 and 20 MeV electrons were measured at 1.5 cm depth in a water tank for 3 cm diameter circular field and compared with those measured with film (in solid water).

C. OUTPUT FACTOR

The output factor, as defined by the ratio of the dose at for the field of interest to the dose at for the field were measured for small fields using MOSFET, as well as with films. The points of measurements (i.e., the depths of corresponding maximum dose) were determined from the respective PDD measured in the Med‐Tec water tank using MOSFET.

III. RESULTS

The MOSFET responses were found to be highly linear () over the measured dose range of 1 cGy to 1000 cGy. The reproducibility, defined as the standard deviation of 12 sequential measurements expressed as a percent of mean value at the dose level of 100 cGy, was found to be . The dose‐responses for 6, 9, 12, 16 and 20 MeV electrons were found to be , , , , and mV/cGy, respectively. These results indicate that the electron energy does not significantly influence the MOSFET sensitivity. The relative dose factors, measured at the nominal , are presented in Fig. 2. The results obtained by MOSFET, film and ion chamber agreed well for field sizes equal to and higher than 4 cm diameter. For the smaller field sizes, however, the relative dose factors measured by the ion chamber deviates significantly (up to 47.9%) from those measured by MOSFET and film. These discrepancies are due to the large cavity dimensions (2 cm diameter) of the ion chamber relative to the field width, which consequently perturb the dose measurement.( – ) Overall, the results obtained by MOSFET were found to agree with films to within 3%.
Figure 2

Relative dose factors for circular apertures at nominal , normalized to the field. Measurements were performed for each of 4 electron beam energies using a parallel plate chamber (red □), MOSFET (green Δ) and film (blue O).

Relative dose factors for circular apertures at nominal , normalized to the field. Measurements were performed for each of 4 electron beam energies using a parallel plate chamber (red □), MOSFET (green Δ) and film (blue O). Comparative central axis percentage depth dose (PDD) data for 12 MeV beams measured with film, ion chamber and MOSFET are shown in Fig. 3. The figure clearly demonstrates the significant shifts of PDD and with change in field size towards the surface. As shown, the agreements among the data were found to be within 2% (in the low‐gradient region) and within 2 mm (in the high‐gradient region). The summary of the agreements among the interesting depth dose parameters (, , and ) are presented in Table 1.
Figure 3

Performance of MOSFET in determining the variation of central axis percentage depth dose (PDD) for 12 MeV electrons, as a function of field size. For field the reference PDD was measured with ion chamber (model: NE2571, 0.6 cc) and for smaller fields the PDD were measured with film.

Table 1

Measured depth of clinical parameters for 12 MeV beams , , and for 1, 1.5, 2, 2.5, 3, 4, 6 cm circular fields and field.

Cutout Size 1.5 cm 3 cm 6 cm 10×10cm2
dmax (cm)Film0.851.62.83.0
MOSFET0.91.73.03.0
d90 (cm)Film1.72.83.83.9
MOSFET1.72.83.73.9
d80 (cm)Film2.13.34.254.3
MOSFET2.13.44.14.3
d50 (cm)Film3.04.54.95.0
MOSFET2.84.44.855.05
Measured depth of clinical parameters for 12 MeV beams , , and for 1, 1.5, 2, 2.5, 3, 4, 6 cm circular fields and field. Performance of MOSFET in determining the variation of central axis percentage depth dose (PDD) for 12 MeV electrons, as a function of field size. For field the reference PDD was measured with ion chamber (model: NE2571, 0.6 cc) and for smaller fields the PDD were measured with film. The 6, 12, 16, and 20 MeV crossbeam profiles are presented in Fig. 4 for the 3 cm circular field measured at 1.5 cm depth utilizing MOSFET and film dosimeters. The MOSFET data agrees within 1 mm with the profile measured by the film.
Figure 4

Dose profiles at the depth of 1.5 cm, measured with film (—) and MOSFET (O) for 3 cm circular field.

Dose profiles at the depth of 1.5 cm, measured with film (—) and MOSFET (O) for 3 cm circular field. The output factors for small fields measured by MOSFET and film are shown in Fig. 5. The average agreements between the results obtained by MOSFET with those by film were found to be (), () and () for 6, 12 and 20 MeV beams, respectively. The maximum deviation was 3.1% for 6 MeV beam with a 1.5 cm diameter cutout.
Figure 5

A comparison of output factors measured using MOSFET and film for the circular field sizes of 1, 1.5, 2, 2.5, 3, 4, and 6.

A comparison of output factors measured using MOSFET and film for the circular field sizes of 1, 1.5, 2, 2.5, 3, 4, and 6.

IV. DISCUSSION

The results presented in this report clearly demonstrate the challenges associated with the dosimetry of small field electron beams. One must take into account the shifts in the PDD for small fields, with respect to the reference PDD. Otherwise, substantial error may be made in the dose coverage of the target volume. Furthermore, a significant error can be made if an appropriate detector is not used in measuring the relative dose output. The performance of MOSFET with its response linearity, insensitivity to energy variation and small size (being a “point detector”) makes it suitable as a practical dosimeter for small field electron dosimetry. The work performed for this report required time‐consuming, point‐by‐point measurement with MOSFET for PDD and crossbeam profiles. However, suitable arrays of MOSFET and MOSFET holder can be developed for PDD and output measurements.

V. CONCLUSIONS

In this work the challenges associated with the dosimetry of small electron beams have been demonstrated. The PDD and shift towards the surface with changes in field size. Measurement of output for field sizes less than 4 cm diameter can be substantially lower if a regular size detector is used. The use of MOSFET detectors has been investigated for various small fields with a range of electron energies. The results indicate that MOSFET detectors can be used for routine clinical dosimetry of electron beams.
  24 in total

1.  The volume effect of detectors in the dosimetry of small fields used in IMRT.

Authors:  Wolfram U Laub; Tony Wong
Journal:  Med Phys       Date:  2003-03       Impact factor: 4.071

2.  Influence of detector size in photon beam profile measurements.

Authors:  C H Sibata; H C Mota; A S Beddar; P D Higgins; K H Shin
Journal:  Phys Med Biol       Date:  1991-05       Impact factor: 3.609

3.  Application of commercial MOSFET detectors for in vivo dosimetry in the therapeutic x-ray range from 80 kV to 250 kV.

Authors:  Christian Ehringfeld; Susanne Schmid; Karin Poljanc; Christian Kirisits; Hannes Aiginger; Dietmar Georg
Journal:  Phys Med Biol       Date:  2005-01-21       Impact factor: 3.609

4.  In vivo dosimetry with MOSFETs: dosimetric characterization and first clinical results in intraoperative radiotherapy.

Authors:  Rita Consorti; Assunta Petrucci; Falbo Fortunato; Antonella Soriani; Simona Marzi; Giuseppe Iaccarino; Valeria Landoni; Marcello Benassi
Journal:  Int J Radiat Oncol Biol Phys       Date:  2005-11-01       Impact factor: 7.038

5.  Dosimetry characteristics of GAFCHROMIC EBT film responding to therapeutic electron beams.

Authors:  Fan-Chi Su; Yaxi Liu; Sotirios Stathakis; Chengyu Shi; Carlos Esquivel; Niko Papanikolaou
Journal:  Appl Radiat Isot       Date:  2007-05-29       Impact factor: 1.513

6.  Monte Carlo simulation of MOSFET dosimeter for electron backscatter using the GEANT4 code.

Authors:  James C L Chow; Michael K K Leung
Journal:  Med Phys       Date:  2008-06       Impact factor: 4.071

7.  Dosimetric characteristics of a MOSFET dosimeter for clinical electron beams.

Authors:  D Manigandan; G Bharanidharan; P Aruna; K Devan; D Elangovan; Vikram Patil; R Tamilarasan; S Vasanthan; S Ganesan
Journal:  Phys Med       Date:  2009-01-06       Impact factor: 2.685

8.  Clinical dosimetry using MOSFETs.

Authors:  R Ramani; S Russell; P O'Brien
Journal:  Int J Radiat Oncol Biol Phys       Date:  1997-03-01       Impact factor: 7.038

9.  Clinical implementation of MOSFET detectors for dosimetry in electron beams.

Authors:  Esther J Bloemen-van Gurp; Andre W H Minken; Ben J Mijnheer; Cary J G Dehing-Oberye; Philippe Lambin
Journal:  Radiother Oncol       Date:  2006-08-17       Impact factor: 6.280

10.  Measurements of dose distributions in small beams of 6 MV x-rays.

Authors:  R K Rice; J L Hansen; G K Svensson; R L Siddon
Journal:  Phys Med Biol       Date:  1987-09       Impact factor: 3.609

View more
  2 in total

1.  Response of Nanodot Optically Stimulated Luminescence Dosimeters to Therapeutic Electron Beams.

Authors:  Y Retna Ponmalar; Ravikumar Manickam; S Sathiyan; K M Ganesh; R Arun; Henry Finlay Godson
Journal:  J Med Phys       Date:  2017 Jan-Mar

2.  A strategy to determine off-axis dosimetric leaf gap using OSLD and EPID.

Authors:  Janahiraman Divyalakshmi; K Mohamathu Rafic; Ebenezer Suman Babu; Timothy Peace Balasingh; Amalan Sebastin; Christopher J Sujith; L Jose Solomon Raj
Journal:  Rep Pract Oncol Radiother       Date:  2021-12-30
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.