| Literature DB >> 34722939 |
Maddison Shaw1,2, Jessica Lye1,3, Andrew Alves1, Stephanie Keehan1, Joerg Lehmann4,5,6,7, Maximilian Hanlon1, John Kenny8, John Baines9, Claudiu Porumb10, Moshi Geso2, Rhonda Brown1.
Abstract
BACKGROUND ANDEntities:
Keywords: Dosimetry audit; Quality assurance; SBRT; SRS; Small field dosimetry; microDiamond
Year: 2021 PMID: 34722939 PMCID: PMC8536779 DOI: 10.1016/j.phro.2021.10.002
Source DB: PubMed Journal: Phys Imaging Radiat Oncol ISSN: 2405-6316
Fig. 1(a) PTW microDiamond detector in the beam parallel orientation, (b) ACDS cranial phantom with two craniocaudal microDiamond detectors and (c) ACDS thorax phantom with microDiamond detector measurement in lung target. Typical orientations of the beam relative to the detector axis are shown in the schematics.
PTW 60019 microDiamond cross-calibration factors for 6 MV and Co-60 Gamma Knife.
| microDiamond sn | NDw,Q_cross (6MV) | Total calibration factor (Co-60 Gamma Knife) |
|---|---|---|
| 122179 | 1.017 | – |
| 123242 | 0.777 | 0.770 |
| 123244 | 0.766 | – |
| 123669 | 1.199 | 1.201 |
| 124007 | 0.780 | 0.781 |
| 124008 | 0.935 | 0.935 |
| 124146 | 0.860 | – |
Fig. 2(a) Average energy dependence, relative to 6MV, of NDw,Q_cross calibration factors for seven PTW 60019 microDiamond detectors in the beam parallel orientation. The error bars show standard uncertainty from calibration of seven detectors. (b) Average parallel vs perpendicular correction factors (kperpendicular) for nominal beam field size at detector EPOM and nominal beam energies, for the seven detectors. The error bars show the standard uncertainty.
Fig. 3Non-coplanar angular dependence of microDiamond for 6 MV and 6 FFF with 3 × 3 cm2 and 1.5 × 1.5 cm2 fields, compared to RAZOR chamber. In each grouping of couch angle, the data series shown left to right represent the gantry angles of 30°, 60°, and 90° respectively.
Fig. 4(a) Roll response of three microDiamond detectors measured with 6 MV, 6 FFF, 10 MV and 10 FFF on a VersaHD Linac. The roll response at each position is relative to the average response of all positions. (b) Roll response of microDiamond sn123244 in Conventional 6 FFF Linac, and inline and perpendicular MR-Linac configurations.
Overall point dose measurement in SRS/SBRT treatment plan uncertainty budget.
| Quantity | Relative standard uncertainty | |
|---|---|---|
| 100 uiA1 | 100 uiB2 | |
| Phantom/detector positioning uncertainty | – | 0.8 |
| kT | – | 0.0 |
| kP | – | 0.0 |
| kH | – | 0.0 |
| Electrometer calibration factor | – | 0.15 |
| NDw, Q_cross calibration factor | – | 0.73 |
| Non-coplanar angular dependence | – | 0.9 |
| Coplanar roll response | – | 0.3 |
| Linearity | – | 0.5 |
| Dose rate | – | 0.7 |
| ks (recombination) | – | 0.0 |
| kpol (polarity) | – | 0.0 |
| kn (beam non uniformity) | – | 0.0 |
| (field output correction factor) | – | 0.4 |
| Reproducibility of SABR delivery and detector charge measurements | 0.3 | – |
| Facility daily output | – | 0.4 |
1uiA represents the relative standard uncertainty estimated by statistical methods, type A.
2uiB represents the relative standard uncertainty estimated by other means, type B.