Literature DB >> 29244899

Spatial response of synthetic microDiamond and diode detectors measured with kilovoltage synchrotron radiation.

Duncan J Butler1, Toby Beveridge1, Joerg Lehmann2, Christopher P Oliver1, Andrew W Stevenson3,4, Jayde Livingstone3.   

Abstract

PURPOSE: To map the spatial response of four solid-state radiation detectors of types commonly used for radiotherapy dosimetry.
METHODS: PTW model 60016 Diode P, 60017 Diode E, 60018 Diode SRS, and 60019 microDiamond detectors were radiographed using a high resolution conventional X-ray system. Their spatial response was then investigated using a 0.1 mm diameter beam of 95 keV average energy photons generated by a synchrotron. The detectors were scanned through the beam while their signal was recorded as a function of position, to map the response. These 2D response maps were created in both the end-on and side-on orientations.
RESULTS: The results show the location and size of the active region. End-on, the active area was determined to be centrally located and within 0.2 mm of the manufacturer's specified diameter. The active areas of the 60016 Diode P, 60017 Diode E, 60018 Diode SRS detectors are uniform to within approximately 5%. The 60019 microDiamond showed local variations up to 30%. The extra-cameral signal in the microDiamond was calculated from the side-on scan to be approximately 8% of the signal from the active element.
CONCLUSIONS: The spatial response of four solid-state detectors has been measured. The technique yielded information about the location and uniformity of the active area, and the extra-cameral signal, for the beam quality used.
© 2017 Commonwealth of Australia. Medical Physics © 2017 American Association of Physicists in Medicine. This work is copyright. Apart from any use as permitted under the Copyright Act 1968, no part may be reproduced without prior written permission. Requests and enquiries concerning reproduction and rights should be directed in the first instance to John Wiley & Sons Ltd of The Atrium, Southern Gate, Chichester, West Sussex P019 8SQ UNITED KINGDOM; alternatively to ARPANSA.

Entities:  

Keywords:  dosimetry; high resolution; kilovoltage; microDiamond; solid-state detectors; synchrotron

Mesh:

Substances:

Year:  2018        PMID: 29244899     DOI: 10.1002/mp.12733

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  5 in total

1.  The role of radiation-induced charge imbalance on the dose-response of a commercial synthetic diamond detector in small field dosimetry.

Authors:  Hui Khee Looe; Daniela Poppinga; Rafael Kranzer; Isabel Büsing; Tuba Tekin; Ann-Britt Ulrichs; Björn Delfs; Dennis Vogt; Jan Würfel; Björn Poppe
Journal:  Med Phys       Date:  2019-05-02       Impact factor: 4.071

2.  Three-dimensional characterization of the active volumes of PTW microDiamond, microSilicon, and Diode E dosimetry detectors using a proton microbeam.

Authors:  Daniela Poppinga; Rafael Kranzer; Ann-Britt Ulrichs; Björn Delfs; Ulrich Giesen; Frank Langner; Björn Poppe; Hui Khee Looe
Journal:  Med Phys       Date:  2019-07-22       Impact factor: 4.071

3.  Evaluation of the PTW microDiamond in edge-on orientation for dosimetry in small fields.

Authors:  Owen J Brace; Sultan F Alhujaili; Jason R Paino; Duncan J Butler; Dean Wilkinson; Brad M Oborn; Anatoly B Rosenfeld; Michael L F Lerch; Marco Petasecca; Jeremy A Davis
Journal:  J Appl Clin Med Phys       Date:  2020-05-22       Impact factor: 2.102

4.  Evaluation of silicon strip detectors in transmission mode for online beam monitoring in microbeam radiation therapy at the Australian Synchrotron.

Authors:  Jeremy Davis; Andrew Dipuglia; Matthew Cameron; Jason Paino; Ashley Cullen; Susanna Guatelli; Marco Petasecca; Anatoly Rosenfeld; Michael Lerch
Journal:  J Synchrotron Radiat       Date:  2022-01-01       Impact factor: 2.616

5.  Characterisation of a synthetic diamond detector for end-to-end dosimetry in stereotactic body radiotherapy and radiosurgery.

Authors:  Maddison Shaw; Jessica Lye; Andrew Alves; Stephanie Keehan; Joerg Lehmann; Maximilian Hanlon; John Kenny; John Baines; Claudiu Porumb; Moshi Geso; Rhonda Brown
Journal:  Phys Imaging Radiat Oncol       Date:  2021-10-19
  5 in total