A Sauerwald1, M Niggl2,3, J Puppe4, A Prescher5, M Scaal6, G K Noé7, S Schiermeier8, M Warm2,4, C Eichler2,3.
Abstract
INTRODUCTION: Pectopexy, a laparoscopic method for prolapse surgery, showed promising results in recent literature. Further improving this approach by reducing surgical time may decrease complication rates and patient morbidity. Since laparoscopic suturing is a time consuming task, we propose a single suture /mesh ileo-pectineal ligament fixation as opposed to the commonly used continues approach.Entities:
Mesh:
Year: 2016 PMID: 26844890 PMCID: PMC4741420 DOI: 10.1371/journal.pone.0144143
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Fig 1Shown are representative images of the 4 evaluated groups.
Fig 1.1 shows mesh + continuous suture (MCS). Fig 1.2 shows mesh + a single interrupted suture (MIS). Fig 1.3 shows a continuous suture (CS) and Fig 1.4 shows a single interrupted suture (IS).
This table show overall results for all four evaluated groups.
| Evaluated Entity | Number Tested | Ultimate Load | Displacement at Failure | Stiffness | Calculated Load at 2 mm | Failure Mode |
|---|---|---|---|---|---|---|
| Total Trials = 33 | n | N | mm | N /mm | N | |
| Mesh + Interrupted Suture | 9 | 35 (± 12) | 37 (± 12) | 1,76 (± 0.32) | 3.5 (± 0.6) | Individual loop failure (9/9) |
| Mesh + Continuous Suture | 8 | 48 (± 7) | 36 (± 5) | 2,51 (± 0.53) | 5.0 (± 1.1) | Global mesh failure (8/8) |
| Only Interrupted Suture | 8 | 85 (± 27) | 21 (± 4) | 4,65 (± 1.06) | 9,3 (± 2.1) | Ligament failure (5/8) Suture failure (3/8) |
| Only Continuous Suture | 8 | 74 (± 24) | 28 (± 8) | 2,83 (± 0.46) | 5.6 (± 0.9) | Ligament failure (3/8) Suture failure (5/8) |
Fig 2Shown are maximum loads and displacement at failure for all four subgroups.
ANOVA analysis results are shown as well. Error bars represent standard deviations. M+IS = mesh + interrupted (single) suture, M + CS = mesh + continuous suture, IS = interrupted (single) suture, CS = continuous suture.
Fig 3Shown is fixation method stiffness for all four subgroups.
ANOVA analysis results are shown as well. Error bars represent standard deviations. M+IS = mesh + interrupted (single) suture, M + CS = mesh + continuous suture, IS = interrupted (single) suture, CS = continuous suture.