| Literature DB >> 34845538 |
Nadja Trageser1, Axel Sauerwald2, Sebastian Ludwig3, Wolfram Malter3, Kilian Wegmann4, Leonidas Karapanos5, Julia Radosa6, Alina Katharina Jansen3, Christian Eichler3,7.
Abstract
PURPOSE: Many different surgical approaches have been established for the repair of a pelvic organ prolapse. Especially in laparoscopic surgery, it is important to generate easy surgical techniques with similar stability. This study shall simplify the choice of mesh by evaluating three polypropylene meshes regarding their biomechanical properties.Entities:
Keywords: Biomechanical testing; Laparoscopy; Pelvic organ prolapse; Polypropylene mesh; Sacrocervicopexy; Uro-gynecological surgery
Mesh:
Substances:
Year: 2021 PMID: 34845538 PMCID: PMC8918124 DOI: 10.1007/s00404-021-06344-9
Source DB: PubMed Journal: Arch Gynecol Obstet ISSN: 0932-0067 Impact factor: 2.344
Fig. 1Testing setup. a SERATEX® SlimSling® with unilateral fixation. b The same mesh with a bilateral fixation. c SERATEX® E11 PA and d SERATEX® B3 PA
Fig. 2Knotting method for SERATEX® SlimSling®. Instead of a common surgical knot, the needle is threaded through the fifth loop of the mesh
Fig. 3Testing frame and evaluated meshes. a Shown is the testing frame that was used. The black asterisk marks the parallel screw clamps that were used for the apical fixation of the meshes. b All three evaluated meshes are shown. From left to right: SERATEX® B3 PA, SERATEX® E11 PA, SERATEX® SlimSling® (by courtesy of SERAG-WIESSNER GmbH & Co. KG)
Overall results for all four evaluated groups
| Evaluated entity | Displacement at failure (mm) | Maximum load (N) | Stiffness (N/mm) | Failure mode | |
|---|---|---|---|---|---|
| SlimSling® unilateral fixation | 12 | 30 (± 4) | 30 (± 2) | 0.91 (± 0.19) | Mesh failure (12/12) |
| SlimSling® bilateral fixation | 10 | 29 (± 5) | 53 (± 7) | 1.82 (± 0.14) | Mesh failure (10/10) |
| SERATEX® E11 PA | 9 | 71 (± 12) | 199 (± 29) | 3.93 (± 0.59) | Mesh failure (6/9) |
| Tissue failure (3/9) | |||||
| SERATEX® B3 PA | 10 | 33 (± 4) | 66 (± 9) | 2.11 (± 0.35) | Mesh failure (10/10) |
Fig. 4Comparison of all evaluated groups for displacement at failure and maximum load. Shown are the results of displacement at failure and maximum load for all evaluated groups. Error bars represent standard deviations. The blue box in the upper left of the figure shows the results of the ANOVA analysis. SSU SERATEX® SlimSling® with unilateral fixation, SSB SERATEX® SlimSling® with bilateral fixation, E11 SERATEX® E11 PA, B3 SERATEX® B3 PA. The green asterisk marks the maximum load of SERATEX® SlimSling® with unilateral fixation. It is significantly lower than maximum loads generated by the other fixation methods. The red asterisk marks the displacement at failure of SERATEX® E11 PA. The displacement is significantly higher than displacements generated by the other fixation methods. The yellow asterisk marks the maximum load of SERATEX® E11 PA. It is significantly higher than maximum loads generated by the other fixation methods
Fig. 5Comparison of all evaluated groups for the parameter stiffness. Error bars represent standard deviations. The blue box in the upper left of the figure shows the results of the ANOVA analysis. SSU SERATEX® SlimSling® with unilateral fixation, SSB SERATEX® SlimSling® with bilateral fixation, E11 SERATEX® E11 PA, B3 SERATEX® B3 PA. The red asterisk marks the evaluated value of the parameter stiffness of SERATEX® SlimSling® with unilateral fixation. It is significantly lower than the stiffness of the other meshes. The yellow asterisk marks the evaluated value of the parameter stiffness of SERATEX® E11 PA. It is significantly higher than the stiffness of the other meshes
Fig. 6Most common failure mode (mesh failure) for all evaluated groups. a SERATEX® SlimSling® with unilateral fixation. b SERATEX® SlimSling® with bilateral fixation. c SERATEX® E11 PA. d SERATEX® B3 PA
Fig. 7Tissue failure of SERATEX® E11 PA
Fig. 8Representative force–deflection graph