How the detector resolution affects the clinical significance of SBRT pre-treatment quality assurance results.

A Bruschi1, M Esposito2, S Pini2, A Ghirelli2, G Zatelli2, S Russo3.   

Abstract

PURPOSE: Aim of this work was to study how the detector resolution can affect the clinical significance of SBRT pre-treatment volumetric modulated arc therapy (VMAT) verification results.
METHODS: Three detectors (PTW OCTAVIUS 4D 729, 1500 and 100 SRS) used in five configurations with different resolution were compared: 729, 729 merged, 1500, 1500 merged and 1000 SRS. Absolute local gamma passing rates of 3D pre-treatment quality assurance (QA) were evaluated for 150 dose distributions in 30 plans. Five different kinds of error were introduced in order to establish the detection sensitivity of the three devices. Percentage dosimetric differences were evaluated between planned dosevolume histogram (DVH) and patients' predicted DVH calculated by PTW DVH 4D® software.
RESULTS: The mean gamma passing rates and the standard deviations were 92.4% ± 3.7%, 94.6% ± 1.8%, 95.3% ± 4.2%, 97.4% ± 2.5% and 97.6% ± 1.4 respectively for 729, 729 merged, 1500, 1500 merged and 1000 SRS with 2% local dose/2mm criterion. The same trend was found on the sensitivity analysis: using a tight gamma analysis criterion (2%L/1mm) only the 1000 SRS detected every kind of error, while 729 and 1500 merged detected three and four kinds of error respectively. Regarding dose metrics extracted from DVH curves, D50% was within the tolerance level in more than 90% of cases only for the 1000 SRS.
CONCLUSIONS: The detector resolution can significantly affect the clinical significance of SBRT pre-treatment verification results. The choice of a detector with resolution suitable to the investigated field size is of main importance to avoid getting false positive.
Copyright © 2017 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  2D array resolution; DVH metric; SBRT QA

Mesh:

Year:  2017        PMID: 29203119     DOI: 10.1016/j.ejmp.2017.11.012

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  8 in total

1.  Application of TG-218 action limits to SRS and SBRT pre-treatment patient specific QA.

Authors:  Yuqing Xia; Justus Adamson; Yana Zlateva; Will Giles
Journal:  J Radiosurg SBRT       Date:  2020

2.  A patient-specific QA comparison between 2D and 3D diode arrays for single-lesion SRS and SBRT treatments.

Authors:  Yongsook C Lee; Yongbok Kim
Journal:  J Radiosurg SBRT       Date:  2021

3.  Operation and calibration of the novel PTW 1600SRS detector for the verification of single isocenter stereotactic radiosurgery treatments of multiple small brain metastases.

Authors:  Esther Decabooter; Ans Cc Swinnen; Michel C Öllers; Fabian Göpfert; Frank Verhaegen
Journal:  Br J Radiol       Date:  2021-06-11       Impact factor: 3.629

4.  Patient-specific dose quality assurance of single-isocenter multiple brain metastasis stereotactic radiosurgery using PTW Octavius 4D.

Authors:  James McCulloch; Jamie Pawlowski; Neil Kirby; Karl Rasmussen; Zheng Shi; Pamela Myers; Sotirios Stathakis; Niko Papanikolaou; Daniel L Saenz
Journal:  J Appl Clin Med Phys       Date:  2020-07-18       Impact factor: 2.102

5.  Characterisation of a synthetic diamond detector for end-to-end dosimetry in stereotactic body radiotherapy and radiosurgery.

Authors:  Maddison Shaw; Jessica Lye; Andrew Alves; Stephanie Keehan; Joerg Lehmann; Maximilian Hanlon; John Kenny; John Baines; Claudiu Porumb; Moshi Geso; Rhonda Brown
Journal:  Phys Imaging Radiat Oncol       Date:  2021-10-19

6.  Evaluation of parameters affecting gamma passing rate in patient-specific QAs for multiple brain lesions IMRS treatments using ray-station treatment planning system.

Authors:  Elahheh Salari; E Ishmael Parsai; Diana Shvydka; Nicholas Niven Sperling
Journal:  J Appl Clin Med Phys       Date:  2021-11-18       Impact factor: 2.102

7.  Practical application of Octavius® -4D: Characteristics and criticalities for IMRT and VMAT verification.

Authors:  Patrizia Urso; Rita Lorusso; Luca Marzoli; Daniela Corletto; Paolo Imperiale; Annalisa Pepe; Lorenzo Bianchi
Journal:  J Appl Clin Med Phys       Date:  2018-07-16       Impact factor: 2.102

8.  Assessment of combined use of ArcCheck® detector and portal dosimetry for delivery quality assurance of head and neck and prostate volumetric-modulated arc therapy.

Authors:  Gilles Moliner; Lise Sorro; Rodolfe Verstraet; Paul Alexandre Daviau; Mélanie Casas; Bérengère Piron; Karine Dubois; Charles Debrigode; Corinne Barrau; Françoise Bons; Joël Greffier
Journal:  J Appl Clin Med Phys       Date:  2018-10-19       Impact factor: 2.102

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.