Juliane Conzelmann1, Ulrich Genske1,2, Arthur Emig1, Michael Scheel3, Bernd Hamm1, Paul Jahnke4,5.
Abstract
OBJECTIVES: To evaluate the effects of anatomical phantom structure on task-based image quality assessment compared with a uniform phantom background.Entities:
Keywords: Medical physics; Neck; Phantoms, imaging; Radiation protection; Tomography, X-ray computed
Mesh:
Year: 2021 PMID: 34476563 PMCID: PMC8794946 DOI: 10.1007/s00330-021-08248-3
Source DB: PubMed Journal: Eur Radiol ISSN: 0938-7994 Impact factor: 5.315
Fig. 1CT images of the uniform and the anatomical phantom. The yellow circle indicates the lesion position
Fig. 2Detection accuracy and reader confidence in uniform and anatomical phantom images. Averaged results across all readers, dose levels, and reconstruction methods at 4, 9 (uniform), 10 (anatomical), 18, 30, and 38 HU lesion contrast are presented. Error bars indicate 95% confidence intervals. 25% detection accuracy corresponds to random guessing (no detection)
Fig. 3Uniform and anatomical phantom images acquired at 1.4 and 5.6 mGy and reconstructed with filtered back projection (FBP) and adaptive iterative dose reduction 3D (AIDR 3D)
Detection accuracy (%) per dose, reconstruction method, and lesion contrast in uniform phantom images. Means and 95% confidence intervals are presented
| 4 HU lesion contrast | 9 HU lesion contrast | 18 HU lesion contrast | 30 HU lesion contrast | 38 HU lesion contrast | ||
|---|---|---|---|---|---|---|
| 1.4 mGy | FBP | 42.9 (27.3 to 58.4) | 67.9 (48.8 to 86.9) | 99.3 (97.5 to 101) | 100 | 100 |
| AIDR 3D | 67.9 (57.9 to 77.8) | 93.6 (84.4 to 102.7) | 100 | 100 | 100 | |
| 5.6 mGy | FBP | 70 (55.9 to 84.1) | 97.1 (93.5 to 100.8) | 99.3 (97.5 to 101) | 100 | 100 |
| AIDR 3D | 70.7 (56.7 to 84.7) | 99.3 (97.5 to 101) | 100 | 100 | 100 | |
Detection accuracy (%) per dose, reconstruction method, and lesion contrast in anatomical phantom images. Means and 95% confidence intervals are presented
| 10 HU lesion contrast | 18 HU lesion contrast | 30 HU lesion contrast | 38 HU lesion contrast | ||
|---|---|---|---|---|---|
| 1.4 mGy | FBP | 48.6 (26.3 to 70.9) | 45 (23 to 67) | 85.7 (66.2 to 105.2) | 93.6 (89.2 to 98) |
| AIDR 3D | 47.9 (25.5 to 70.2) | 54.3 (36.5 to 72.1) | 90.7 (79.3 to 102.2) | 99.3 (97.5 to 101) | |
| 5.6 mGy | FBP | 48.6 (26.3 to 70.9) | 56.4 (41.6 to 71.2) | 95.7 (89 to 102.5) | 97.9 (92.6 to 103.1) |
| AIDR 3D | 66.4 (51.4 to 81.5) | 66.4 (43.2 to 89.7) | 92.1 (82.2 to 102.1) | 99.3 (97.5 to 101) | |
Detection accuracy (%) per dose and reconstruction method in uniform phantom images. Averaged results across 4 and 9 HU lesion contrast and 95% confidence intervals are presented
| FBP | AIDR 3D | ||
|---|---|---|---|
| 1.4 mGy | 55.4 (42.5 to 68.2) | 80.7 (71.1 to 90.3) | < |
| 5.6 mGy | 83.6 (73.4 to 93.8) | 85 (74.6 to 95.4) | |
| < |
Fig. 4Detection accuracy per dose and reconstruction method in uniform phantom images. Averaged results across 4 and 9 HU lesion contrast are presented. Error bars indicate 95% confidence intervals. 25% detection accuracy corresponds to random guessing (no detection)
Fig. 5Detection accuracy per dose and reconstruction method in anatomical phantom images. Averaged results across all lesion contrasts are presented. Error bars indicate 95% confidence intervals. 25% detection accuracy corresponds to random guessing (no detection)
Detection accuracy (%) per dose and reconstruction method in anatomical phantom images. Averaged results across all lesion contrasts and 95% confidence intervals are presented
| FBP | AIDR 3D | ||
|---|---|---|---|
| 1.4 mGy | 68.2 (56.9 to 79.5) | 73 (62.3 to 83.7) | |
| 5.6 mGy | 74.6 (64.2 to 85.1) | 81.1 (72.8 to 89.3) | |
Fig. 6Noise and noise power spectrum (NPS) results. Average noise values from 200 images per phantom type, dose, and image reconstruction are presented. Error bars indicate standard deviations
Noise (SD) and spatial frequency (mm−1) of the peak amplitude of the noise power spectrum (NPS). Averaged noise results from 200 images per phantom type, dose, and image reconstruction are presented along with 95% confidence intervals
| 1.4 mGy FBP | 1.4 mGy AIDR 3D | 5.6 mGy FBP | 5.6 mGy AIDR 3D | |
|---|---|---|---|---|
Noise (SD) Uniform | 26.33 (26.15 to 26.51) | 13.89 (13.77 to 14) | 12.79 (12.7 to 12.88) | 9.8 (9.72 to 9.88) |
Noise (SD) Anatomical | 39.79 (39.5 to 40.08) | 23.76 (23.54 to 23.98) | 27.52 (27.28 to 27.75) | 23.5 (23.27 to 23.74) |
Spatial frequency of peak NPS (mm−1) Uniform | 0.28 | 0.23 | 0.29 | 0.28 |
Spatial frequency of peak NPS (mm−1) Anatomical | 0.12 | 0.12 | 0.12 | 0.12 |