Direct calibration in megavoltage photon beams using Monte Carlo conversion factor: validation and clinical implications.

Tracy Wright1, Jessica E Lye, Ganesan Ramanathan, Peter D Harty, Chris Oliver, David V Webb, Duncan J Butler.   

Abstract

The Australian Radiation Protection and Nuclear Safety Agency (ARPANSA) has established a method for ionisation chamber calibrations using megavoltage photon reference beams. The new method will reduce the calibration uncertainty compared to a (60)Co calibration combined with the TRS-398 energy correction factor. The calibration method employs a graphite calorimeter and a Monte Carlo (MC) conversion factor to convert the absolute dose to graphite to absorbed dose to water. EGSnrc is used to model the linac head and doses in the calorimeter and water phantom. The linac model is validated by comparing measured and modelled PDDs and profiles. The relative standard uncertainties in the calibration factors at the ARPANSA beam qualities were found to be 0.47% at 6 MV, 0.51% at 10 MV and 0.46% for the 18 MV beam. A comparison with the Bureau International des Poids et Mesures (BIPM) as part of the key comparison BIPM.RI(I)-K6 gave results of 0.9965(55), 0.9924(60) and 0.9932(59) for the 6, 10 and 18 MV beams, respectively, with all beams within 1σ of the participant average. The measured kQ values for an NE2571 Farmer chamber were found to be lower than those in TRS-398 but are consistent with published measured and modelled values. Users can expect a shift in the calibration factor at user energies of an NE2571 chamber between 0.4-1.1% across the range of calibration energies compared to the current calibration method.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 25565406     DOI: 10.1088/0031-9155/60/2/883

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  3 in total

1.  Uncertainty in positioning ion chamber at reference depth for various water phantoms.

Authors:  Naoki Kinoshita; Hiroshi Oguchi; Toshiki Adachi; Hiroki Shioura; Hirohiko Kimura
Journal:  Rep Pract Oncol Radiother       Date:  2018-03-30

2.  Characterisation of a synthetic diamond detector for end-to-end dosimetry in stereotactic body radiotherapy and radiosurgery.

Authors:  Maddison Shaw; Jessica Lye; Andrew Alves; Stephanie Keehan; Joerg Lehmann; Maximilian Hanlon; John Kenny; John Baines; Claudiu Porumb; Moshi Geso; Rhonda Brown
Journal:  Phys Imaging Radiat Oncol       Date:  2021-10-19

3.  Commissioning of a PTW 34070 large-area plane-parallel ionization chamber for small field megavoltage photon dosimetry.

Authors:  Tom Kupfer; Joerg Lehmann; Duncan J Butler; Ganesan Ramanathan; Tracy E Bailey; Rick D Franich
Journal:  J Appl Clin Med Phys       Date:  2017-10-04       Impact factor: 2.102

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.