Juliane Conzelmann1, Felix Benjamin Schwarz1, Bernd Hamm1, Michael Scheel2, Paul Jahnke1.
Abstract
PURPOSE: To develop a customized method to produce uniform phantoms for task-based assessment of CT image quality.Entities:
Keywords: contrast sensitivity; imaging; phantoms; tomography; x-ray computed
Mesh:
Year: 2020 PMID: 32721106 PMCID: PMC7497917 DOI: 10.1002/acm2.12974
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Fig. 1Neck computed tomography image of a patient that was used as a template for the phantom shape. The region of interest in the left parapharyngeal space indicates the position for insertion of low‐contrast signals.
Fig. 2Phantom construction. (a) The metallic markers were removed from a 12 mm diameter angioplasty balloon, which was filled with fructose solutions. (b) Nineteen PMMA slabs of 5 mm thickness were cut to the shape of a patient's neck and an opening was inserted into the left parapharyngeal space to hold the balloon. (c) The PMMA slabs were compressed between two wood panels and the angioplasty balloon was inserted. (d) Computed tomography scout image of the phantom holding the angioplasty balloon.
Fig. 3Computed tomography images of the phantom with different fructose concentrations. Displayed images were acquired with 120 mA tube current (CTDIvol 5.6 mGy) and reconstructed with FBP. Labels at the top indicate measured contrast values. The bottom row shows magnified details of the lesions in the left parapharyngeal space. All images are displayed with window level 100 and window width 350.
Fig. 4Calculated and measured lesion contrast values. Means and standard deviations across all acquisitions and reconstructions, Pearson's correlation coefficient r and the equation of the linear regression are shown.
Calculated and measured HU and contrast values. Mean ± SD of measured values averaged across all acquisitions and reconstructions are shown. Fructose HU and contrasts were calculated according to Methods Part A.
| Fructose concentration (mg/mL) | Density (g/cm3) | Calculated fructose HU | Measured fructose HU | Measured PMMA HU | Calculated contrast | Measured contrast |
|---|---|---|---|---|---|---|
| 240 | 1.090 | 81.3 | 80.8 ± 2.0 | 118.3 ± 0.8 | 37.3 | 37.5 ± 2.1 |
| 250 | 1.094 | 84.7 | 83.5 ± 2.1 | 118.2 ± 0.8 | 33.9 | 34.7 ± 2.2 |
| 260 | 1.097 | 88.1 | 87.6 ± 2.1 | 117.4 ± 1.1 | 30.5 | 29.9 ± 2.2 |
| 280 | 1.105 | 94.8 | 93.7 ± 2.3 | 117.4 ± 1.1 | 23.8 | 23.7 ± 2.4 |
| 290 | 1.109 | 98.2 | 97.2 ± 2.0 | 117.4 ± 1.1 | 20.4 | 20.1 ± 2.0 |
| 300 | 1.112 | 101.6 | 99.4 ± 2.4 | 117.4 ± 1.2 | 17 | 18.0 ± 2.4 |
| 310 | 1.116 | 105.0 | 103.9 ± 2.3 | 117.3 ± 1.2 | 13.6 | 13.4 ± 2.2 |
| 320 | 1.120 | 108.3 | 106.1 ± 2.2 | 117.5 ± 1.1 | 10.3 | 11.3 ± 2.2 |
| 330 | 1.124 | 111.7 | 108.5 ± 1.9 | 117.4 ± 1.2 | 6.9 | 8.9 ± 2.2 |
| 340 | 1.128 | 115.1 | 113.1 ± 2.1 | 117.3 ± 1.2 | 3.5 | 4.2 ± 2.0 |
Measured contrast per fructose concentration, tube current, and reconstruction method. Mean ± SD values of 45 measurements are shown (five repeated acquisitions per condition × 9 images per acquisition).
| Fructose concentration (mg/mL) | FBP | AIDR 3D | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| 10 mA | 20 mA | 30 mA | 40 mA | 100 mA | 120 mA | 10 mA | 20 mA | 30 mA | 40 mA | 100 mA | 120 mA | |
| 240 | 38.9 ± 4.5 | 37.7 ± 2.2 | 38.1 ± 1.9 | 37.0 ± 1.5 | 36.5 ± 0.9 | 36.4 ± 0.8 | 39.3 ± 2.2 | 37.6 ± 1.2 | 37.5 ± 1.2 | 37.0 ± 1.0 | 36.9 ± 0.8 | 36.7 ± 0.7 |
| 250 | 35.9 ± 4.7 | 34.8 ± 3.0 | 34.7 ± 2.2 | 34.8 ± 1.4 | 33.9 ± 0.9 | 33.5 ± 1.0 | 36.0 ± 2.1 | 35.1 ± 1.6 | 34.9 ± 1.6 | 34.7 ± 1.0 | 34.2 ± 0.8 | 33.8 ± 0.9 |
| 260 | 28.8 ± 4.2 | 30.8 ± 2.9 | 30.7 ± 1.9 | 29.6 ± 2.3 | 29.0 ± 1.0 | 29.1 ± 0.9 | 30.3 ± 2.6 | 30.7 ± 1.7 | 30.5 ± 1.4 | 29.9 ± 1.8 | 29.4 ± 0.9 | 29.5 ± 0.9 |
| 280 | 23.8 ± 5.2 | 24.1 ± 2.9 | 23.7 ± 2.5 | 23.6 ± 1.5 | 23.1 ± 1.2 | 22.7 ± 1.3 | 24.8 ± 2.4 | 24.6 ± 1.7 | 23.6 ± 2.0 | 23.9 ± 1.1 | 23.5 ± 1.1 | 23.2 ± 1.1 |
| 290 | 19.6 ± 4.1 | 21.2 ± 2.5 | 20.0 ± 2.2 | 19.7 ± 1.4 | 19.4 ± 1.0 | 19.3 ± 1.2 | 21.0 ± 1.6 | 21.5 ± 1.4 | 20.2 ± 1.6 | 20.0 ± 1.2 | 19.8 ± 0.9 | 19.7 ± 1.1 |
| 300 | 18.5 ± 5.9 | 18.7 ± 2.2 | 18.3 ± 2.3 | 16.8 ± 1.8 | 17.1 ± 1.3 | 17.3 ± 1.0 | 19.6 ± 2.5 | 18.8 ± 1.1 | 18.6 ± 1.6 | 17.5 ± 1.6 | 17.6 ± 1.2 | 17.8 ± 1.0 |
| 310 | 13.2 ± 5.5 | 13.7 ± 2.6 | 13.4 ± 2.4 | 12.6 ± 1.5 | 12.8 ± 1.2 | 12.7 ± 1.0 | 14.3 ± 1.7 | 14.2 ± 1.5 | 14.0 ± 1.4 | 13.2 ± 1.2 | 13.3 ± 1.2 | 13.2 ± 0.9 |
| 320 | 10.8 ± 4.5 | 12.1 ± 3.2 | 11.0 ± 2.1 | 10.8 ± 1.4 | 10.8 ± 1.1 | 10.3 ± 1.1 | 12.3 ± 2.2 | 12.4 ± 1.7 | 11.7 ± 1.5 | 11.5 ± 1.0 | 11.3 ± 1.0 | 10.8 ± 1.1 |
| 330 | 9.8 ± 4.3 | 9.0 ± 2.0 | 8.5 ± 2.4 | 8.1 ± 1.8 | 7.8 ± 1.6 | 8.0 ± 1.2 | 10.7 ± 1.6 | 9.8 ± 1.5 | 9.3 ± 1.7 | 8.8 ± 1.5 | 8.3 ± 1.5 | 8.6 ± 1.1 |
| 340 | 2.6 ± 3.6 | 4.4 ± 2.7 | 3.9 ± 2.2 | 3.4 ± 1.7 | 3.8 ± 1.2 | 3.8 ± 1.2 | 5.0 ± 1.7 | 5.7 ± 1.4 | 4.8 ± 1.7 | 4.4 ± 1.5 | 4.3 ± 1.1 | 4.3 ± 1.2 |
Fig. 5Comparison of detection accuracy (left) and confidence scores (right) between low and higher contrast signals. Pooled reader results across all tube currents and image reconstructions are shown. Crosses indicate mean values.
Detection accuracy per signal contrast, tube current, and reconstruction method. Mean ± SD values are shown. CTDIvol values were 1.4 mGy for 30 mA and 5.6 mGy for 120 mA tube current.
| Signal contrast (HU) | 30 mA FBP | 30 mA AIDR 3D | 120 mA FBP | 120 mA AIDR 3D |
|---|---|---|---|---|
| 4 | 42.9 ± 16.8 % | 67.9 ± 10.8 % | 70.0 ± 15.3 % | 70.7 ± 15.1 % |
| 20 | 98.6 ± 2.4 % | 100.0 ± 0.0 % | 100.0 ± 0.0 % | 100.0 ± 0.0 % |
Reader confidence per signal contrast, tube current, and reconstruction method. Mean ± SD values are shown. CTDIvol values were 1.4 mGy for 30 mA and 5.6 mGy for 120 mA tube current.
| Signal contrast (HU) | 30 mA FBP | 30 mA AIDR 3D | 120 mA FBP | 120 mA AIDR 3D |
|---|---|---|---|---|
| 4 | 1.41 ± 0.58 | 1.59 ± 0.64 | 1.69 ± 0.84 | 1.63 ± 0.77 |
| 20 | 3.91 ± 1.19 | 4.26 ± 1.01 | 4.37 ± 0.76 | 4.54 ± 0.75 |